RX Biliary M00545550Recalls & reported problems
Boston Scientific Corporation · Biliary Stent with RX Delivery System · Class 2 · Implant
All models of RX Biliary (13)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729337850), received between Apr 3, 2020 and Oct 22, 2021 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
- Adverse Event Without Identified Device or Use Problem1
- Tear, Rip or Hole in Device Packaging1
- Unsealed Device Packaging1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 22, 2021Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA RX BILIARY STENT WAS OPENED TO BE USED FOR AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE. THE EVENT DATE WAS NOT REPORTED. DURING PREPARATION, IT WAS NOTED THAT THERE WAS A HOLE IN THE PACKAGING. IT WAS REPORTED THAT THE STERILITY OF THE DEVICE WAS COMPROMISED. THEY OPENED ANOTHER FLEXIMA RX BILIARY STENT AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12686045 ↗InjurySep 14, 2020Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA RX BILIARY STENT WAS USED DURING A BILIARY DRAINAGE IN THE BILE DUCT PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE GUIDE CATHETER BECAME SEPARATED DURING ATTEMPTED STENT DEPLOYMENT. THE BROKEN GUIDE CATHETER REMAINED INSIDE THE STENT WHICH WAS PLACED IN THE BILE DUCT. THE PHYSICIAN THOUGHT THE STENT WAS PUSHED BY A STRICTURE AND THERE WAS A SEVERE RESISTANCE WHEN THE INNER CATHETER WAS REMOVED. A PAIR OF FORCEPS WAS USED TO REMOVE THE BROKEN GUIDE CATHETER AND THE STENT FROM THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER FLEXIMA RX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED ON THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10528797 ↗InjuryMay 21, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA RX BILIARY STENT WAS USED DURING A STENT PLACEMENT PROCEDURE IN THE COMMON BILE DUCT PERFORMED ON (B)(6), 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHILE ATTEMPTING TO DEPLOY THE STENT, THE GUIDE CATHETER BECAME SEPARATED AND REMAINED INSIDE THE STENT. THE PHYSICIAN USED GRASPING FORCEPS TO REMOVE THE BROKEN GUIDE CATHETER FROM THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10080735 ↗MalfunctionApr 3, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA RX BILIARY STENT WAS USED DURING A BILIARY STENT PLACEMENT IN THE LOWER BILE DUCT, PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE STENT WAS RELEASED, THEY HAVE OBSERVED THAT THE STENT WAS UNABLE TO SEPARATE FROM THE DELIVERY SYSTEM. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER FLEXIMA RX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THAT THE GUIDE CATHETER WAS BROKEN THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9922648 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page