Segura Hemisphere M0063801180Recalls & reported problems
Boston Scientific Corporation · Stone Retrieval Basket · Class 2
All models of Segura Hemisphere (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729341970), received between May 4, 2018 and Jun 3, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break2
- Device Contamination with Chemical or Other Material1
- Material Deformation1
Patient problems
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 3, 2020Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEGURA HEMI BASKET WAS TO BE USED IN THE URETER DURING A URETEROSCOPIC LITHOTRIPSY (URSL) PROCEDURE PERFORMED ON (B)(6), 2020. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, WHEN THE HANDLE WAS TO BE SEPARATED DURING USE, THE INNER CORE OF THE BASKET WAS CROOKED, RESULTING IN POOR SEPARATION. A PHOTO OF THE COMPLAINT DEVICE SUBMITTED BY THE CUSTOMER SHOWS THE PULL WIRE BROKEN ON THE PROXIMAL END. THE PROCEDURE WAS COMPLETED WITH ANOTHER SEGURA HEMI BASKET. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10114021 ↗MalfunctionAug 19, 2019Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEGURA HEMI BASKET WAS USED IN A URETEROSCOPIC LITHOTRIPSY (URSL) PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, A BROWN STAIN WAS FOUND INSIDE THE CAP OF THE HANDLE AND WAS CONFIRMED BY A PHOTOGRAPH SENT FROM THE CUSTOMER. IT WAS NOTED THAT THERE WAS NO DAMAGED TO THE PACKAGING. ANOTHER SEGURA BASKET WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8906196 ↗MalfunctionMay 4, 2018Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEGURA HEMISPHERE¿ BASKET WAS USED IN THE KIDNEY FOR A STONE EXTRACTION PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE SEGURA BASKET WAS ACTUATED INSIDE THE PATIENT, THE PULL WIRE BROKE. THE BASKET WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS EXTENDED FOR ABOUT 15 MINUTES. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7488606 ↗
FDA clearance and classification
The FDA regulates product code FFL (dislodger, stone, basket, ureteral, metal) under 21 CFR 876.4680 as a Class II device. 21 CFR 876.4680 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Segura Hemisphere M0063801180
Has the Segura Hemisphere M0063801180 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729341970), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Segura Hemisphere M0063801180?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between May 4, 2018 and Jun 3, 2020. In those reports, the most common specific device problems are “Break” (2), “Device Contamination with Chemical or Other Material” (1) and “Material Deformation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Segura Hemisphere M0063801180?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Segura Hemisphere M0063801180 FDA approved?
Devices of this type (product code FFL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Boston Scientific Corporation.
Who makes the Segura Hemisphere M0063801180, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Segura Hemisphere, model M0063801180). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DISLODGER, STONE, BASKET, URETERAL, METAL devices
- Zero Tip M0063901050Boston Scientific Corporation · DISLODGER, STONE, BASKET, URETERAL, METAL286 reports
- NCircle G18778Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL222 reports
- NCircle G46206Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL195 reports
- NCircle G17520Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL127 reports
- NGage G48297Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL96 reports
- NGage G48294Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL92 reports
- NCircle G19110Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL87 reports
- Bard® SkyLite™ Tipless Nitinol Stone Basket 041900C. R. Bard, Inc. · DISLODGER, STONE, BASKET, URETERAL, METAL78 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page