ACL TOP 500 CTS 00000280040RRecalls & reported problems

Instrumentation Laboratory Company · ACL TOP 500 CTS (Refurbished) · Class 2

All models of ACL TOP 500 CTS (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 22, 2022Class IIOpen, Classified

    Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

    Z-1014-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other INSTRUMENT, COAGULATION, AUTOMATED devices

All INSTRUMENT, COAGULATION, AUTOMATED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page