RX Biliary M00545680Recalls & reported problems

Boston Scientific Corporation · Biliary Stent with RX Delivery System · Class 2 · Implant

All models of RX Biliary (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729301851), received between Jan 6, 2022 and Feb 8, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break2
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 8, 2022Break; Positioning Failure; Break; Positioning Failure

    NOTE: THIS REPORT PERTAINS TO THE SECOND OF TWO DEVICES THAT WERE USED IN THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA RX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, IT WAS NOTICED THAT THE GUIDE CATHETER WAS BROKEN. THE BROKEN GUIDE CATHETER WAS REMOVED FROM THE PATIENT. THE PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED AND SHOWED THAT THE PUSH CATHETER WAS KINKED AND THE GUIDE CATHETER WAS DETACHED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13478900 ↗
  • InjuryJan 6, 2022Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA RX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE COMMON BILE DUCT, PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT, IT WAS NOTICED THE GUIDE CATHETER WAS BROKEN. THE STENT WAS SUCCESSFULLY DEPLOYED AND THE BROKEN PIECE OF GUIDE CATHETER WAS RETRIEVED WITH THE USE OF RAT TOOTH AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13175062 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices