WallFlex™ Biliary M00570700Recalls & reported problems

Boston Scientific Corporation · Stent System · Class 2 · Implant

All models of WallFlex Biliary (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729759294), received between Sep 28, 2021 and Jun 18, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Obstruction/Occlusion1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 18, 2025Migration; Migration

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A WALLFLEX BILIARY STENT WAS IMPLANTED IN THE DUODENUM AND COMMON BILE DUCT (CBD) TO TREAT A MALIGNANT OBSTRUCTION DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) ON (B)(6) 2024. ON (B)(6) 2025, IT WAS FOUND THAT THE STENT DISTALLY MIGRATED OUT OF THE CBD INTO THE STOMACH. THIS WAS NOTICED AS THE PATIENT PASSAGEWAY BECAME OBSTRUCTED, BUT NO SPECIFIC SYMPTOMS WERE REPORTED. THE STENT WAS RETRIEVED AND REMOVED FROM THE PATIENT BY PERFORMING AN ESOPHAGOGASTRODUODENOSCOPY COMBINED WITH AN ERCP, AND ANOTHER WALLFLEX BILIARY STENT WAS IMPLANTED. THE PATIENT ENDURE AN ENDOSCOPIC PROCEDURE TO RETRIEVE AND IMPLANT A NEW STENT.

    Patient problems: Obstruction/Occlusion

    MAUDE report 22284901 ↗
  • MalfunctionSep 28, 2021Difficult to Remove; Difficult to Remove

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2021 THAT A WALLFLEX BILIARY RX PARTIALLY COVERED STENT WAS IMPLANTED TO TREAT A MALIGNANT STRICTURE IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH METAL STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2021. REPORTEDLY, THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE STENT WAS SUCCESSFULLY DEPLOYED; HOWEVER, DURING REMOVAL OF THE DELIVERY SYSTEM, THE DELIVERY SYSTEM GOT CAUGHT WITHIN THE STENT AND IT WAS DIFFICULT TO REMOVE. THE WALLFLEX BILIARY STENT REMAINS IMPLANTED AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12544084 ↗