Segura Hemisphere M0063801120Recalls & reported problems

Boston Scientific Corporation · Stone Retrieval Basket · Class 2

All models of Segura Hemisphere (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08714729341949), received between Dec 7, 2016 and Feb 9, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Detachment of Device or Device Component4
  • Material Split, Cut or Torn2
  • Torn Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 9, 2026Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT A SEGURA HEMI BASKET WAS SET TO BE USED DURING A URETEROSCOPIC LITHOTRIPSY (URSL) PROCEDURE. REPORTEDLY, BEFORE THE SURGERY, IT WAS FOUND THAT THE BASKET HANDLE COMPONENT FELL OFF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ANALYSIS OF THE RETURNED DEVICE REVEALED THAT ONE OF THE BASKET WIRES WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24299742 ↗
  • MalfunctionSep 16, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT A SEGURA HEMI BASKET WAS ABOUT TO BE USED DURING A URETEROSCOPIC LITHOTRIPSY (URSL) PROCEDURE. REPORTEDLY, DURING UNPACKING IT WAS FOUND THAT THE BASKET HANDLE COMPONENT FALLS OFF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS. ANALYSIS OF THE RETURNED DEVICE REVEALED THAT ONE OF THE BASKET WIRES WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23064480 ↗
  • MalfunctionJul 8, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT A SEGURA HEMI BASKET WAS ABOUT TO BE USED DURING A URETEROSCOPIC LITHOTRIPSY (URSL) PROCEDURE. REPORTEDLY, BEFORE THE PROCEDURE, IT WAS FOUND THAT THE BASKET HANDLE COMPONENT FALLS OFF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS. ANALYSIS OF THE RETURNED DEVICE REVEALED THAT ONE OF THE BASKET WIRES WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22449363 ↗
  • MalfunctionMay 8, 2025Break; Break

    IT WAS REPORTED THAT A SEGURA HEMI BASKET DEVICE WAS ABOUT TO BE USED DURING A URETEROSCOPIC LITHOTRIPSY (URSL) PROCEDURE. DURING PREPARATION, IT WAS FOUND THAT THE BASKET CAN'T BE WORKED. ADDITIONALLY, IT WAS NOTICED THAT ONE OF THE BASKET WIRES GOT DETACHED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21986451 ↗
  • MalfunctionMar 27, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT A STONE CONE RETRIEVAL COIL WAS ABOUT TO BE USED DURING A URETEROSCOPIC LITHOTRIPSY (URSL) PROCEDURE. REPORTEDLY, THE BASKET HANDLE COMPONENT FALLS OFF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS. ANALYSIS OF THE RETURNED DEVICE REVEALED THAT ONE OF THE BASKET WIRES WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21713618 ↗
  • MalfunctionMar 17, 2022Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEGURA HEMISPHERE BASKET WAS USED IN THE URETER DURING A RETROGRADE INTRARENAL SURGERY (RIRS) PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, THE SHEATH WAS TORN ON THE DISTAL SECTION. THE PROCEDURE WAS COMPLETED WITH ANOTHER SEGURA HEMI BASKET. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION FOLLOWING PROCEDURE IS REPORTEDLY STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13792481 ↗
  • MalfunctionJan 13, 2020Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEGURA HEMISPHERE BASKET WAS USED IN A URETEROSCOPY (URS) PROCEDURE. THE EXACT DATE OF THE PROCEDURE IS UNKNOWN BUT WAS REPORTED TO BE EITHER (B)(6) 2019 OR (B)(6) 2019. PRELIMINARY INVESTIGATION OF THE DEVICE FOUND THAT THE SHEATH WAS TORN AT THE DISTAL END. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THE EVENT. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9582843 ↗
  • MalfunctionNov 8, 2019Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEGURA HEMISPHERE STONE RETRIEVAL BASKET WAS USED IN THE URETER DURING A URETEROSCOPY (URS) PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, AT THE TIME OF STONE CAPTURE, THE BASKET WIRE BROKE. REPORTEDLY, NO PART OF THE DEVICE WAS DETACHED INSIDE THE PATIENT. ANOTHER SEGURA BASKET WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9296541 ↗
  • MalfunctionDec 7, 2016Torn Material; Torn Material

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEGURA HEMISPHERE¿ BASKET WAS USED IN THE BILIARY TRACT DURING STONE MANAGEMENT PROCEDURE PERFORMED ON (B)(6) 2016. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, THE BASKET WAS UNABLE TO OPEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER SEGURA HEMISPHERE¿ BASKET. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THIS EVENT HAS BEEN DEEMED AN MDR-REPORTABLE EVENT BASED ON INVESTIGATION RESULTS WHICH REVEALED THAT THE SHEATH WAS FOUND TORN AT THE DISTAL END.