Skip to content

Symmetry™ M001103330Recalls & reported problems

Boston Scientific Corporation · Balloon Dilatation Catheter · Class 2

All models of Symmetry (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729181958), received between Jun 12, 2019 and Apr 5, 2023 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Rupture6
  • Component Missing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2
  • No Patient Involvement1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 5, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT UNDERWENT SHUNT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA). THE 99% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED VESSEL. A SV/6.0-2/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INITIAL INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING THE NORMAL METHOD WITHOUT ANY PROBLEM AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16683522 ↗
  • MalfunctionMar 16, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY IN THE 95% STENOSED TARGET LESION LOCATED IN A SHUNT. A 6.0-2/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16559122 ↗
  • MalfunctionApr 12, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN AN ARTERIOVENOUS SHUNT IN THE MODERATELY TORTUOUS VESSEL. A 6.0-2/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION AND WAS INITIALLY INFLATED AT 8 ATMOSPHERES (ATM) FOR 1 MINUTE. THE SECOND INFLATION AT 10 ATMOSPHERES FOR A MINUTE. HOWEVER, DURING THE THIRD INFLATION AT 10 ATMOSPHERES FOR 1 MINUTE, THE BALLOON RUPTURED, THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION WAS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14084295 ↗
  • MalfunctionFeb 10, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A SV/6.0-2/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING FIRST INFLATION AT 8 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S CONDITION WAS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11306738 ↗
  • MalfunctionApr 24, 2020Component Missing; Component Missing

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 02-APR 2020. IT WAS REPORTED THAT BALLOON CATHETER WAS MISSING. A 6.0-2/4T/90 SYMMETRY BALLOON CATHETER WAS FOUND MISSING WHEN OUT OF THE PACKAGE DURING PREPARATION. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED SHAFT DETACHED.

    Patient problems: No Patient Involvement

    MAUDE report 9995825 ↗
  • MalfunctionSep 6, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN A SHUNT IN THE FOREARM VEIN. A 6.0-2/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8971186 ↗
  • MalfunctionJun 12, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE CORONARY ARTERY. A 6.0-2/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE SECOND INFLATION AT 15 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8692755 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates