Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI H862GPL21030H0), received between Sep 10, 2019 and Sep 10, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Pneumothorax1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 10, 2019Fracture; Fracture

    A URESIL DRAINAGE CATHETER WAS PLACED IN THE PATIENT AS A CHEST DRAINAGE TUBE ON (B)(6) 2019. THE NEXT DAY ((B)(6) 2019), A LEAK WAS DETECTED AND THE CATHETER WAS FOUND TO BE BLOCKED. THE CATHETER WAS FLUSHED WITH SALINE AND HUB LEAKAGE WAS AGAIN NOTED. EVEN THOUGH THE HOSPITAL WAS AWARE OF THIS LEAK, THE CATHETER WAS NOT REPLACED UNTIL 2 DAYS LATER - (B)(6) 2019. THE HOSPITAL ALSO STATED THAT THEY INTENTIONALLY "OVER-TIGHTEN" THESE TYPES OF CATHETER CONNECTIONS, AND MOST LIKELY USED A METAL INSTRUMENT TO OVER-TIGHTEN THE CONNECTION. THE HOSPITAL RECOGNIZES THAT THE LEAKING CATHETER SITUATION SHOULD HAVE BEEN ADDRESSED SOONER, AND IT IS URESIL'S UNDERSTANDING THAT THEY INTEND TO CHANGE THEIR INTERNAL PROCEDURES REGARDING TIMELY ADDRESSING OF NOTED ISSUES LIKE LEAKING DEVICES AND TO NOT USE METAL INSTRUMENTS TO TIGHTEN CONNECTIONS.

    Patient problems: Pneumothorax

    MAUDE report 8981077 ↗

Other Catheter, Nephrostomy, General & Plastic Surgery devices