MaxForce™ M00567380Recalls & reported problems
Boston Scientific Corporation · Balloon Dilatation Catheter · Class 2
All models of MaxForce (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 08714729097075), received between Jan 3, 2019 and Jul 6, 2021 — classified as 19 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel9
- Material Puncture/Hole8
- Leak/Splash2
- Detachment of Device or Device Component1
- Difficult to Advance1
- Entrapment of Device1
- Material Deformation1
- Material Integrity Problem1
Patient problems
- No Consequences Or Impact To Patient17
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 6, 2021Entrapment of Device; Detachment of Device or Device Component; Difficult to Advance; Material Deformation; Entrapment of Device; Detachment of Device or Device Component; Difficult to Advance; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6), 2021. DURING THE PROCEDURE, THE DEVICE WAS ADVANCE OVER THE GUIDEWIRE; HOWEVER, THE DEVICE COULD NOT ADVANCE FARTHER THAN 30 CM. SUBSEQUENTLY, THE CATHETER BUNCHED UP IN ACCORDION AND THE GUIDEWIRE COULD NOT BE REMOVED EITHER. THE DEVICE WAS THEN SUCCESSFULLY REMOVED FROM THE SCOPE USING FORCE WHICH RIPPED OFF THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12125866 ↗MalfunctionFeb 15, 2021Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON WOULD NOT INFLATE AND WAS LEAKING. REPORTEDLY, THE DEVICE WAS REMOVED FROM THE BILE DUCT TO INSPECT ENDOSCOPICALLY, AND IT WAS FOUND THAT THE BALLOON HAD AT LEAST 3 SMALL HOLE LEAKS. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11327901 ↗MalfunctionOct 1, 2020Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
NOTE: THIS REPORT PERTAINS TO ONE OF TWO MAXFORCE BILIARY DILATATION BALLOONS AND A CRE PRO WIREGIDED DILATATION BALLOON USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO MAXFORCE BILIARY DILATATION BALLOONS AND A CRE PRO WIREGIDED DILATATION BALLOON WERE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTICED THAT ALL THE THREE BALLOONS WERE LEAKING. REPORTEDLY, THE CRE BALLOON WAS TESTED PRIOR TO USE INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PRO WIREGIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THAT THE BALLOON HAD A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10614371 ↗MalfunctionOct 1, 2020Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
NOTE: THIS REPORT PERTAINS TO ONE OF TWO MAXFORCE BILIARY DILATATION BALLOONS AND A CRE PRO WIREGIDED DILATATION BALLOON USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO MAXFORCE BILIARY DILATATION BALLOONS AND A CRE PRO WIREGIDED DILATATION BALLOON WERE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTICED THAT ALL THE THREE BALLOONS WERE LEAKING. REPORTEDLY, THE CRE BALLOON WAS TESTED PRIOR TO USE INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PRO WIREGIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THAT THE BALLOON HAD PINHOLES; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10614368 ↗MalfunctionAug 12, 2020Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE BILE DUCT DURING A DILATION PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON BECOME BROKEN AND WOULD NOT INFLATE AGAIN. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10398818 ↗MalfunctionAug 12, 2020Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE BILE DUCT DURING A DILATION PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON BECOME BROKEN AND WOULD NOT INFLATE AGAIN. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10398780 ↗MalfunctionAug 12, 2020Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE BILE DUCT DURING A DILATION PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON BECOME BROKEN AND WOULD NOT INFLATE AGAIN. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10398776 ↗MalfunctionAug 12, 2020Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE BILE DUCT DURING A DILATION PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON BECOME BROKEN AND WOULD NOT INFLATE AGAIN. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10398770 ↗MalfunctionJun 30, 2020Burst Container or Vessel; Burst Container or Vessel
THIS REPORT PERTAINS TO ONE OF TWO MAXFORCE BILIARY DILATATION BALLOONS USED DURING THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO MAXFORCE BILIARY DILATATION BALLOONS WERE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6), 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON BURST AND BROKE. THE SAME ISSUE OCCURRED WITH THE SECOND MAXFORCE BILIARY DILATATION BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10214636 ↗MalfunctionJun 30, 2020Burst Container or Vessel; Burst Container or Vessel
THIS REPORT PERTAINS TO ONE OF TWO MAXFORCE BILIARY DILATATION BALLOONS USED DURING THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO MAXFORCE BILIARY DILATATION BALLOONS WERE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6), 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON BURST AND BROKE. THE SAME ISSUE OCCURRED WITH THE SECOND MAXFORCE BILIARY DILATATION BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10214592 ↗MalfunctionJun 29, 2020Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE PANCREAS DURING A DILATION FOR A CYST DRAINAGE PLACEMENT PROCEDURE PERFORMED ON (B)(6), 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON WOULD NOT INFLATE. THE DEVICE WAS THEN REMOVED FROM THE PATIENT FOR INSPECTION AND IT WAS NOTICED THAT THE BALLOON HAD A HOLE. REPORTEDLY, THE STOMACH WAS FILLED WITH PANCREATIC FLUID, WHICH DID NOT LEAD TO ANY FURTHER COMPLICATIONS. THE PROCEDURE WAS COMPLETED WITH A NON-BSC DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE. INVESTIGATION RESULTS REVEALED THAT THE BALLOON HAD A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10208436 ↗MalfunctionApr 8, 2020Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, A LEAK IN THE DISTAL END OF THE BALLOON WAS NOTICED. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. INVESTIGATION RESULTS REVEALED THAT THE BALLOON HAD A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9934447 ↗MalfunctionMar 25, 2020Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, A LEAK WAS NOTED ON THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. INVESTIGATION RESULTS REVEALED THAT THE BALLOON WAS RUPTURED; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9881072 ↗MalfunctionFeb 27, 2020Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON DEVICE WAS USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, AN ATTEMPT WAS MADE TO INFLATE THE BALLOON; HOWEVER, THE BALLOON WAS NOTED TO BE LEAKING. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE. INVESTIGATION RESULTS REVEALED THAT THE BALLOON FOUND A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9763349 ↗MalfunctionFeb 27, 2020Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON WAS LEAKING, AS IT WAS DAMAGED. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE HAVE BEEN NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9762690 ↗MalfunctionOct 28, 2019Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, A LEAK WAS NOTED ON THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. INVESTIGATION RESULTS REVEALED THAT THE BALLOON HAD A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9246703 ↗MalfunctionSep 23, 2019Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, A LEAK WAS NOTED AT THE DISTAL END OF THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. INVESTIGATION RESULTS REVEALED THAT THE BALLOON HAD A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9103674 ↗MalfunctionJul 22, 2019Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED DURING A BILE DUCT DILATION PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON WOULD NOT INFLATE. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BILIARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THAT THE BALLOON WAS RUPTURED; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8812251 ↗MalfunctionJan 3, 2019Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BILIARY DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON POPPED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8213777 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page