ACL TOP 550 CTS 00000280045Recalls & reported problems
Instrumentation Laboratory Company · ACL TOP 550 CTS · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08426950729242), received between Mar 1, 2019 and Nov 6, 2024 — classified as 3 injuries.
The most frequently reported specific device problem is “Use of Device Problem” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 13, 2023Class IIOpen, Classified
Sample misidentification could occur under specific conditions and patient management altered based on an incorrectly assigned result
Z-2578-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Feb 22, 2022Class IIOpen, Classified
Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe
Z-1014-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Use of Device Problem2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Abrasion1
- Needle Stick/Puncture1
- Therapeutic Response, Increased1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryNov 6, 2024Use of Device Problem; Use of Device Problem
A COMPLAINT WAS FILED ON OCTOBER 25, 2024 REGARDING ACL TOP 550 CTS SN (B)(6) INDICATING USER ACCIDENTALLY PUNCTURED HIS FINGER ON PROBE WHILE TRYING TO UNLOCK IT. COMPLAINT MENTIONED THAT A COLLEAGUE PLACED A CLOSED TUBE IN THE WRONG RACK (WITHOUT A CAP) AND THE NEEDLE GOT STUCK, AND THE USER PRICKED HIS FINGER WHEN UNLATCHING THE PERFORATOR AND THE TUBE. THERE WAS NO KNOWN ADVERSE EFFECT FROM THE PUNCTURE.
Patient problems: Abrasion
MAUDE report 20622329 ↗InjuryJul 21, 2020Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT AN ACL TOP 550 CTS INSTRUMENT OPERATOR DEVELOPED A LACERATION RESULTING IN BLEEDING ON THE FINGER OF HER GLOVED HAND. THIS INCIDENT OCCURRED WHILE TROUBLESHOOTING TO CLEAN AND REMOVE CUVETTES FROM THE WASTE AREA OF THE INSTRUMENT. THE STAFF MEMBER RECEIVED MEDICAL ATTENTION IN THE EMERGENCY DEPARTMENT AT THE HOSPITAL AND FILED AN INCIDENT REPORT. NO STITCHES WERE NECESSARY. THE STAFF MEMBER WAS PLACED ON MEDICATION FOR (B)(6) PROPHYLAXIS.
Patient problems: Needle Stick/Puncture
MAUDE report 10303571 ↗InjuryMar 1, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A CUSTOMER REPORTED THAT THEIR ACL TOP 550 CTS GENERATED ERRONEOUS APTT RESULTS USING HEMOSIL SYNTHASIL. THREE PATIENTS WERE ADMINISTERED AN INCREASED DOSE OF HEPARIN BASED ON THE ERRONEOUS APTT RESULTS. THE APTT RESULTS WERE LATER CORRECTED AND THE HEPARIN DOSES WERE ADJUSTED AS APPROPRIATE. AFTER THE ADDITIONAL DOSE OF HEPARIN, THE PATIENTS WERE MONITORED AND THERE WERE NO COMPLICATIONS FROM THE ADDITIONAL HEPARIN ADMINISTERED.
Patient problems: Therapeutic Response, Increased
MAUDE report 8383931 ↗
Other INSTRUMENT, COAGULATION, AUTOMATED devices
- BCS® XP System 10459330Siemens Healthcare Diagnostics Products GmbH · INSTRUMENT, COAGULATION, AUTOMATED1 recall63 reports
- ACL TOP 350 CTS 00000280065Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED2 recalls1 report
- ACL TOP 500 CTS 00000280040Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall8 reports
- ACL TOP 750 LAS 00000280035Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall4 reports
- ACL TOP 750 CTS 00000280055Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall3 reports
- ACL TOP 300 CTS 00000280060Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall1 report
- ACL TOP 500 CTS 00000280040RInstrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall0 reports
- ACL TOP 300 CTS 00000280060RInstrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page