Flexima™ Biliary M00539320Recalls & reported problems

Boston Scientific Corporation · Biliary Stent with Delivery System · Class 2 · Implant

All models of Flexima Biliary (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729162636), received between Jan 17, 2017 and May 14, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component1
  • Entrapment of Device1
  • Kinked1
  • Material Deformation1
  • Obstruction of Flow1
  • Positioning Failure1
  • Premature Activation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient2
  • Foreign Body In Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2026Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED FOR DILATATION DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE STENT WAS IMPLANTED; HOWEVER, FOLLOWING PLACEMENT, IT WAS NOTED THAT THE RADIOPAQUE MARKER BAND HAD DETACHED AND REMAINED INSIDE THE PATIENT. THE MARKER BAND WAS LEFT IN SITU, AS THE PHYSICIAN DETERMINED THAT IT WOULD NOT IMPACT THE PATIENT. THE PROCEDURE WAS COMPLETED USING THE ORIGINAL STENT, AND NO ADDITIONAL SURGERY OR INTERVENTION WAS PERFORMED. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 25167982 ↗
  • MalfunctionMar 24, 2025Break; Positioning Failure; Entrapment of Device; Stretched; Break; Positioning Failure; Entrapment of Device; Stretched

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED IN THE BILE DUCT TO TREAT CHOLEDOCHOLITHIASIS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE (ERCP) PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE FLEXIMA BILIARY STENT DID NOT DEPLOY. THE DUODENOSCOPE WAS THEN REMOVED BECAUSE THE STENT CATHETER GOT STUCK IN THE WORKING CHANNEL. THE DELIVERY SYSTEM WAS CHECKED, AND IT WAS OBSERVED THAT THE GRAY STENT DELIVERY TUBE WAS NOTED TO BE STUCK AND STRETCHED, AND THE STENT ITSELF WAS NOTED TO BE DAMAGED. A DIFFERENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21681885 ↗
  • MalfunctionFeb 1, 2022Premature Activation; Material Deformation; Premature Activation; Material Deformation

    IT WAS REPORTED BOSTON SCIENTIFIC CORPORATION THAT AN FLEXIMA BILIARY STENT WAS OPENED TO BE USED FOR AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE WITH PLASTIC STENT PLACEMENT IN THE COMMON BILE DUCT PERFORMED ON (B)(6) 2021. DURING UNPACKING, IT WAS FOUND THAT THE FLEXIMA BILIARY STENT HAS BEEN DEPLOYED IN THE PACKAGING. ADDITIONALLY, IT WAS NOTED THAT THE FLEXIMA BILIARY STENT WAS DAMAGED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER FLEXIMA BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE AS NO PATIENT COMPLICATIONS. THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13418864 ↗
  • MalfunctionFeb 24, 2021Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE BILE DUCT, PERFORMED ON (B)(6) 2021 IT WAS REPORTED THAT, WHEN THE PHYSICIAN OPENED THE PACKAGE, THE STENT WAS BROKEN. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER FLEXIMA BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11376374 ↗
  • MalfunctionAug 5, 2019Obstruction of Flow; Obstruction of Flow

    NOTE: THIS REPORT IS ONE OF TWO COMPLAINTS THAT PERTAIN TO THE SAME EVENT (MFR REPORT # 3005099803-2019-03911 AND MFR. REPORT # 3005099803-2019-03910). IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO PREVIOUSLY IMPLANTED FLEXIMA BILIARY STENTS WERE REMOVED FROM A PATIENT DURING AN ENDOSCOPIC BILIARY DRAINAGE WITH STENT EXCHANGE PROCEDURE PERFORMED ON (B)(6) 2019. THE EXACT IMPLANT DATES FOR THE FLEXIMA STENTS WERE NOT REPORTED. THE PATIENT HAD A PREVIOUSLY PLACED A FLEXIMA BILIARY STENT 10FR/10CM IN B6, AND A FLEXIMA BILIARY STENT 10FR/15CM IN B8. BOTH STENTS WERE REPORTED TO BE OCCLUDED. THE PHYSICIAN REMOVED THE OCCLUDED STENTS USING A SNARE. THE PHYSICIAN ATTEMPTED TO PLACE A NEW FLEXIMA BILIARY STENT 10FR/15CM IN B8, HOWEVER THE DEVICE BECAME DIFFICULT TO ADVANCE IN THE PATIENT'S ANATOMY DUE TO SEVERE TORTUOSITY. THE DEVICE WAS PULLED OUT TO CHANGE THE GUIDEWIRE. THE GUIDEWIRE WAS REPLACED AND THE PHYSICIAN ATTEMPTED AGAIN TO PLACE THE FLEXIMA BILIARY STENT 10FR/15CM IN B8 HOWEVER THE STENT PREMATURELY DEPLOYED FROM THE DELIVERY SYSTEM. UPON REMOVAL AND EXAMINATION OF THE FLEXIMA BILIARY STENT 10FR/15CM, THE SUTURE WAS NOTED TO BE SEPARATED AND REMAINED INSIDE THE WOR

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8862167 ↗
  • MalfunctionJan 17, 2017Kinked; Kinked

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA¿ BILIARY STENT WAS USED IN AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2016. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTICED THAT THE STENT WAS SEPARATED FROM THE DELIVERY SYSTEM. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE "STABLE". THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS; STENT KINKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6255788 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →