Skip to content

ACL TOP 550 CTS 00000280045Recalls & reported problems

Instrumentation Laboratory Company · ACL TOP 550 CTS · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08426950729242), received between Mar 1, 2019 and Nov 6, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 13, 2023Class IIOpen, Classified

    Sample misidentification could occur under specific conditions and patient management altered based on an incorrectly assigned result

    Z-2578-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Feb 22, 2022Class IIOpen, Classified

    Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

    Z-1014-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Use of Device Problem2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Abrasion1
  • Needle Stick/Puncture1
  • Therapeutic Response, Increased1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 6, 2024Use of Device Problem; Use of Device Problem

    A COMPLAINT WAS FILED ON OCTOBER 25, 2024 REGARDING ACL TOP 550 CTS SN (B)(6) INDICATING USER ACCIDENTALLY PUNCTURED HIS FINGER ON PROBE WHILE TRYING TO UNLOCK IT. COMPLAINT MENTIONED THAT A COLLEAGUE PLACED A CLOSED TUBE IN THE WRONG RACK (WITHOUT A CAP) AND THE NEEDLE GOT STUCK, AND THE USER PRICKED HIS FINGER WHEN UNLATCHING THE PERFORATOR AND THE TUBE. THERE WAS NO KNOWN ADVERSE EFFECT FROM THE PUNCTURE.

    Patient problems: Abrasion

    MAUDE report 20622329 ↗
  • InjuryJul 21, 2020Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT AN ACL TOP 550 CTS INSTRUMENT OPERATOR DEVELOPED A LACERATION RESULTING IN BLEEDING ON THE FINGER OF HER GLOVED HAND. THIS INCIDENT OCCURRED WHILE TROUBLESHOOTING TO CLEAN AND REMOVE CUVETTES FROM THE WASTE AREA OF THE INSTRUMENT. THE STAFF MEMBER RECEIVED MEDICAL ATTENTION IN THE EMERGENCY DEPARTMENT AT THE HOSPITAL AND FILED AN INCIDENT REPORT. NO STITCHES WERE NECESSARY. THE STAFF MEMBER WAS PLACED ON MEDICATION FOR (B)(6) PROPHYLAXIS.

    Patient problems: Needle Stick/Puncture

    MAUDE report 10303571 ↗
  • InjuryMar 1, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CUSTOMER REPORTED THAT THEIR ACL TOP 550 CTS GENERATED ERRONEOUS APTT RESULTS USING HEMOSIL SYNTHASIL. THREE PATIENTS WERE ADMINISTERED AN INCREASED DOSE OF HEPARIN BASED ON THE ERRONEOUS APTT RESULTS. THE APTT RESULTS WERE LATER CORRECTED AND THE HEPARIN DOSES WERE ADJUSTED AS APPROPRIATE. AFTER THE ADDITIONAL DOSE OF HEPARIN, THE PATIENTS WERE MONITORED AND THERE WERE NO COMPLICATIONS FROM THE ADDITIONAL HEPARIN ADMINISTERED.

    Patient problems: Therapeutic Response, Increased

    MAUDE report 8383931 ↗

Other INSTRUMENT, COAGULATION, AUTOMATED devices

All INSTRUMENT, COAGULATION, AUTOMATED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page