UroMax Ultra Kit M0062251220Recalls & reported problems

Boston Scientific Corporation · Balloon Dilatation Catheter Kit · Class 2

All models of UroMax Ultra Kit (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729341291), received between Feb 19, 2019 and Oct 31, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Unsealed Device Packaging” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 5, 2025Class IIOpen, Classified

    The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

    Z-2536-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Unsealed Device Packaging2
  • Device Contaminated During Manufacture or Shipping1
  • Device Contamination with Chemical or Other Material1
  • Difficult to Remove1
  • Material Deformation1
  • Material Puncture/Hole1
  • Material Rupture1
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 31, 2025Material Rupture; Device Contamination with Chemical or Other Material; Material Rupture; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT FOUR UROMAX ULTRA BALLOON DILATATION CATHETER DEVICES WERE SET FOR USE IN A PROCEDURE. HOWEVER, FOREIGN BODIES WERE FOUND INSIDE THE ENCORE INFLATION DEVICES OF THE UROMAX ULTRA BALLOON KIT. ADDITIONALLY, THE UROMAX ULTRA BALLOON DEVICES WERE ALSO NOTED TO BE RUPTURED. THE PROCEDURE WAS COMPLETED WITH NO PATIENT COMPLICATIONS. THIS REPORT PERTAINS TO THE FOURTH OF FOUR UROMAX ULTRA BALLOON DILATATION CATHETER DEVICES IN REFERENCE TO MEDWATCH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23441968 ↗
  • MalfunctionNov 20, 2023Unsealed Device Packaging; Device Contaminated During Manufacture or Shipping; Unsealed Device Packaging; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN UROMAX ULTRA DILATION BALLOON CATHETER DEVICE WAS ABOUT TO BE USED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, UPON OPENING OF THE PACKAGE IT WAS FOUND THAT THE ANCHOR INFLATOR STERILIZATION WAS OPEN AND NOT CLEAN. THE PROCEDURE WAS STILL COMPLETED WITH ANOTHER UROMAX ULTRA DILATION BALLOON CATHETER DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18177632 ↗
  • MalfunctionFeb 3, 2023Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NUROMAX ULTRA KIT WITH ENCORE 26 INFLATOR WAS USED IN THE KIDNEY DURING A LITHOTRIPSY PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, THE PROTECTIVE SLEEVE OF THE BALLOON WAS DIFFICULT TO REMOVE. THE PROCEDURE WAS COMPLETED WITH ANOTHER NUROMAX ULTRA DILATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF SHAFT KINKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16292662 ↗
  • MalfunctionJul 6, 2020Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX DILATATION BALLOON CATHETER WAS USED IN THE URETER DURING A BALLOON DILATION PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTICED THAT THE BALLOON HAD A PINHOLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX DILATATION BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10232666 ↗
  • MalfunctionFeb 19, 2019Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging; Unsealed Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A UROMAX ULTRA DILATATION BALLOON WAS USED DURING A STONE EXTRACTION PROCEDURE PERFORMED LAST (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, THE DEVICE POUCH WAS TORN WHILE OPENING THE PACKAGE AND THE STERILE SEAL WAS COMPROMISED. REPORTEDLY, THIS MIGHT BE DUE TO MANUFACTURE DEFECT. THE PROCEDURE WAS COMPLETED WITH ANOTHER UROMAX ULTRA DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENTS CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.

    Patient problems: No Consequences Or Impact To Patient