Skip to content

Symmetry™ Stiff Shaft M001105200Recalls & reported problems

Boston Scientific Corporation · Balloon Dilatation Catheter · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08714729182757), received between Oct 19, 2020 and Nov 24, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture9
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED ANASTOMOSIS SITE. A 4.0-10/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON THIRD INFLATION AT 10 ATMOSPHERES FOR 60 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23628917 ↗
  • MalfunctionOct 15, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL. A 4.0-10/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE INITIAL INFLATION AT 10 ATMOSPHERES FOR ONE MINUTE, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23288753 ↗
  • MalfunctionJul 16, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND NON-CALCIFIED CEPHALIC VEIN OF THE FOREARM. A 4.0-10/4T/90 SYMMETRY? STIFF SHAFT BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 8 ATMOSPHERES FOR 30 SECONDS, THE BALLOON VERTICALLY RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT INJURIES REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22523183 ↗
  • MalfunctionFeb 25, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED SHUNT WAS LOCATED IN THE NON-CALCIFIED VESSEL IN THE FOREARM. A 4.0-10/4T/90 SYMMETRY STIFF SHAFT BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE SECOND INFLATION AT 15 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21457269 ↗
  • MalfunctionJan 30, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE SHUNT VEIN. A SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 10 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21279768 ↗
  • MalfunctionApr 12, 2023Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN AN ARTERIOVENOUS GRAFT LOOP SHUNT IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED VESSEL IN THE FOREARM. A 4.0-10/4T/90 SYMMETRY BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING FOURTH INFLATION AT 15 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED VERTICALLY. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16726403 ↗
  • MalfunctionJan 8, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED ARTERIOVENOUS ANASTOMOSIS. TWO 4.0-10/4T/90 SYMMETRY STIFF SHAFT BALLOON CATHETERS WERE ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT 12 ATMOSPHERES, THE BALLOONS RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11139665 ↗
  • MalfunctionJan 8, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED ARTERIOVENOUS ANASTOMOSIS. TWO 4.0-10/4T/90 SYMMETRY STIFF SHAFT BALLOON CATHETERS WERE ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT 12 ATMOSPHERES, THE BALLOONS RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11139654 ↗
  • MalfunctionOct 19, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL. A SS/4.0-10/4T/90 SYMMETRY STIFF SHAFT BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT 15 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED WITHOUT ANY ISSUE NOTED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT'S STATUS WAS GOOD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10702487 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page