ACL TOP 750 LAS 00000280035Recalls & reported problems
Instrumentation Laboratory Company · ACL TOP 750 LAS · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08426950784050), received between Oct 3, 2022 and Nov 22, 2024 — classified as 4 injuries.
The most frequently reported specific device problem is “Data Problem” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 13, 2023Class IIOpen, Classified
Sample misidentification could occur under specific conditions and patient management altered based on an incorrectly assigned result
Z-2578-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Data Problem2
- Use of Device Problem2
Patient problems
- Abrasion2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryNov 22, 2024Use of Device Problem; Use of Device Problem
A COMPLAINT WAS REPORTED ON (B)(6)2024 REGARDING ACL TOP 750 LAS SN (B)(6), INDICATING THAT THE PROBE SCRAPED ACROSS FINGERTIPS OF THE TECHNOLOGIST WHEN SHE REACHED TOO FAR IN PAST CONVEYOR BELT TRYING TO ALIGN CUVETTES ON ACL TOP 750 LAS SN (B)(6). IT WAS NOTED THAT THE GLOVE THE TECHNICIAN WAS WEARING WAS STILL INTACT BUT A SMALL SCRATCH WAS SEEN ON FINGERTIP. TECHNICIAN WENT TO EMERGENCY ROOM AND FOLLOWED UP WITH OCCUPATIONAL HEALTH. OCCUPATIONAL HEALTH PLACED HER ON A PEP FOR POTENTIAL EXPOSURE FOR 28 DAYS AND SCHEDULED FOLLOW-UP. THERE WAS NO KNOWN ADVERSE EFFECT FROM THE SCRAPE.
Patient problems: Abrasion
MAUDE report 20762833 ↗InjuryNov 21, 2024Use of Device Problem; Use of Device Problem
A COMPLAINT FROM JAPAN WAS REPORTED ON 11/11/2024 REGARDING ASSY, SYSTEM, ACL TOP 750 LAS (PART NUMBER 00000280035) THAT MENTIONS THAT A STAFF MEMBER WAS INJURED BY A PART IN THE BACK OF THE CUVETTE WASTE BOX FROM INSTRUMENT ACL TOP 750 LAS SN (B)(6). WHEN SHE REMOVED THE BOX FROM THE INSTRUMENT AND PUT HER HAND ON IT, SHE CUT HER HAND ON THE METAL PART OF THE PLATE SPRING ON THE BACK OF THE BOX. IT WAS NOTED THAT BECAUSE OF THE RISK OF INFECTION, A STICKING PLASTERS WAS PUT ON WOUND SITE AND INJURED STAFF MEMBER WILL BE ALSO TAKING POST EXPOSURE DRUGS TO PREVENT AGAINST HIV VIRUS FOR 1 MONTH.
Patient problems: Abrasion
MAUDE report 20745498 ↗InjuryOct 11, 2022Data Problem; Data Problem
IT WAS REPORTED THAT THE ACL TOP 750 LAS ANALYZER INCORRECTLY READ A PATIENT SAMPLE BAR CODE ON (B)(6) 2022. THE OPERATOR WAS REVIEWING OUTSTANDING REQUESTS ON THE LIS AND NO RESULTS WERE PRESENT FOR SAMPLE NUMBER (B)(6) ON LIS. THE OPERATOR REMOVED THE PHYSICAL TUBE AND VIEWED THE BARCODE LABEL, THE SAMPLE THAT HAD READ AS (B)(6) ON THE ACL TOP SAMPLE LIST WAS ACTUALLY EYE-READABLE AS (B)(6). ON CHECKING THE GENERAL LOG FOR THE PERIOD ON TIME INVOLVED, THERE WAS NO NON-IDENTIFIED SAMPLE ERROR RECORDED DURING THIS TIME. SAMPLE ID (B)(6) HAD NO DEMOGRAPHIC UPLOADED TO ACL TOP FROM HEMOHUB. THE OPERATOR REPRINTED THE BAR CODES FOR EACH SAMPLE NUMBER AND LOADED ON THE ACL TOP. BOTH SAMPLE BARCODES READ CORRECTLY. NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15578943 ↗InjuryOct 3, 2022Data Problem; Data Problem
IT WAS REPORTED THAT THE ACL TOP 750 LAS ANALYZER INCORRECTLY READ A PATIENT SAMPLE BAR CODE ON (B)(6) 2022, AND POPULATED THE TEST RESULTS TO THE INCORRECT PATIENT IN THE HOSPITAL INFORMATION SYSTEM (HIS). THE ERROR WAS NOTICED BEFORE CLINICIANS HAD SEEN/ACTED UPON THE INCORRECT RESULTS. THE LABELS IN QUESTION WERE REPRINTED ON (B)(6) 2022 AND WERE READ CORRECTLY BY THE ACL TOP 750 LAS ANALYZER. THERE WAS NO PATIENT IMPACT AS RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15527228 ↗
Other INSTRUMENT, COAGULATION, AUTOMATED devices
- BCS® XP System 10459330Siemens Healthcare Diagnostics Products GmbH · INSTRUMENT, COAGULATION, AUTOMATED1 recall63 reports
- ACL TOP 550 CTS 00000280045Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED2 recalls3 reports
- ACL TOP 350 CTS 00000280065Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED2 recalls1 report
- ACL TOP 500 CTS 00000280040Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall8 reports
- ACL TOP 750 CTS 00000280055Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall3 reports
- ACL TOP 300 CTS 00000280060Instrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall1 report
- ACL TOP 500 CTS 00000280040RInstrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall0 reports
- ACL TOP 300 CTS 00000280060RInstrumentation Laboratory Company · INSTRUMENT, COAGULATION, AUTOMATED1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page