Flexima™ Biliary M00539260Recalls & reported problems
Boston Scientific Corporation · Biliary Stent with Delivery System · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Flexima biliary stent is implanted in the bile duct during ERCP. Reports mostly describe the stent or inner guide catheter fracturing or failing to deploy, with procedures completed using another device. It has 41 reports: no deaths, 3 injuries, and 38 malfunctions. There are no recalls, and reports fell from 10 to 4 in the recent year.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 41
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 08714729162575), received between Jan 3, 2017 and Jun 10, 2026 — classified as 3 injuries, 38 malfunctions.
The most frequently reported specific device problem is “Break” (36 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break36
- Positioning Failure21
- Adverse Event Without Identified Device or Use Problem11
- Use of Device Problem10
- Material Deformation8
- Stretched4
- Entrapment of Device3
- Defective Device2
Patient problems
- No Clinical Signs, Symptoms or Conditions35
- No Consequences Or Impact To Patient6
Recent adverse event reports
The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2026Break; Positioning Failure
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS INTENDED TO BE IMPLANTED IN THE COMMON BILE DUCT TO TREAT A STENOSIS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE WITH STENT PLACEMENT, PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE STENT WAS FRACTURED AND COULD NOT BE DEPLOYED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULTS OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25460994 ↗MalfunctionApr 24, 2026Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS ATTEMPTED TO BE IMPLANTED IN THE COMMON BILE DUCT TO TREAT STONES DURING AN ENDOSCOPIC ULTRASOUND (EUS)-GUIDED BILIARY DRAINAGE PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE STENT FRACTURED INTO MULTIPLE PIECES WHILE BEING ADVANCED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULTS OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24987476 ↗MalfunctionJan 12, 2026Break; Nonstandard Device; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE DEVICE WAS PRESENTING A DEFECT DURING USE, MAKING IT IMPOSSIBLE TO USE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS WHICH REVEALED THAT THE GUIDE CATHETER WAS DETACHED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24039857 ↗MalfunctionDec 17, 2025Break; Positioning Failure; Adverse Event Without Identified Device or Use Problem; Break; Positioning Failure; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING PREPARATION, THE DEVICE COULD NOT BE DEPLOYED, AND THE CATHETER GUIDE WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23829802 ↗MalfunctionOct 29, 2025Break; Positioning Failure; Adverse Event Without Identified Device or Use Problem; Break; Positioning Failure; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED TRANSDUODENAL TO THE BILIARY TRACT TO TREAT A BILIARY STENOSIS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE DEVICE COULD NOT BE DEPLOYED NORMALLY. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: THIS EVENT HAS BEEN DEEMED REPORTABLE BASED ON THE INVESTIGATION FINDING THAT THE GUIDE CATHETER DETACHED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23418732 ↗MalfunctionJul 29, 2025Break; Positioning Failure; Adverse Event Without Identified Device or Use Problem; Break; Positioning Failure; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED IN THE GALLBLADDER TO TREAT GALLLADDER STONES DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE (ERCP) PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, IT WAS NOTED THAT THE SUTURE OF THE FLEXIMA BILIARY STENT WAS FRACTURED, AND THE STENT COULD NOT BE OPENED NORMALLY. ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION AFTER THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THIS COMPLAINT HAS BEEN DEEMED MDR REPORTABLE EVENT BASED ON THE INVESTIGATION RESULT WHICH REVEALED THAT GUIDE CATHETER WAS DETACHED. PLEASE SEE BLOCK H11 FOR THE FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22659252 ↗InjuryJun 11, 2025Break; Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Break; Use of Device Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED TO THE BILE TRACT TO TREAT A STENOSIS CAUSED BY MULTIPLE BILIARY CALCULI DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING PROCEDURE, IT WAS REPORTED THAT THE INNER SHEATH CORE COULD NOT BE PULLED OUT AND IT WAS FRACTURED. THE BROKEN GUIDE CATHETER HAS BEEN REMOVED USING SNARE AND FOREIGN OBJECT FORCEPS. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT THE GUIDEWIRE WAS INSIDE THE PATIENT DURING THE ATTEMPTED DEPLOYMENT. HOWEVER, THE FLEXIMA BILIARY STENT WITH DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU) STATES, "TO DEPLOY THE STENT, LOOSEN THE TUOHY-BORST ADAPTER SLIGHTLY AND PULL ON THE GUIDE CATHETER LUER-LOK HUB AND GUIDEWIRE. HOLD THE OUTER POSITIONER STATIONARY, WHILE GENTLY RETRACTING THE INNER GUIDE CATHETER AND GUIDEWIRE. ENDOSCOPICALLY MONITOR THE STENT POSITION. WARNING: IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE DELIVERY SYSTEM, THE STENT CANNOT BE FULLY DEPLOYED. THE PHYSICIAN DID NOT FOLLOW THE STE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22185361 ↗MalfunctionJun 11, 2025Break; Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Break; Use of Device Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED TO THE BILIARY TRACT TO TREAT A MULTIPLE BILIARY CALCULI DURING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE INNER SHEATH CORE COULD NOT BE PULLED OUT AND IT WAS FRACTURED. IT WAS NOTED THAT THE BILIARY DUCT WAS NARROW. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE CONDITION OF THE PATIENT WAS REPORTED TO BE STABLE. NOTE: IT WAS REPORTED THAT THE GUIDEWIRE WAS IN THE PATIENT DURING THE ATTEMPTED DEPLOYMENT. PER THE IFU, "IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE DELIVERY SYSTEM, THE STENT CANNOT BE FULLY DEPLOYED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22185037 ↗MalfunctionApr 9, 2025Break; Positioning Failure; Premature Activation; Adverse Event Without Identified Device or Use Problem; Break; Positioning Failure; Premature Activation; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED TO TREAT GALLSTONES IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, AFTER THE STONES WERE REMOVED, THE STENT WAS PLACED BUT GREAT RESISTANCE WAS FELT, AND THE STENT COULD NOT BE DEPLOYED SUCCESSFULLY. THE DEVICE WAS THEN WITHDRAWN, BUT IT WAS NOTICED THAT THE STET FELL OFF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION WAS STABLE. NOTE: THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS WHICH REVEALED THAT THE GUIDE CATHETER WAS DETACHED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21809753 ↗MalfunctionFeb 20, 2025Break; Positioning Failure; Use of Device Problem; Break; Positioning Failure; Use of Device Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED IN THE BILIARY TRACT FOR BILIARY DILATION DURING A BILIARY STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2025. THE PATIENT'S ANATOMY WAS SEVERELY TORTUOUS. DURING THE PROCEDURE, THE PULL WIRE BROKE AND THE FLEXIMA STENT WAS UNABLE TO BE DEPLOYED. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION AFTER THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: IT WAS REPORTED THAT THE BARB FLAP COVER WAS NOT USED TO INSERT THE DEVICE THROUGH THE BIOPSY CAP. HOWEVER, THE FLEXIMA BILIARY STENT WITH DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU) STATES, "SLIDE THE OUTER ORANGE SLEEVE OVER THE MOST PROXIMAL STENT BARB TO HOLD IT DOWN FOR STENT INSERTION." THE PHYSICIAN DID NOT FOLLOW THE STEPS CITED IN THE IFU.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21425848 ↗MalfunctionFeb 11, 2025Break; Positioning Failure; Adverse Event Without Identified Device or Use Problem; Break; Positioning Failure; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED IN THE BILE DUCT TO TREAT BILE DUCT STONES DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE (ERCP) PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE INNER GUIDE CATHETER WAS NOT EASY TO BE PULLED OUT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION AFTER THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THIS COMPLAINT HAS BEEN DEEMED MDR REPORTABLE EVENT BASED ON THE INVESTIGATION RESULT WHICH REVEALED THAT GUIDE CATHETER WAS DETACHED. PLEASE SEE BLOCK H11 FOR THE FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21357211 ↗InjuryJan 28, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY PRELOADED STENT WAS TO BE USED IN THE BILIARY TRACT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE INNER SHEATH WAS FRACTURED DURING THE RELEASE. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION REMAINED STABLE AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21260242 ↗MalfunctionDec 2, 2024Break; Difficult or Delayed Positioning; Material Deformation; Break; Difficult or Delayed Positioning; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED IN THE BILE DUCT TO TREAT BILE DUCT STONES DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE (ERCP) PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE INNER GUIDE CATHETER WAS FRACTURED AND WAS DIFFICULT TO PUSH OUT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION AFTER THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: A PHOTO PROVIDED BY THE COMPLAINANT SHOWED THAT THE GUIDE CATHETER WAS KINKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20820314 ↗MalfunctionNov 14, 2024Positioning Failure; Use of Device Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn; Positioning Failure; Use of Device Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED IN THE COMMON BILE DUCT TO TREAT A BILIARY STONE STENOSIS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE (ERCP) PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE STENT WAS STUCK AND COULD NOT BE DEPLOYED. ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION AFTER THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: IT WAS REPORTED THAT THE GUIDEWIRE WAS IN THE DELIVERY SYSTEM DURING THE ATTEMPTED DEPLOYMENT. HOWEVER, THE FLEXIMA BILIARY STENT WITH DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU) STATES, "WARNING IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE DELIVERY SYSTEM, THE STENT CAN NOT BE FULLY DEPLOYED." THE PHYSICIAN DID NOT FOLLOW THE STEPS CITED IN THE IFU. NOTE: THIS COMPLAINT HAS BEEN DEEMED MDR REPORTABLE EVENT BASED ON THE INVESTIGATION RESULT WHICH REVEALED THAT THE STENT BARB WAS TORN. PLEASE SEE BLOCK H11 FOR THE FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20682486 ↗MalfunctionOct 11, 2024Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY PRELOADED STENT WAS TO BE USED TO TREAT A CHOLEDOCHOLITHIASIS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT PLACEMENT PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, WHILE PULLING THE INNER GUIDE CATHETER TO DEPLOY THE STENT, THE INNER GUIDE CATHETER TORE. THERE WAS NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON OCTOBER 29, 2024. IT WAS REPORTED THAT THE FLEXIMA BILIARY PRELOADED STENT WAS TO BE USED IN THE COMMON BILE DUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20428012 ↗MalfunctionSep 11, 2024Break; Positioning Failure; Use of Device Problem; Break; Positioning Failure; Use of Device Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY PRELOADED STENT WAS USED IN THE BILE DUCT TO TREAT BILE DUCT STONES DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GRAY INNER SHEATH COULD NOT BE PULLED OUT AND THE STENT COULD NOT BE DEPLOYED. THE PROCEDURE WAS COMPLETED WITH ANOTHER FLEXIMA BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: IT WAS REPORTED THAT THE BARB FLAP COVER WAS NOT USED TO INSERT THE DEVICE THROUGH THE BIOPSY CAP. HOWEVER, THE FLEXIMA BILIARY STENT WITH DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU) STATES, "SLIDE THE OUTER ORANGE SLEEVE OVER THE MOST PROXIMAL STENT BARB TO HOLD IT DOWN FOR STENT INSERTION." THIS COMPLAINT HAS BEEN DEEMED MDR REPORTABLE BASED ON THE INVESTIGATION RESULT WHICH REVEALED THE CATHETER GUIDE WAS DETACHED. PLEASE SEE BLOCK H11 FOR THE FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20191735 ↗MalfunctionJun 14, 2024Positioning Failure; Use of Device Problem; Material Split, Cut or Torn; Positioning Failure; Use of Device Problem; Material Split, Cut or Torn
