Skip to content

UroMax Ultra Kit M0062251310Recalls & reported problems

Boston Scientific Corporation · Balloon Dilatation Catheter Kit · Class 2

All models of UroMax Ultra Kit (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729341383), received between Mar 24, 2020 and Dec 27, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 5, 2025Class IIOpen, Classified

    The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

    Z-2536-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Inflation Problem1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 27, 2021Burst Container or Vessel; Inflation Problem; Burst Container or Vessel; Inflation Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT THE A UROMAX ULTRA DILATION BALLOON CATHETER WAS USED IN THE URETHRA DURING A URETER STENOSIS PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, IT WAS NOTICED THAT THE BALLOON WAS DAMAGED WHEN INFLATED TO 16 ATM. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA DILATION BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. ADDITIONAL INFORMATION RECEIVED ON DECEMBER 24, 2021. IT WAS CLARIFIED THAT THE ONLY DAMAGED NOTICED WAS THE BALLOON WOULD JUST NOT INFLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13087216 ↗
  • MalfunctionMar 24, 2020Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA BALLOON WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING PROCEDURE, AFTER THE BALLOON WAS INFLATED WITH CONTRAST AGENT, IT WAS FOUND THAT THE BALLOON HAD A SMALL PUNCTURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT TO THIS EVENT. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9873567 ↗

Other DILATOR, URETHRAL devices

All DILATOR, URETHRAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page