Skip to content

First Choice EUHP75164Recalls & reported problems

Creagh Medical Limited · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08570590020235), received between Jun 20, 2023 and Jun 20, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Failure to Deflate” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Failure to Deflate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 20, 2023Failure to Deflate; Failure to Deflate

    DURING AN ANGIOPLASTY PROCEDURE, A BALLOON CATHETER WAS USED AND WOULD NOT DEFLATE. THE PATIENT WAS SENT TO THE HOSPITAL TO RESOLVE THE ISSUE. THE FISTULA WAS LOCATED IN THE PATIENT'S ARM IN THE SUBCLAVIAN VEIN. THE LEVEL OF VESSEL CALCIFICATION AND LESION LENGTH WERE UNKNOWN. THE DEGREE OF STENOSIS WAS APPROXIMATELY 50%. THE BALLOON WAS INFLATED TO 14PSI AND PREPPED PER IFU REQUIREMENTS. THE REPORTED ISSUE OCCURRED DURING INITIAL INFLATION. THE PATIENT WAS NOTED TO BE LARGER AND THE PHYSICIAN WAS UNABLE TO LOCATE THE EXACT AREA CONFIDENTLY WITH ULTRASOUND TO DEFLATE THE BALLOON. THE PHYSICIAN THOUGHT IT WOULD BE BETTER FOR THE PATIENT TO GET TO CARE WHERE THEY COULD RECEIVE A CT SCAN. THE BALLOON WAS RUPTURED AT THE HOSPITAL WITHOUT ISSUE AND THE PATIENT RECEIVED DIALYSIS VIA ACCESS. THE PATIENT WAS REPORTED TO BE DOING WELL. THE PATIENT THEN CAME BACK TO THE SITE ON (B)(6) 2023 FOR ADDITIONAL INTERVENTION ON THE ACCESS. THE SECOND PROCEDURE WAS COMPLETED WITHOUT ISSUE. THERE WAS NO FURTHER IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17165762 ↗

Other Catheter, Angioplasty, Peripheral, Transluminal devices

All Catheter, Angioplasty, Peripheral, Transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page