Symmetry™ M001103300Recalls & reported problems
Boston Scientific Corporation · Balloon Dilatation Catheter · Class 2
All models of Symmetry (13)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 08714729181859), received between Nov 26, 2018 and Jan 19, 2026 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture10
- Entrapment of Device3
Patient problems
- No Consequences Or Impact To Patient7
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 19, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND NON-CALCIFIED RADIAL ANTECUBITAL VEIN. A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE INITIAL INFLATION AT 10 ATMOSPHERES FOR ONE MINUTE, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24108173 ↗MalfunctionMar 27, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY CALCIFIED FOREARM SHUNT. A 4.5-4/4T/90 SYMMETRY BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE SECOND INFLATION AT ABOUT 8 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21706203 ↗MalfunctionMay 3, 2024Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT DEVICE STUCK ON GUIDEWIRE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED SHUNT VESSEL. AFTER A V-18 CONTROL WIRE WAS INSERTED, A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, IT WAS NOTED THAT THE WIRE WAS STUCK AROUND THE MARKER ON THE JOINT PART BETWEEN THE SHAFT AND THE BALLOON CATHETER. RESISTANCE WAS ENCOUNTERED UPON REMOVAL, BUT THE DEVICES WERE ABLE TO BE REMOVED AS ONE UNIT. NO FRAGMENTS LEFT INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19242706 ↗MalfunctionDec 6, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED FOREARM CEPHALIC VEIN. A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 16 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED AND VERTICALLY SEPARATED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES, REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15923589 ↗MalfunctionNov 16, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SHUNT VEIN. A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING FIFTH INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. THE PATIENT WAS IN GOOD CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15808503 ↗MalfunctionJul 15, 2022Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT CATHETER ENTRAPMENT ON GUIDEWIRE OCCURRED. A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS SELECTED FOR USE. HOWEVER, AFTER A NON-BOSTON SCIENTIFIC GUIDEWIRE WAS INSERTED INTO THE GUIDEWIRE LUMEN OF THE BALLOON, A RESISTANCE WAS FELT AT 40CM FROM THE HANDLE SIDE. IT WAS NOTED THAT THE BALLOON CATHETER WAS STUCK WITH THE GUIDEWIRE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15031030 ↗MalfunctionDec 13, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON PINHOLE WAS OBSERVED. A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS SELECTED FOR USE; HOWEVER, DURING PREPARATION, THE AIR IN THE BALLOON COULD NOT BE TOTALLY RELEASED. WHEN THE BALLOON WAS CHECKED, IT WAS FOUND TO HAVE A HOLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10999693 ↗MalfunctionAug 11, 2020Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT CATHETER ENTRAPMENT OCCURRED. A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS SELECTED FOR DILATION. HOWEVER, THE GUIDEWIRE WAS STUCK INSIDE THE SHAFT AROUND THE BALLOON PART AND THE CATHETER COULD NOT BE PLACED ON THE WIRE. ANOTHER 4.0-4/4T/90 SYMMETRY BALLOON CATHETER WAS SELECTED BUT THE BALLOON WAS TORN DURING PREPARATION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10390252 ↗MalfunctionJul 15, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 5 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED WITHOUT ANY ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10280484 ↗MalfunctionMar 14, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MODERATELY TORTUOUS AND MILDLY CALCIFIED VESSEL. A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, UPON FIRST INFLATION, LEAKAGE OF CONTRAST MEDIA WAS OBSERVED AT THE TIP PART AND DURING THE THIRD INFLATION AT 10 ATMOSPHERES FOR 1 MINUTE, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9832958 ↗MalfunctionFeb 20, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN THE MODERATELY TORTUOUS VESSEL IN THE FOREARM. A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING FIRST INFLATION AT 10 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED FROM THE PATIENT WITHOUT ANY ISSUE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9734325 ↗MalfunctionDec 18, 2018Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. VASCULAR ACCESS WAS OBTAINED USING VENOUS RETROGRADE APPROACH. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL IN THE LEFT FOREARM. AFTER A GUIDE WIRE WAS PASSED THROUGH, A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NON-BSC DEVICE. NO SERIOUS INJURY NOR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8173660 ↗MalfunctionNov 26, 2018Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN A SHUNT IN A MIDLY CALCIFIED VESSEL. A 4.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NON-BSC DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8103018 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page