Flexima™ Biliary M00539330Recalls & reported problems
Boston Scientific Corporation · Biliary Stent with Delivery System · Class 2 · Implant
All models of Flexima Biliary (15)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729162643), received between Nov 18, 2019 and Jul 8, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Adverse Event Without Identified Device or Use Problem1
- Premature Activation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 8, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS TO BE IMPLANTED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC BILIARY DRAINAGE (EBD) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE CONNECTOR AT THE PROXIMAL PART OF THE INNER SHEATH BECAME DISCONNECTED. THE BROKEN GUIDE CATHETER REMAINED WITHIN THE PUSH CATHETER, ALLOWING THE DELIVERY SYSTEM TO BE REMOVED IN ONE PIECE. THE PROCEDURE WAS COMPLETED USING THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22436718 ↗MalfunctionMay 26, 2022Break; Premature Activation; Break; Premature Activation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA BILIARY STENT WAS USED DURING A BILIARY DRAINAGE PROCEDURE IN THE BILE DUCT PERFORMED ON (B)(6) 2022. DURING THE PREPARATION, WHEN THE DEVICE WAS UNPACKED, IT WAS FOUND THAT THE DELIVERY SYSTEM WAS NOT CONNECTED TO THE STENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER FLEXIMA BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THAT THE GUIDE CATHETER WAS DETACHED/SEPARATED, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14515605 ↗MalfunctionNov 18, 2019Break; Break
NOTE: THIS REPORT IS ONE OF THREE COMPLAINTS THAT PERTAIN TO THE SAME EVENT (MFR REPORT # 3005099803-2019-05601, 3005099803-2019-05622 AND 3005099803-2019-05625). IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT THREE FLEXIMA BILIARY STENTS WERE USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE PERFORMED IN THE BILE DUCT ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, WHILE DEPLOYING THE FLEXIMA BILIARY STENT INSIDE THE PATIENT, THE GUIDE CATHETER STRETCHED. THE STENT WAS IMPLANTED IN THE PATIENT. A PHOTO OF THE DEVICE AFTER BEING REMOVED FROM THE PATIENT SHOWS THE GUIDE CATHETER WAS BROKEN. A SECOND FLEXIMA BILIARY STENT WAS BEING PLACED, BUT THE GUIDE CATHETER BROKE DURING ATTEMPTED DEPLOYMENT. THE STENT WAS IMPLANTED IN THE PATIENT. THE GUIDE CATHETER REMAINED WITHIN THE PUSH CATHETER ENABLING THE CATHETER TO BE REMOVED FROM THE PATIENT IN ONE PIECE. A THIRD FLEXIMA BILIARY STENT WAS ATTEMPTED TO BE PLACED, BUT THE GUIDE CATHETER BROKE OFF DURING ATTEMPTED DEPLOYMENT AND REMAINED WITHIN THE STENT. THE BROKEN GUIDE CATHETER AND STENT WERE REMOVED BY THE PHYSICIAN USING BIOPSY FORCEPS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9336491 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page