Skip to content

Flexima™ APDL M001281340Recalls & reported problems

Boston Scientific Corporation · Drainage Catheter System with Dissolving Tip · Class 2 · Implant

All models of Flexima APDL (12)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729191780), received between Mar 18, 2021 and Mar 18, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Collapse” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Collapse1
  • Detachment of Device or Device Component1
  • Difficult to Advance1

Patient problems

  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 18, 2021Collapse; Detachment of Device or Device Component; Difficult to Advance; Collapse; Detachment of Device or Device Component; Difficult to Advance

    IT WAS REPORTED THAT THE TIP DETACHMENT AND RETRIEVED DEVICE COMPONENT OCCURRED. THE DRAINAGE CATHETER WAS ATTACHED TO THE PATIENT'S KIDNEYS. A 8/25 FLEXIMA APDL WAS SELECTED FOR USE. THE FLEXIMA APDL CATHETER WAS ADVANCED OVER THE WIRE. THE CATHETER WAS MEETING RESISTANCE IN THE RETROPERITONEAL SOFT TISSUES AND THE TIP WAS TELESCOPING. THE PHYSICIAN PULLED THE CATHETER BACK AND NOTICED THAT THE CLEAR TAPERED PIECE AT THE TIP WAS MISSING. THIS PIECE WAS SEEN ON THE CATHETER WHILE IT WAS LOADED ON THE WIRE. IT WAS CONCLUDED THAT THE PIECE SOMEHOW DETACHED FROM THE CATHETER INSIDE THE PATIENT. THE DETACHED CATHETER TIP COULD NOT BE LOCATED WITHIN THE PATIENT. CONSEQUENTLY, THE PHYSICIAN DISSECTED THE SOFT TISSUES IN ATTEMPT TO LOCATED THE DETACHED TIP. LOCATING THE DETACHED TIP WAS UNSUCCESSFUL. IT WAS NOTED THAT THE DETACHED CATHETER TIP WAS STILL ON THE WIRE ABOUT 4-5 CM DEEP TO THE SKIN AND THE PHYSICIAN BELIEVES THIS LIKELY PUSHED THE PART THROUGH THE KIDNEY AND INTO THE COLLECTING SYSTEM. THE PROCEDURE WAS COMPLETED WITH A NON BOSTON SCIENTIFIC DEVICE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 11517099 ↗

Other TUBE, DRAINAGE, SUPRAPUBIC devices

All TUBE, DRAINAGE, SUPRAPUBIC devices →

Sources and updates