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED IN THE BILIARY TRACT DURING AN IMPLANTATION PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED. THE PROCEDURE WAS COMPLETED USING ANOTHER FLEXIMA BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS WHICH REVEALED THAT THE STENT BARB WAS TORN. PLEASE SEE BLOCK H10 FOR FULL INVESTIGATION DETAILS. IT WAS REPORTED THAT THE GUIDEWIRE WAS INSIDE THE PATIENT DURING THE ATTEMPTED DEPLOYMENT. HOWEVER, THE FLEXIMA BILIARY STENT WITH DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU) STATES, "COMPLETELY RETRACT THE GUIDEWIRE INTO THE ENDOSCOPE. IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE DELIVERY SYSTEM, THE STENT CANNOT BE FULLY DEPLOYED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19537391 ↗MalfunctionApr 11, 2024Break; Positioning Failure; Stretched; Use of Device Problem; Defective Device; Activation, Positioning or Separation Problem; Material Deformation; Break; Positioning Failure; Stretched; Use of Device Problem; Defective Device; Activation, Positioning or Separation Problem; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY PRELOADED STENT WAS TO BE IMPLANTED IN THE COMMON BILE DUCT TO TREAT STONES DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED AS THE BLACK WIRE BETWEEN THE OUTER SHEATH AND THE STENT WAS NOT FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER FLEXIMA BILIARY PRELOADED STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THIS EVENT HAS BEEN DEEMED AN MDR-REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS WHICH REVEALED THAT THE GUIDE CATHETER WAS DETACHED FROM THE DELIVERY SYSTEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19093698 ↗MalfunctionMar 15, 2024Break; Positioning Failure; Use of Device Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem; Break; Positioning Failure; Use of Device Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY PRELOADED STENT WAS TO BE IMPLANTED IN THE BILIARY TRACT FOR STENT PLACEMENT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE TOUHY BORST WAS BROKEN, AND THE STENT COULD NOT BE DEPLOYED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT, AND THE PATIENT'S CONDITION AFTER THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: IT WAS REPORTED THAT THE GUIDEWIRE WAS INSIDE THE PATIENT DURING THE ATTEMPTED DEPLOYMENT. HOWEVER, THE FLEXIMA BILIARY STENT WITH DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU) STATES, "TO DEPLOY THE STENT, LOOSEN THE TUOHY-BORST ADAPTER SLIGHTLY AND PULL ON THE GUIDE CATHETER LUER-LOK HUB AND GUIDEWIRE. HOLD THE OUTER POSITIONER STATIONARY, WHILE GENTLY RETRACTING THE INNER GUIDE CATHETER AND GUIDEWIRE. ENDOSCOPICALLY MONITOR THE STENT POSITION. WARNING: IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE DELIVERY SYSTEM, THE STENT CAN NOT BE FULLY DEPLOYED." THE PHYSICIAN DID NOT FOLLOW THE STEPS CITED IN THE IFU. THIS EVENT HAS BEEN DEEMED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18912919 ↗MalfunctionFeb 29, 2024Break; Positioning Failure; Use of Device Problem; Break; Positioning Failure; Use of Device Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY PRELOADED STENT WAS TO BE IMPLANTED IN THE BILIARY TRACT TO TREAT TIGHT BILIARY STENOSIS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER FLEXIMA BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT THE GUIDEWIRE WAS INSIDE THE PATIENT DURING THE ATTEMPTED DEPLOYMENT. HOWEVER, THE FLEXIMA BILIARY STENT WITH DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU) STATES, "TO DEPLOY THE STENT, LOOSEN THE TUOHY-BORST ADAPTER SLIGHTLY AND PULL ON THE GUIDE CATHETER LUER-LOK HUB AND GUIDEWIRE. HOLD THE OUTER POSITIONER STATIONARY, WHILE GENTLY RETRACTING THE INNER GUIDE CATHETER AND GUIDEWIRE. ENDOSCOPICALLY MONITOR THE STENT POSITION. WARNING: IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE DELIVERY SYSTEM, THE STENT CANNOT BE FULLY DEPLOYED." THIS EVENT HAS BEEN DEEMED AN MDR-REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF THE GUIDE CATHETER BREAK. PLEASE SEE BLOCK H10 FOR THE FULL INVESTIGA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18809745 ↗InjuryDec 21, 2023Break; Positioning Failure; Stretched; Use of Device Problem; Detachment of Device or Device Component; Material Deformation; Break; Positioning Failure; Stretched; Use of Device Problem; Detachment of Device or Device Component; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY PRELOADED STENT WAS TO BE IMPLANTED IN THE GALLBLADDER FOR BILIARY DRAINAGE DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED, THE GUIDE CATHETER WAS BROKEN, STRETCHED, AND KINKED, AND THE SUTURE WAS DETACHED. A BIOPSY FORCEPS WAS USED TO REMOVE THE BROKEN PIECES OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER FLEXIMA BILIARY PRELOADED STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT, AND THE PATIENTS CONDITION AFTER THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: IT WAS REPORTED THAT THE GUIDEWIRE WAS INSIDE THE PATIENT DURING THE ATTEMPTED DEPLOYMENT. HOWEVER, THE FLEXIMA BILIARY STENT WITH DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU) STATES, "COMPLETELY RETRACT THE GUIDEWIRE INTO THE ENDOSCOPE. IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE DELIVERY SYSTEM, THE STENT CANNOT BE FULLY DEPLOYED." THE PHYSICIAN DID NOT FOLLOW THE STEPS CITED IN THE IFU.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18372146 ↗MalfunctionJul 28, 2023Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A FLEXIMA BILIARY STENT WAS ATTEMPTED TO BE USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE COMMON BILE DUCT, PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE GUIDEWIRE WAS STUCK IN THE ADVANCER. ANOTHER FLEXIMA BILIARY STENT WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS; GUIDE CATHETER DETACHED/SEPARATED. PLEASE SEE BLOCK H10 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17414955 ↗MalfunctionMar 27, 2023Break; Difficult or Delayed Positioning; Entrapment of Device; Break; Difficult or Delayed Positioning; Entrapment of Device
IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A FLEXIMA BILIARY STENT WAS USED DURING A PROCEDURE. THE PROCEDURE DATE IS UNKNOWN. DURING THE PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT, GREAT RESISTANCE WAS FELT. THE PHYSICIAN REMOVED THE DEVICE FROM THE DUODENOSCOPE, REASSEMBLED, AND AFTER A FEW ATTEMPTS EXPERIENCING CONTINUED RESISTANCE DUE TO THE INABILITY TO REMOVE THE GUIDEWIRE, THE PHYSICIAN DEPLOYED THE STENT AND COMPLETED THE PROCEDURE. THE 0.025 THREAD GUIDEWIRE (NON-BOSTON SCIENTIFIC, MANUFACTURER IS UNKNOWN) USED DURING THE PROCEDURE WAS REMOVED FROM THE SCOPE, HOWEVER, COULD NOT BE REMOVED FROM THE DELIVERY CATHETER AFTER THE PROCEDURE. THERE WERE NO KNOWN PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF GUIDE CATHETER DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16620498 ↗MalfunctionMar 3, 2023Break; Positioning Failure; Break; Positioning Failure
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE IN THE BILIARY TRACT PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, IT WAS NOTED THAT THE STENT DID NOT DEPLOY SUCCESSFULLY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER FLEXIMA BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THE GUIDE CATHETER WAS DETACHED, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16479156 ↗MalfunctionDec 19, 2022Break; Positioning Failure; Break; Positioning Failure
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE BILIARY DRAINAGE (ERBD) PROCEDURE IN THE COMMON BILE DUCT PERFORMED ON (B)(6) 2022. A PATIENT WHO SUFFERED FROM COMMON BILE DUCT DILATATION COMBINED WITH STONES HAD A LARGE NUMBER OF STONES WERE REMOVED AND RE-ANGIOGRAPHY SHOWED NO OBVIOUS RESIDUAL STONES. DURING THE PROCEDURE, WHEN THE DEVICE WAS PLACED FOR DRAINAGE, THE STENT WAS UNABLE TO BE DEPLOYED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER FLEXIMA BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THE GUIDE CATHETER WAS DETACHED/SEPARATED, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16010070 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page