Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729414995), received between Nov 1, 2021 and Nov 30, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 30, 2023Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ZERO TIP PULMONARY BASKET WAS USED IN THE BRONCHUS DURING THE REMOVAL OF INTRABRONCHIAL FOREIGN BODY PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, WHEN THE BASKET WAS OPENED TO RETRIEVE THE FOREIGN BODY, THE CONNECTION PORT BETWEEN THE HANDLE AND THE WIRE FRACTURED. THE DEVICE WAS REMOVED, AND ANOTHER ZERO TIP PULMONARY BASKET WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18240218 ↗
  • MalfunctionNov 1, 2021Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ZERO TIP PULMONARY BASKET WAS USED IN THE LUNGS DURING AN AIRWAY FOREIGN BODY RETRIEVAL PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, A BASKET WIRE BROKE WHEN ATTEMPTING TO CLOSE THE BASKET INSIDE THE PATIENT. THE BROKEN WIRE WAS STILL ATTACHED TO THE DEVICE. THE BASKET WAS PULLED OUT AND ANOTHER ZERO TIP PULMONARY BASKET WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions