Skip to content

Flexima™ Biliary M00539250Recalls & reported problems

Boston Scientific Corporation · Biliary Stent with Delivery System · Class 2 · Implant

All models of Flexima Biliary (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 08714729162568), received between Nov 4, 2019 and Apr 27, 2026 — classified as 2 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Break” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break11
  • Positioning Failure5
  • Adverse Event Without Identified Device or Use Problem2
  • Material Twisted/Bent2
  • Use of Device Problem2
  • Material Deformation1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 27, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS USED IN THE COMMON BILE DUCT TO TREAT STONES DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING DEPLOYMENT PROCESS, THE INNER GUIDE CATHETER GOT DETACHED. IT WAS REMOVED WITH A SNARE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULTS OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25009164 ↗
  • MalfunctionMar 3, 2026Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS IMPLANTED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING PROCEDURE, DIFFICULTIES WERE ENCOUNTERED DURING THE DEPLOYMENT PHASE. IMMEDIATELY FOLLOWING THE FINAL POSITIONING OF THE STENT, THE COMPONENTS OF THE DEVICE SEPARATED AND FELL APART. THE PROCEDURE WAS COMPLETED USING THE ORIGINAL DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24500605 ↗
  • MalfunctionJul 31, 2025Break; Positioning Failure; Use of Device Problem; Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem; Break; Positioning Failure; Use of Device Problem; Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED FOR BILIARY DRAINAGE AFTER A CHOLEDOCHOLITHOTOMY, DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. PROVIDED PHOTOGRAPHS OF THE DEVICE SHOWED WHAT APPEARED TO BE A TWISTED SUTURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: IT WAS REPORTED THAT THE GUIDEWIRE WAS IN THE PATIENT DURING THE ATTEMPTED DEPLOYMENT. PER THE IFU, "IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE DELIVERY SYSTEM, THE STENT CANNOT BE FULLY DEPLOYED." NOTE: THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF GUIDE CATHETER DETACHED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22682796 ↗
  • MalfunctionMar 18, 2025Positioning Failure; Use of Device Problem; Material Split, Cut or Torn; Positioning Failure; Use of Device Problem; Material Split, Cut or Torn

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY PRELOADED STENT WAS IMPLANTED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE FLEXIMA BILIARY STENT COULD NOT BE DEPLOYED. THE PROCEDURE WAS COMPETED WITH ANOTHER OF THE SAME DEVICE. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. IT WAS REPORTED THAT THE GUIDEWIRE WAS IN THE PATIENT'S BILE DUCT DURING THE ATTEMPTED DEPLOYMENT. HOWEVER, THE FLEXIMA BILIARY PRELOADED STENT INSTRUCTIONS FOR USE (IFU) STATES, "FOR PRELOADED SYSTEM ONLY: IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE ENDOSCOPE, THE STENT CANNOT BE FULLY DEPLOYED." THE PHYSICIAN DID NOT FOLLOW THE STEPS CITED IN THE IFU. NOTE: THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS WHICH REVEALED THAT THE STENT BARB WAS TORN. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21637118 ↗
  • MalfunctionJun 14, 2024Break; Positioning Failure; Break; Positioning Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY PRELOADED STENT WAS USED IN THE BILE DUCT TO TREAT A CHOLEDOCHOLITHIASIS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE GUIDE CATHETER BROKE WHEN THE STENT WAS RELEASED. THE PROCEDURE WAS COMPLETED WITH A NON-BOSTON SCIENTIFIC DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19536628 ↗
  • MalfunctionDec 12, 2023Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED TO TREAT STONES IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE INNER CORE OF THE DEVICE WAS FRACTURED, AND THE STENT COULD NOT BE DEPLOYED. THE PROCEDURE WAS COMPLETED USING ANOTHER FLEXIMA BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18314398 ↗
  • InjuryDec 11, 2023Break; Positioning Failure; Break; Positioning Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY PRELOADED STENT WAS USED FOR THE TREATMENT OF BILIARY STONES DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE INNER CATHETER TORE OFF WHEN THE STENT WAS DEPLOYED WHICH CAUSED DIFFICULTY IN RELEASING THE STENT. THE GUIDEWIRE WAS REMOVED; HOWEVER, THE INNER CATHETER REMAINED IN THE STENT INSIDE THE PATIENT. SUBSEQUENTLY, THE STENT ALONG WITH THE INNER CATHETER WAS RETRIEVED AND A SECOND FLEXIMA BILIARY STENT WAS USED. HOWEVER, IT WAS NOTED THAT THE STENT WAS UNABLE TO DEPLOY. A THIRD FLEXIMA BILIARY STENT WAS USED AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18305204 ↗
  • MalfunctionFeb 7, 2023Break; Positioning Failure; Break; Positioning Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE APPENDICITIS THERAPY (ERAT) PROCEDURE IN THE APPENDIX ON (B)(6) 2022. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED. ANOTHER FLEXIMA BILIARY STENT WAS USED AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF GUIDE CATHETER DETACHED/SEPARATED. PLEASE SEE BLOCK H10 FOR FULL INVESTIGATION DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16311761 ↗
  • MalfunctionSep 12, 2022Break; Material Deformation; Material Twisted/Bent; Break; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE USED DURING A BILE DUCT STENTING PROCEDURE IN THE BILE DUCT PERFORMED ON (B)(6) 2022. DURING UNPACKING, IT WAS FOUND THAT THE SUTURE WAS TWISTED. THE PROCEDURE WAS COMPLETED BUT IT WAS UNKNOWN WHAT COMPLETED THE PROCEDURE. A PHOTO OF THE COMPLAINT WAS PROVIDED AND SHOWED THAT THE SUTURE WAS TWISTED AND THE PUSH CATHETER WAS KINKED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THE GUIDE CATHETER WAS DETACHED/SEPARATED, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT. PLEASE SEE BLOCK H10 FOR FULL INVESTIGATION DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15407675 ↗
  • MalfunctionJun 15, 2021Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE COMMON BILE DUCT, PERFORMED ON (B)(6) 2021. DURING PREPARATION, WHEN PHYSICIAN ATTEMPTED TO RELEASE THE SUTURE, IT WAS NOTED THAT THE SUTURE WAS BROKEN. ANOTHER FLEXIMA BILIARY STENT WAS OPENED AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF GUIDE CATHETER BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11999062 ↗
  • MalfunctionJul 14, 2020Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE BILE DUCT, PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THEY ATTEMPTED TO RELEASE STENT FOR DEPLOYMENT, THE STENT WAS UNABLE TO BE DEPLOYED DUE TO BECOMING STUCK ON THE GUIDEWIRE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER FLEXIMA BILIARY STENT WITHOUT INCONVENIENCE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE OKAY. THIS COMPLAINT IS BEING REPORTED BASED ON THE INVESTIGATION RESULTS OF GUIDE CATHETER BREAK.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10273224 ↗
  • MalfunctionNov 4, 2019Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE IN THE COMMON BILE DUCT PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER FLEXIMA BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THAT THE GUIDE CATHETER WAS BROKEN, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9273492 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page