ACL TOP Family 50 Series Models as follows: In vitro… recallZ-2578-2023 · Instrumentation Laboratory · initiated Jul 13, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Instrumentation Laboratory or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sample misidentification could occur under specific conditions and patient management altered based on an incorrectly assigned result

FDA root cause category: Under Investigation by firm

What the company is doing

UPDATE: 10/16/23: Update relates to additional ACL TOP Family 50 Series instruments that were installed after the field notification was initiated (Phase I), but prior to inclusion of the notification instructions in start-up kits with new production. During the Installation process, Service field representatives will provide the titled Important Instruction, to the customer. The letter states the same instructions as provided in the Urgent Medical Device Correction Letter initially issued. __________________________________________________________________________ Werfen Issued Urgent Medical Device Correction Letter and Mandatory response tracking form on July 13th, 2023 via email or Certified Mail. Letter states reason for recall, health risk and action to take: Based on the above please take one of the following immediate workflows to mitigate this issue. To avoid the potential sample misidentification, users should not change any samples in the sample rack that have been properly identified. If the user decides they must remove a sample from a sample rack that has been properly identified by the ACL TOP Family 50 Series Instrument, they may do so only after the bar code reader has returned to the home position or the samples have been completely resulted. Alternatively, a user may remove a properly identified sample from a sample rack and must leave that sample rack position empty. x Share this information with your laboratory staff and update your internal procedures, as needed. x Forward this notification to all affected locations in your facility. x Post this notification on all affected systems in your facility. x Retain a copy of this notification for your records. x Complete and return the Mandatory Response Tracking Form to acknowledge the above actions. x Mandatory Response Tracking Record x Return the completed and signed Mandatory Response Tracking Form to: x Fax No.: 1-781-861-4207 Or E-mail: [email protected] x Contact Information for

Recall details

Recall number
Z-2578-2023
Status
Open, Classified
Initiated by the company
Jul 13, 2023
Posted by the FDA
Sep 12, 2023
Recalling firm
Instrumentation Laboratory
Quantity
2,068 kits US: 6,270 units OUS
Product code
GKP
FDA submissions
K150877
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Finland, Ireland, Lithuania, Malta, Qatar, Sweden.

Product description

ACL TOP Family 50 Series Models as follows: In vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis (1) ACL TOP 550 CTS Model Number: 00000280045; (2) ACL TOP 750 CTS Model Number: 00000280055; (3) ACL TOP 350 CTS Model Number: 00000280065; (4) ACL TOP 750 Model Number: 00000280015; (5) ACL TOP 750 LAS Model Number: 00000280035 (6) ACL TOP 970 CL(only available ROW)Model Number: 00000280097

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Model Names UDI (1)ACL TOP 550 CTS 08426950729242; (2)ACL TOP 750 CTS 08426950784067; (3)ACL TOP 350 CTS 08426950784081; (4) ACL TOP 750 08426950784074; (5) ACL TOP 750 LAS 08426950784050 (6) ACL TOP 970 CL 08430793045476 Software: 1)ACL TOP 550 CTS v6.3.0 or later; (2)ACL TOP 750 CTS v6.3.0 or later; (3)ACL TOP 350 CTS v6.3.0 or later; (4) ACL TOP 750 v6.3.0 or later; (5) ACL TOP 750 LAS v6.3.0 or later; (6) ACL TOP 970 CL Version 1.1.0 (OUS). UPDATE: 10/16/23 Affected Serial Number Range Start SN End SN ACL TOP 350 CTS 21113058 23074369; ACL TOP 550 CTS 21112046 23072623; ACL TOP 750 22010516 23060671; ACL TOP 750 CTS 21110860 23071015; ACL TOP 750 LAS 21120879 23071077

Devices named in this recall

6 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page