MaxForce™ M00567370Recalls & reported problems
Boston Scientific Corporation · Balloon Dilatation Catheter · Class 2
All models of MaxForce (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729097068), received between Dec 24, 2019 and Jul 1, 2021 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Material Puncture/Hole” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Puncture/Hole6
Patient problems
- No Consequences Or Impact To Patient5
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2021Material Puncture/Hole; Material Puncture/Hole
BOSTON SCIENTIFIC CORPORATION RECEIVED A MAXFORCE BILIARY DILATION BALLOON DEVICE WITH NO ASSOCIATED INFORMATION. THIS EVENT HAS BEEN DEEMED REPORTABLE BASED ON THE INVESTIGATION RESULTS OF BALLOON PINHOLE. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: Insufficient Information
MAUDE report 12099645 ↗MalfunctionDec 24, 2019Material Puncture/Hole; Material Puncture/Hole
NOTE: THIS REPORT PERTAINS TO ONE OF TWO MAXFORCE BILIARY DILATION BALLOONS USED DURING THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO MAXFORCE BILIARY DILATION BALLOONS WERE USED IN THE BILE DUCT DURING A BILE DUCT DILATION PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON WOULD NOT INFLATE. THE SAME ISSUE "OCURRED" IN THE SECOND MAXFORCE BILIARY DILATION BALLOON. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE. INVESTIGATION RESULTS REVEALED THAT THE BALLOON HAD A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9516362 ↗MalfunctionDec 24, 2019Material Puncture/Hole; Material Puncture/Hole
NOTE: THIS REPORT PERTAINS TO ONE OF TWO MAXFORCE BILIARY DILATION BALLOONS USED DURING THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO MAXFORCE BILIARY DILATION BALLOONS WERE USED IN THE BILE DUCT DURING A BILE DUCT DILATION PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON WOULD NOT INFLATE. THE SAME ISSUE "OCURRED" IN THE SECOND MAXFORCE BILIARY DILATION BALLOON. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE. INVESTIGATION RESULTS REVEALED THAT THE BALLOON HAD TWO PINHOLES; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9516352 ↗MalfunctionDec 24, 2019Material Puncture/Hole; Material Puncture/Hole
NOTE: THIS REPORT PERTAINS TO ONE OF TWO MAXFORCE BILIARY BALLOONS USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO MAXFORCE BILIARY DILATION BALLOONS WERE USED DURING A BILIARY DILATION PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, BOTH BALLOONS WOULD NOT INFLATE DUE TO SMALL HOLES. REPORTEDLY, THE DEVICES WERE REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9516301 ↗MalfunctionDec 24, 2019Material Puncture/Hole; Material Puncture/Hole
NOTE: THIS REPORT PERTAINS TO ONE OF TWO MAXFORCE BILIARY BALLOONS USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO MAXFORCE BILIARY DILATION BALLOONS WERE USED DURING A BILIARY DILATION PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, BOTH BALLOONS WOULD NOT INFLATE DUE TO SMALL HOLES. REPORTEDLY, THE DEVICES WERE REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9516281 ↗MalfunctionDec 24, 2019Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE HEPATIC PORTAL DURING A BILIARY STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON HAD A PINHOLE-LIKE HOLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE HAVE BEEN NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9515931 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.