Skip to content

Symmetry™ M001103860Recalls & reported problems

Boston Scientific Corporation · Balloon Dilatation Catheter · Class 2

All models of Symmetry (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 08714729181934), received between Jul 31, 2019 and Dec 4, 2025 — classified as 1 injury, 11 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture11
  • Difficult to Remove1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 4, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED VESSEL IN THE LEFT ELBOW. A 5.5-40/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23716950 ↗
  • MalfunctionMay 21, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY CALCIFIED VESSEL. A 5.5-4/4T/90 SYMMETRY BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE SECOND INFLATION AT THE MIDDLE OF A RATED PRESSURE FOR 15 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19366674 ↗
  • MalfunctionSep 11, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL. A 5.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT BELOW NOMINAL PRESSURE, THE BALLOON RUPTURED IMMEDIATELY AFTER INFLATION WAS STARTED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17721175 ↗
  • MalfunctionApr 25, 2023Leak/Splash; Leak/Splash

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2023. IT WAS REPORTED THAT BALLOON LEAK WAS ENCOUNTERED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED ARTERIOVENOUS FISTULA VEIN SIDE. A 5.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION BUT WHEN AN INFLATION DEVICE WAS ATTACHED AND PRESSURIZATION WAS PERFORMED, THE PRESSURE DID NOT INCREASE. THE DEVICE WAS REMOVED FROM THE PATIENT BODY AND CHECKED; THUS, IT WAS NOTED THAT THE INFLATION SOLUTION WAS LEAKING FROM THE MIDDLE OF THE SHAFT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED BALLOON PINHOLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16807778 ↗
  • InjuryMar 24, 2023Difficult to Remove; Material Rupture; Difficult to Remove; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE AND REMOVAL DIFFICULTIES REQUIRING SURGICAL INTERVENTION OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MILDLY TORTUOUS AND NON-CALCIFIED CEPHALIC VEIN. A 5.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, AFTER SEVERAL DILATATION AT 16 ATMOSPHERES, THE BALLOON RUPTURED HORIZONTALLY. ATTEMPTS WERE MADE TO RETRIEVE THE DEVICE; HOWEVER, THE BALLOON PORTION WAS CAUGHT IN THE SHEATH AND WAS NOT ABLE TO REMOVE. THE PHYSICIAN REMOVED THE BALLOON SURGICALLY AND AFTER IT RUPTURED, CONTRAST SHOWED THAT THE LESION WAS DILATED, THEN THE PROCEDURE WAS TERMINATED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16611218 ↗
  • MalfunctionApr 8, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN AN ARTERIOVENOUS FISTULA IN A MODERATELY TORTUOUS AND MILDLY CALCIFIED VENOUS SIDE. A 5.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. THE BALLOON WAS INITIALLY INFLATED AT 6 ATMOSPHERES FOR 30 SECONDS. HOWEVER, DURING SECOND INFLATION AT 10 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED FROM THE PATIENT WITHOUT ANY PROBLEM AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT STATUS WAS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14053454 ↗
  • MalfunctionJan 13, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN AN ARTERIOVENOUS FISTULA IN A MILDLY TORTUOUS AND MILDLY CALCIFIED VEIN. A SV/5.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. INFLATION WAS PERFORMED AT 6 ATMOSPHERES AND 15 ATMOSPHERES FOR 30 SECONDS RESPECTIVELY. HOWEVER, DURING THE SECOND INFLATION, THE BALLOON WOULD NOT HOLD PRESSURE. THE DEVICE WAS REMOVED AND A PINHOLE WAS FOUND NEAR THE TIP OF THE BALLOON. THE PHYSICIAN NOTED THAT IT IS PROBABLE THAT THERE WAS A PINHOLE DURING PREPARATION BECAUSE THE NEGATIVE PRESSURE DID NOT WORK WELL. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13245543 ↗
  • MalfunctionDec 15, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 5.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 4 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED AND THE PATIENT CONDITION WAS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13008193 ↗
  • MalfunctionMay 4, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A SV/5.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 14 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED AND THE PATIENT WAS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11773246 ↗
  • MalfunctionDec 9, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VEIN. A SV/5.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 15 ATMOSPHERES FOR 20 SECONDS, THE BALLOON RUPTURED. THE DEVICE REPLACED WITH A NON-BSC PRODUCT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10977463 ↗
  • MalfunctionNov 12, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED VEIN. A 5.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 15 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A 6MM NON-BSC DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9313289 ↗
  • MalfunctionJul 31, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED VEIN. A 5.5-4/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING SECOND DILATION AT 15 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8849073 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page