Flexima™ M001272610Recalls & reported problems

Boston Scientific Corporation · Biliary Catheter System with Radiopaque Marker · Class 2 · Implant

All models of Flexima (20)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729454960), received between Sep 27, 2019 and Mar 27, 2025 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Deformation Due to Compressive Stress2
  • Difficult to Remove2
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding2
  • Pain2
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 27, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A TEAR IN THE PACKAGING OCCURRED. A 10.3 FLEXIMA DRAINAGE CATHETER WAS USED FOR PERCUTANEOUS TRANSHEPATIC BILIARY DRAINAGE (PTBD) IN THE GALLBLADDER. DURING UNPACKING, IT WAS OBSERVED THAT THERE WAS A TEAR IN THE PACKAGING FROM THE TOP PART AND THE DRAINAGE CATHETER WAS OUT FROM THE TORN PART. THE DEVICE WAS NO LONGER STERILE; THEREFORE, IT WAS NOT USED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21711001 ↗
  • InjurySep 8, 2023Difficult to Remove; Deformation Due to Compressive Stress; Difficult to Remove; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT REMOVAL DIFFICULTIES OCCURRED RESULTING IN INCREASED PAIN. THE PATIENT WAS UNDERGOING A BILE DUCT DRAINAGE PROCEDURE. A 10F FLEXIMA DRAINAGE CATHETER WAS INSERTED. THE INNER CANNULA AND THE WIRE BECAME STUCK DURING REMOVAL. WHILE ATTEMPTING TO REMOVE THE COMPONENTS, THE CATHETER ACCORDIONED IN THE PATIENT. THE PATIENT EXPERIENCED INCREASED PAIN WHICH WAS MANAGED WITH MIDAZOLAM AND FENTANYL. ADDITIONALLY, HEMATOBILIA WAS OBSERVED DUE TO IRRITATION OF THE BILE DUCT. THE ISSUE OCCURRED WITH TWO CATHETERS. THE DEVICE WAS ABLE TO BE REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT IS IN GOOD CONDITION.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 17714320 ↗
  • InjurySep 8, 2023Difficult to Remove; Deformation Due to Compressive Stress; Difficult to Remove; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT REMOVAL DIFFICULTIES OCCURRED RESULTING IN INCREASED PAIN. THE PATIENT WAS UNDERGOING A BILE DUCT DRAINAGE PROCEDURE. A 10F FLEXIMA DRAINAGE CATHETER WAS INSERTED. THE INNER CANNULA AND THE WIRE BECAME STUCK DURING REMOVAL. WHILE ATTEMPTING TO REMOVE THE COMPONENTS, THE CATHETER ACCORDIONED IN THE PATIENT. THE PATIENT EXPERIENCED INCREASED PAIN WHICH WAS MANAGED WITH MIDAZOLAM AND FENTANYL. ADDITIONALLY, HEMATOBILIA WAS OBSERVED DUE TO IRRITATION OF THE BILE DUCT. THE ISSUE OCCURRED WITH TWO CATHETERS. THE DEVICE WAS ABLE TO BE REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT IS IN GOOD CONDITION.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 17714313 ↗
  • MalfunctionOct 2, 2019Break; Break

    IT WAS REPORTED THAT THE SUTURE BROKE. UPON PREPARATION OUTSIDE THE PATIENT'S BODY, THE SUTURE OF A 10.3 FLEXIMA DRAINAGE CATHETER BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME SIZE. NO COMPLICATIONS REPORTED AND PATIENT DID WELL.

    Patient problems: No Patient Involvement

    MAUDE report 9145763 ↗
  • MalfunctionSep 27, 2019Break; Break

    IT WAS REPORTED THAT THE DEVICE WAS MISSING A RADIOPAQUE (RO) MARKER ON THE DRAIN. A 10.3 FLEXIMA REGULAR W/RO VTCB WAS SELECTED FOR USE FOR A BILIARY PROCEDURE. IT WAS NOTED THAT THERE WAS RO MARKER ON DRAIN. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9129349 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →