Flexima™ M001272610Recalls & reported problems
Boston Scientific Corporation · Biliary Catheter System with Radiopaque Marker · Class 2 · Implant
All models of Flexima (20)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729454960), received between Sep 27, 2019 and Mar 27, 2025 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Deformation Due to Compressive Stress2
- Difficult to Remove2
- Tear, Rip or Hole in Device Packaging1
Patient problems
- Hemorrhage/Blood Loss/Bleeding2
- Pain2
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 27, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT A TEAR IN THE PACKAGING OCCURRED. A 10.3 FLEXIMA DRAINAGE CATHETER WAS USED FOR PERCUTANEOUS TRANSHEPATIC BILIARY DRAINAGE (PTBD) IN THE GALLBLADDER. DURING UNPACKING, IT WAS OBSERVED THAT THERE WAS A TEAR IN THE PACKAGING FROM THE TOP PART AND THE DRAINAGE CATHETER WAS OUT FROM THE TORN PART. THE DEVICE WAS NO LONGER STERILE; THEREFORE, IT WAS NOT USED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21711001 ↗InjurySep 8, 2023Difficult to Remove; Deformation Due to Compressive Stress; Difficult to Remove; Deformation Due to Compressive Stress
IT WAS REPORTED THAT REMOVAL DIFFICULTIES OCCURRED RESULTING IN INCREASED PAIN. THE PATIENT WAS UNDERGOING A BILE DUCT DRAINAGE PROCEDURE. A 10F FLEXIMA DRAINAGE CATHETER WAS INSERTED. THE INNER CANNULA AND THE WIRE BECAME STUCK DURING REMOVAL. WHILE ATTEMPTING TO REMOVE THE COMPONENTS, THE CATHETER ACCORDIONED IN THE PATIENT. THE PATIENT EXPERIENCED INCREASED PAIN WHICH WAS MANAGED WITH MIDAZOLAM AND FENTANYL. ADDITIONALLY, HEMATOBILIA WAS OBSERVED DUE TO IRRITATION OF THE BILE DUCT. THE ISSUE OCCURRED WITH TWO CATHETERS. THE DEVICE WAS ABLE TO BE REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT IS IN GOOD CONDITION.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain
MAUDE report 17714320 ↗InjurySep 8, 2023Difficult to Remove; Deformation Due to Compressive Stress; Difficult to Remove; Deformation Due to Compressive Stress
IT WAS REPORTED THAT REMOVAL DIFFICULTIES OCCURRED RESULTING IN INCREASED PAIN. THE PATIENT WAS UNDERGOING A BILE DUCT DRAINAGE PROCEDURE. A 10F FLEXIMA DRAINAGE CATHETER WAS INSERTED. THE INNER CANNULA AND THE WIRE BECAME STUCK DURING REMOVAL. WHILE ATTEMPTING TO REMOVE THE COMPONENTS, THE CATHETER ACCORDIONED IN THE PATIENT. THE PATIENT EXPERIENCED INCREASED PAIN WHICH WAS MANAGED WITH MIDAZOLAM AND FENTANYL. ADDITIONALLY, HEMATOBILIA WAS OBSERVED DUE TO IRRITATION OF THE BILE DUCT. THE ISSUE OCCURRED WITH TWO CATHETERS. THE DEVICE WAS ABLE TO BE REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT IS IN GOOD CONDITION.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain
MAUDE report 17714313 ↗MalfunctionOct 2, 2019Break; Break
IT WAS REPORTED THAT THE SUTURE BROKE. UPON PREPARATION OUTSIDE THE PATIENT'S BODY, THE SUTURE OF A 10.3 FLEXIMA DRAINAGE CATHETER BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME SIZE. NO COMPLICATIONS REPORTED AND PATIENT DID WELL.
Patient problems: No Patient Involvement
MAUDE report 9145763 ↗MalfunctionSep 27, 2019Break; Break
IT WAS REPORTED THAT THE DEVICE WAS MISSING A RADIOPAQUE (RO) MARKER ON THE DRAIN. A 10.3 FLEXIMA REGULAR W/RO VTCB WAS SELECTED FOR USE FOR A BILIARY PROCEDURE. IT WAS NOTED THAT THERE WAS RO MARKER ON DRAIN. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9129349 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page