Skip to content

UroMax Ultra Kit M0062251210Recalls & reported problems

Boston Scientific Corporation · Balloon Dilatation Catheter Kit · Class 2

All models of UroMax Ultra Kit (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729341284), received between Jul 26, 2017 and Dec 23, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 5, 2025Class IIOpen, Classified

    The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

    Z-2536-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Burst Container or Vessel2
  • Material Puncture/Hole2
  • Device Contamination with Chemical or Other Material1
  • Fluid/Blood Leak1
  • Leak/Splash1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 23, 2025Fluid/Blood Leak; Leak/Splash; Fluid/Blood Leak; Leak/Splash

    IT WAS REPORTED THAT A UROMAX ULTRA DILATION BALLOON CATHETER DEVICE WAS USED DURING A URETEROSCOPY (URS), STRICTURE DILATION PROCEDURE. DURING THE PROCEDURE, IT WAS NOTICED THAT THE BALLOON WAS NOT INFLATING. REPORTEDLY, UPON REMOVAL IT WAS FOUND THAT CONTRAST SOLUTION WAS LEAKING OUT FROM THE BALLOON. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23877655 ↗
  • MalfunctionFeb 28, 2022Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA DILATION BALLOON CATHETER WAS USED DURING A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE BALLOON WOULD NOT INFLATE. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA DILATION BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF BALLOON PINHOLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13618399 ↗
  • MalfunctionFeb 4, 2020Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA DILATATION BALLOON CATHETER WAS USED IN THE URETER DURING A PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON RUPTURED AFTER BEING INFLATED THE THIRD TIME. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9664283 ↗
  • MalfunctionMay 23, 2019Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA URETERAL DILATATION BALLOON WAS USED IN THE URETER DURING A UTEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON RUPTURED DURING INFLATION. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8637930 ↗
  • MalfunctionApr 30, 2019Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA DILATATION BALLOON CATHETER WAS USED IN THE URETER DURING A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTICED THAT THE BALLOON HAD A HOLE WHEN TRIED TO BE INFLATED. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA DILATATION BALLOON CATHETER DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THAT THE BALLOON FOUND A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8566328 ↗
  • MalfunctionJul 26, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA BALLOON KIT WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING PROCEDURE, WHILE UNPACKING THERE WAS A HAIR FOUND INSIDE OF THE STERILE PACKAGING. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA BALLOON KIT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE UNHARMED AND STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6743678 ↗