Flexima™ APDL M001271330Recalls & reported problems
Boston Scientific Corporation · Drainage Catheter System · Class 2 · Implant
All models of Flexima APDL (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729323099), received between Sep 14, 2020 and Apr 15, 2022 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Difficult to Remove3
- Break2
- Defective Device2
- Difficult to Advance2
- Adverse Event Without Identified Device or Use Problem1
- Deformation Due to Compressive Stress1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Pain2
- Device Embedded In Tissue or Plaque1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 05APR2022. A FLEXIMA ALL PURPOSE DRAINAGE CATHETER WAS RETURNED FOR ANALYSIS WITH NO REPORTED COMPLAINT. HOWEVER, DEVICE ANALYSIS REVEALED THE STENT HAD DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14120354 ↗MalfunctionApr 15, 2022Difficult to Remove; Difficult to Remove
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 06APR2022. IT WAS REPORTED THAT THE TROCAR COULD NOT BE REMOVED FROM THE CATHETER. A FLEXIMA ALL PURPOSE DRAINAGE CATHETER WAS SELECTED FOR USE. THE TROCAR COULD NOT BE REMOVED FROM THE CATHETER. THE PROCEDURE WAS COMPLETED USING AN ALTERNATIVE METHOD. THERE WERE NO PATIENT COMPLICATIONS. HOWEVER, DEVICE ANALYSIS REVEALED THE STENT HAD DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14120346 ↗MalfunctionFeb 3, 2022Difficult to Remove; Defective Device; Deformation Due to Compressive Stress; Difficult to Advance; Difficult to Remove; Defective Device; Deformation Due to Compressive Stress; Difficult to Advance
IT WAS REPORTED THAT THE REMOVAL DIFFICULTY OCCURRED. TWO APDL/6.3 FLEXIMA REGULAR ALL PURPOSE DRAINAGE WERE SELECTED FOR USE. THE FIRST DEVICE WAS ADVANCED OVER THE GUIDEWIRE AND THROUGH THE INCISION IN THE SKIN AND INSERTED IN THE PATIENT. DURING INSERTION, THE DEVICE STARTED TO CRUMPLE LIKE AN ACCORDION AND GOT STUCK. NO KINKS WERE NOTED. THE DEVICE WAS CAREFULLY REMOVED FROM PATIENT WITH INCREASED PULLING FORCE AND WAS A BIT PAINFUL. THERE WAS NO PERMANENT DAMAGE TO THE PATIENT. ANOTHER OF THE SAME DEVICE WAS USED; HOWEVER, THE SAME ISSUE OCCURRED. AFTER CONFIRMING THE GUIDE WIRE WAS STILL IN THE RIGHT PLACE, AN APDL 8 FLEXIMA REGULAR CATHETER WAS USED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO COMPLICATIONS REPORTED.
Patient problems: Pain
MAUDE report 13446461 ↗MalfunctionFeb 3, 2022Difficult to Remove; Defective Device; Difficult to Advance; Difficult to Remove; Defective Device; Difficult to Advance
IT WAS REPORTED THAT THE REMOVAL DIFFICULTY OCCURRED. TWO APDL/6.3 FLEXIMA REGULAR ALL PURPOSE DRAINAGE WERE SELECTED FOR USE. THE FIRST DEVICE WAS ADVANCED OVER THE GUIDEWIRE AND THROUGH THE INCISION IN THE SKIN AND INSERTED IN THE PATIENT. DURING INSERTION, THE DEVICE STARTED TO CRUMPLE LIKE AN ACCORDION AND GOT STUCK. NO KINKS WERE NOTED. THE DEVICE WAS CAREFULLY REMOVED FROM PATIENT WITH INCREASED PULLING FORCE AND WAS A BIT PAINFUL. THERE WAS NO PERMANENT DAMAGE TO THE PATIENT. ANOTHER OF THE SAME DEVICE WAS USED; HOWEVER, THE SAME ISSUE OCCURRED. AFTER CONFIRMING THE GUIDE WIRE WAS STILL IN THE RIGHT PLACE, AN APDL 8 FLEXIMA REGULAR CATHETER WAS USED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO COMPLICATIONS REPORTED.
