Flexima™ APDL M001275030Recalls & reported problems

Boston Scientific Corporation · Drainage Catheter System Kit · Class 2 · Implant

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729300694), received between Dec 14, 2023 and Feb 6, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation3
  • Break1

Patient problems

  • Foreign Body In Patient3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE STENT BROKE. ON (B)(6) 2025, A 7.3 FLEXIMA ALL-PURPOSE DRAINAGE CATHETER WAS SELECTED FOR PERCUTANEOUS LIVER ABSCESS DRAINAGE. HOWEVER, THE FOLLOWING DAY, IT WAS NOTED THROUGH THE COMPUTERIZED TOMOGRAPHY (CT) THAT THE TUBE HAD COME OFF WITHIN THE LIVER. SUBSEQUENTLY, ON (B)(6) 2025, THE TUBE STENT WAS REMOVED. THE PIGTAIL WAS NOT STRAIGHTENED, SO THE TUBE WAS REMOVED WHILE STRAIGHTENING IT WITH A GUIDEWIRE. THE SUTURE SEEMED TO GET CAUGHT ON SOMETHING AND COULD NOT BE PULLED OUT, SO THE PROCEDURE WAS UNAVOIDABLY COMPLETED WITH A REMAINING SUTURE INSIDE THE BODY, APPROXIMATELY 2CM FROM THE BODY SURFACE. ON (B)(6) 2025, THE SUTURE WAS COMPLETELY PULLED INSIDE THE PATIENT WHEN VISUALLY CHECKED. AT THIS TIME, THERE WERE NO ADVERSE EFFECTS TO THE PATIENT. AS PER PHYSICIAN, IT IS POSSIBLE THAT A STRUCTURE WAS FORMED WITHIN THE LIVER AND THE SUTURE MAY HAVE BEEN CAUGHT THERE. HOWEVER, THE PHYSICIAN DOUBTED THAT A STRUCTURE COULD BE FORMED SO QUICKLY BECAUSE IT HAS NOT BEEN LONG SINCE IT WAS PLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21323295 ↗
  • InjuryDec 21, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT THE SUTURE WAS IMPLANTED. A 7.3 FR FLEXIMA ALL PURPOSE DRAINAGE CATHETER WITH LOCK WAS SELECTED FOR USE. DURING DEVICE REMOVAL, A RESISTANCE WAS FELT WHEN THE REMAINING SUTURE WAS RETRIEVED. THERE WAS A CONCERN THAT THIS WOULD DAMAGE THE ORGANS BY PULLING OUT THE DEVICE BY FORCE. THE SUTURE WAS CUT OUT AND THE REMAINING PART WAS IMPLANTED SUBCUTANEOUSLY. THE PROCEDURE WAS COMPLETED, AND THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 18372698 ↗
  • InjuryDec 14, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT THE BROKEN SUTURE REMAINED IN THE PATIENT. THE TARGET LESION WAS LOCATED IN THE NOT TIGHTLY TORTUOUS BILE DUCT. A 7.3 FLEXIMA APDL ALL PURPOSE DRAINAGE WAS SELECTED FOR IMPLANT. DURING REMOVAL, FIRST CATHETER WAS CUT AND PULLED OUT. HOWEVER, RESISTANCE WAS ENCOUNTERED WHEN THE SUTURE WAS PULLED OUT AND IT GOT SEPARATED, AND REMAINED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THE PATIENT WAS STABLE AFTER THE PROCEDURE; HOWEVER, AN ADVERSE HEALTH EFFECTS WAS CONFIRMED.

    Patient problems: Foreign Body In Patient

    MAUDE report 18330795 ↗

Other TUBE, DRAINAGE, SUPRAPUBIC devices

All TUBE, DRAINAGE, SUPRAPUBIC devices →