Zero Tip M0063901040Recalls & reported problems
Boston Scientific Corporation · Nitinol Stone Retrieval Basket · Class 2
All models of Zero Tip (8)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729469636), received between Feb 14, 2022 and Dec 29, 2025 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Break2
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Inflammation1
- Unspecified Infection1
- Urinary Tract Infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 29, 2025Break; Nonstandard Device; Break; Nonstandard Device
IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS USE DURING A URETEROSCOPIC LITHOTRIPSY (URSL) PROCEDURE. REPORTEDLY, THEY FOUND THAT THE BASKET WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS. HOWEVER, INVESTIGATION OF THE RETURNED DEVICE REVEALED THAT TWO BASKET WIRES BROKE AND FULLY DETACHED FROM THE BASKET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23905884 ↗InjuryAug 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NOTE: THIS REPORT PERTAINS TO ONE OF THREE DEVICES THAT WERE USED IN THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ZERO TIP BASKET, FLEXIVA PULSE LASER FIBERS, AND A DUAL LUMEN URETERAL CATHETER WERE USED IN THE LEFT KIDNEY DURING A LITHOTRIPSY PROCEDURE PERFORMED ON (B)(6) 2022. ON (B)(6) 2022, THE PATIENT EXPERIENCED PYELITIS AND CYSTITIS. THE PATIENT TOOK PERCOCET 5-325MG FOR THE PYELITIS UNTIL (B)(6) 2022. FOR CYSTITIS, THE PATIENT TOOK CEPHALEXIN 500MG AND WAS STOPPED ON (B)(6) 2022.
Patient problems: Inflammation; Unspecified Infection
MAUDE report 15199387 ↗InjuryJul 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES THAT WERE USED IN THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2022 THAT A FLEXIVA PULSE ID LASER FIBER AND A ZERO TIP BASKET WERE USED DURING A LITHOTRIPSY PROCEDURE TO TREAT STONES IN THE BLADDER, PERFORMED ON (B)(6) 2022. THE LASER UNIT USED WAS A LUMENIS PULSE 100 H LASER CONSOLE WITH LASER SETTINGS OF 0.6 JOULES, 6 HERTZ AND 4 WATTS. ON (B)(6) 2022, FIVE DAYS POST PROCEDURE, THE PATIENT EXPERIENCED CYSTITIS. SHE WAS TREATED WITH CEPHALEXIN ORAL FOR INFECTION AND PROCHLORPERAZINE IV FOR NAUSEA AND THE PATIENT'S CONDITION WAS RESOLVED.
Patient problems: Urinary Tract Infection
MAUDE report 15125302 ↗InjuryFeb 14, 2022Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ZERO TIP BASKET WAS USED IN THE URETER DURING A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, A LASER WAS USED ON THE LEFT URETERAL STONE. THE ZERO TIP BASKET WAS THEN USED TO RETRIEVED THE STONES. HOWEVER, THE BASKET WIRES BROKE INTO TWO SEPARATE PIECES WITHIN THE URETER. A DIFFERENT TYPE OF BASKET WAS USED TO SNARE THE BROKEN PIECES. THE PROCEDURE WAS DELAYED APPROXIMATELY 15 - 20 MINUTES DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13525206 ↗
FDA clearance and classification
The FDA regulates product code FFL (dislodger, stone, basket, ureteral, metal) under 21 CFR 876.4680 as a Class II device. 21 CFR 876.4680 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Zero Tip M0063901040
Has the Zero Tip M0063901040 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729469636), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Zero Tip M0063901040?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Feb 14, 2022 and Dec 29, 2025. In those reports, the most common specific device problems are “Break” (2) and “Nonstandard Device” (1); the most common patient problems are “Inflammation” (1), “Unspecified Infection” (1) and “Urinary Tract Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Zero Tip M0063901040?
None of the reports are classified as deaths, 3 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Zero Tip M0063901040 FDA approved?
Devices of this type (product code FFL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Boston Scientific Corporation.
Who makes the Zero Tip M0063901040, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Zero Tip, model M0063901040). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DISLODGER, STONE, BASKET, URETERAL, METAL devices
- Zero Tip M0063901050Boston Scientific Corporation · DISLODGER, STONE, BASKET, URETERAL, METAL286 reports
- NCircle G18778Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL222 reports
- NCircle G46206Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL195 reports
- NCircle G17520Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL127 reports
- NGage G48297Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL96 reports
- NGage G48294Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL92 reports
- NCircle G19110Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL87 reports
- Bard® SkyLite™ Tipless Nitinol Stone Basket 041900C. R. Bard, Inc. · DISLODGER, STONE, BASKET, URETERAL, METAL78 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page