Patient problems: Pain
MAUDE report 13446226 ↗InjuryNov 25, 2020Break; Break
IT WAS REPORTED THAT THE DISTAL END OF THE DEVICE DETACHED. A 6.3 FLEXIMA APDL WAS USED FOR TREATMENT. UPON REMOVAL, THE DISTAL PIGTAIL END OF THE DRAINAGE CATHETER DETACHED AND REMAINED INSIDE THE PATIENT. ANOTHER PROCEDURE WAS PERFORMED TO SUCCESSFULLY REMOVE THE DETACHED PIGTAIL END. THERE WERE NO FURTHER COMPLICATIONS REPORTED AND THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10906085 ↗InjurySep 14, 2020Break; Break
IT WAS REPORTED THAT THE DRAINAGE CATHETER DETACHED AND REMAINED IN THE PATIENT. A 6.3 FLEXIMA APDL WAS SELECTED FOR A RENAL CYST PUNCTURE PROCEDURE UNDER LOCAL ANESTHESIA. DURING PROCEDURE, THE CATHETER WAS PLACED TO REACH THE RENAL CYST. THE DRAINAGE CATHETER WAS PUT IN PLACE ON THE GUIDE AND THE GUIDE WAS REMOVED. THE CYST WAS DRAINED WITH THE DRAINAGE CATHETER, THEN IT WAS INJECTED WITH CONTRAST PRODUCT AND IT WAS REMOVED BY THE DRAINAGE CATHETER, THEN INJECTED WITH 40CC ALCOHOL WITH A CONTACT TIME OF 15MIN. THEN, WHEN IT WAS NECESSARY TO REMOVE THE ALCOHOL ONLY 10CC WAS REMOVED, SO THE DRAINAGE CATHETER WAS MOVED IN ORDER TO RECOVER MORE ALCOHOL. THE PHYSICIAN NOTED A RUPTURE IN THE DRAINAGE CATHETER AND IT DETACHED INSIDE THE PATIENT. ONE PART OF THE DETACHED FRAGMENT REMAINED IN THE HANDS OF THE PHYSICIAN. AN ULTRASOUND SCAN PERFORMED THAT REVEALED A TWO FRAGMENT PIECES OF THE DRAINAGE CATHETER WERE INSIDE THE PATIENT. ONE PART OF THE DETACHED FRAGMENT REMAINED IN THE CYST AND ANOTHER PART WAS FOUND IN FRONT OF THE CYST AT THE PERIRENAL LEVEL. NO SURGERY WAS PERFORMED TO RETRIEVE THE TWO FRAGMENT PIECES OF THE DRAINAGE CATHETER.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10528460 ↗
Other TUBE, DRAINAGE, SUPRAPUBIC devices
- Flexima™ APDL M001271380Boston Scientific Corporation · TUBE, DRAINAGE, SUPRAPUBIC17 reports
- Flexima™ APDL M001271340Boston Scientific Corporation · TUBE, DRAINAGE, SUPRAPUBIC2 deaths16 reports
- Flexima™ APDL M001271960Boston Scientific Corporation · TUBE, DRAINAGE, SUPRAPUBIC1 death10 reports
- Flexima™ APDL M001271350Boston Scientific Corporation · TUBE, DRAINAGE, SUPRAPUBIC9 reports
- Flexima™ APD™ M001271360Boston Scientific Corporation · TUBE, DRAINAGE, SUPRAPUBIC8 reports
- Flexima™ APDL M001275010Boston Scientific Corporation · TUBE, DRAINAGE, SUPRAPUBIC3 reports
- Flexima™ APDL M001275030Boston Scientific Corporation · TUBE, DRAINAGE, SUPRAPUBIC3 reports
- Flexima™ APDL M001271310Boston Scientific Corporation · TUBE, DRAINAGE, SUPRAPUBIC2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page