WallFlex™ Biliary M00570890Recalls & reported problems
Boston Scientific Corporation · Stent System · Class 2 · Implant
All models of WallFlex Biliary (23)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 08714729765332), received between Nov 17, 2016 and Jan 7, 2026 — classified as 7 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Partial Blockage” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Partial Blockage3
- Premature Activation3
- Activation Failure2
- Break2
- Defective Device1
- Detachment of Device or Device Component1
- Difficult to Remove1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions3
- Inflammation2
- Hemorrhage/Blood Loss/Bleeding1
- Jaundice1
- Nausea1
- Unspecified Infection1
- Vomiting1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 7, 2026Premature Activation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A WALLFLEX BILIARY STENT WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. IT WAS REPORTED THAT WHILE PREPARING A 10MM X 40MM WALLFLEX BILIARY UNCOVERED STENT FOR PLACEMENT, THE STENT UNINTENTIONALLY DEPLOYED DURING REMOVAL OF THE INNER SHEATH PRIOR TO BEING HANDED TO THE PHYSICIAN. THE DEVICE WAS RENDERED UNUSABLE AND WAS NOT DEPLOYED INSIDE THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23999651 ↗MalfunctionNov 18, 2022Defective Device; Defective Device
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON OCTOBER 28, 2022, THAT A WALLFLEX BILIARY RX UNCOVERED STENT WAS TO BE IMPLANTED DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, PRIOR TO STENT DEPLOYMENT, THE PHYSICIAN FELT THAT THE STENT'S LENGTH WAS NOT THE LABELED LENGTH. THE STENT APPEARED TO BE TOO LONG INSIDE THE CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15828799 ↗InjuryJan 6, 2022Activation Failure; Activation Failure
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2021 THAT A WALLFLEX BILIARY RX UNCOVERED STENT WAS IMPLANTED IN THE COMMON BILE DUCT TO TREAT A MALIGNANT, EXTRINSIC PANCREATIC MASS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT PLACEMENT PROCEDURE PERFORMED ON AN UNKNOWN DATE. REPORTEDLY, THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS NOT DILATED PRIOR TO THE STENT PLACEMENT PROCEDURE. FIVE DAYS POST STENT PLACEMENT, THE PATIENT WAS ADMITTED TO THE INTENSIVE CARE UNIT WITH CHOLANGITIS. IMAGING WAS PERFORMED, AND IT WAS NOTED THAT THE STENT DID NOT FULLY EXPAND. REINTERVENTION VIA ERCP WAS PERFORMED AND TWO PLASTIC STENTS WERE IMPLANTED WITHIN THE WALLFLEX BILIARY STENT TO ALLOW FOR BILIARY DRAINAGE. REPORTEDLY, THE PHYSICIAN WAS COMFORTABLE LEAVING THE WALLFLEX BILIARY UNCOVERED STENT IMPLANTED. THE PATIENT WAS REPORTED TO BE FULLY RECOVERED AND THE PATIENT'S CURRENT CONDITION WAS REPORTED TO BE FINE.
Patient problems: Inflammation
MAUDE report 13176714 ↗InjuryAug 26, 2021Premature Activation; Premature Activation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2021 THAT A WALLFLEX BILIARY RX UNCOVERED STENT WAS TO BE IMPLANTED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE STENT DEPLOYED PREMATURELY AND WAS UNABLE TO BE REPOSITIONED. THE STENT WAS REMOVED FROM THE PATIENT AND ANOTHER WALLFLEX BILIARY STENT WAS IMPLANTED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12375327 ↗InjuryOct 9, 2020Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON SEPTEMBER 14, 2020 THAT A WALLFLEX BILIARY UNCOVERED STENT WAS IMPLANTED TO TREAT A BILIARY OBSTRUCTION HILAR MASS IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE STENT WAS ABLE TO BE DEPLOYED; HOWEVER, DURING REMOVAL OF THE CATHETER, THE INNER CATHETER WAS STUCK IN THE BILE DUCT AND EVENTUALLY BROKE OFF. REPORTEDLY, AFTER 2 HOURS OF SEVERAL ATTEMPTS AND DIFFERENT TECHNIQUES, THE INNER CATHETER WAS REMOVED BY PULLING IT HARD AND THE STENT REMAINS IMPLANTED TO COMPLETE THE PROCEDURE. THE PATIENT WAS SENT HOME AFTER THE PROCEDURE. REPORTEDLY, THE PATIENT WAS ADMITTED TO A DIFFERENT HOSPITAL A DAY OR TWO AFTER THE PROCEDURE DUE TO A BLEED. IT WAS ALSO NOTED THAT THE PATIENT GOT PANCREATITIS. IT IS UNKNOWN IF OR HOW THE BLEEDING AND PANCREATITIS WERE TREATED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Inflammation
MAUDE report 10657374 ↗MalfunctionNov 8, 2019Premature Activation; Premature Activation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2019 THAT A WALLFLEX BILIARY RX UNCOVERED STENT WAS IMPLANTED TO TREAT A MALIGNANT STRICTURE IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2019. REPORTEDLY, THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE STENT WAS ATTEMPTED TO BE DEPLOYED STENT-IN-STENT WITHIN AND EXTENDING PROXIMAL TO A PREVIOUSLY PLACED STENT. THE TARGET ANATOMY WAS A TIGHT TURN INTO THE COMMON HEPATIC DUCT. HOWEVER, DURING DEPLOYMENT, THE STENT PREMATURELY DEPLOYED ABOVE THE PRIOR STENT, INSTEAD OF WITHIN IT. REPORTEDLY, ANOTHER WALLFLEX BILIARY STENT WAS PLACED TO CONNECT THE TWO STENTS AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9297710 ↗InjuryAug 20, 2018Activation Failure; Activation Failure
IT WAS REPORTED TO BOSTON SCIENTIFIC ON (B)(6) 2018 THAT A WALLFLEX BILIARY RX UNCOVERED STENT WAS TO BE USED DURING A PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING PROCEDURE, THE STENT DID NOT OPEN; IT REMAINED CLOSED AND WAS IMPOSSIBLE TO PLACE. THE PROCEDURE WAS COMPLETED WITH ANOTHER WALLFLEX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION AND/ OR SPECIFIC INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THESE EVENTS TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7798890 ↗MalfunctionAug 22, 2017Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2017, THAT A WALLFLEX BILIARY RX UNCOVERED STENT WAS TO BE USED TO TREAT A 2 CM MALIGNANT STRICTURE CAUSED BY PANCREATIC CANCER DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2017. REPORTEDLY, THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE PHYSICIAN TRIED TO DEPLOY THE STENT BUT THE OUTER SHEATH BROKE CLOSE TO THE GUIDEWIRE ACCESS PORT. THE STENT WAS REMOVED FROM THE PATIENT FULLY CONSTRAINED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6810681 ↗MalfunctionJul 26, 2017Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2017 THAT A WALLFLEX BILIARY RX UNCOVERED STENT WAS TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CATHETER BROKE DURING DEPLOYMENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER WALLFLEX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON AUGUST 21, 2017. THE COMPLAINANT REPORTED THAT THEY WERE REFERRING TO THE CLEAR SHEATH WHEN THEY STATED THAT THE CATHETER BROKE. THE BROKEN PIECE REMAINED ATTACHED TO THE DELIVERY SYSTEM AND THE STENT WAS FULLY CONSTRAINED WHEN IT WAS REMOVED FROM THE PATIENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6744641 ↗InjuryFeb 27, 2017Partial Blockage; Partial Blockage
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON FEBRUARY 02, 2017 THAT A WALLFLEX BILIARY UNCOVERED STENT WAS IMPLANTED ON (B)(6) 2016 AS PART OF THE (B)(6) TRIAL. THE PATIENT HAD A HISTORY OF AMPULLARY CANCER AND WAS ENROLLED INTO GROUP B PALLIATIVE TREATMENT OF BILIARY STRICTURES PRODUCED BY MALIGNANT NEOPLASMS PRIOR TO CURATIVE INTENT SURGERY; WITH OR WITHOUT NEOADJUVANT THERAPY. ON (B)(6) 2016, AN ASSESSMENT OF BILIARY OBSTRUCTIVE SYMPTOMS WAS TAKEN AND REPORTED JAUNDICE, ITCHING, DARK URINE, AND PALE STOOLS. LIVER FUNCTION TEST WAS TAKEN ON (B)(6) 2016. IMAGING (ERCP, CT, AND EUS) AND TISSUE DIAGNOSIS USING FINE NEEDLE ASPIRATION (FNA) WAS WITH A RESULT OF MALIGNANT NEOPLASM THAT IS DEEMED RESECTABLE. AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT PLACEMENT PROCEDURE WAS PERFORMED ON (B)(6) 2016. DURING THE STENT PLACEMENT PROCEDURE, A PANCREATOGRAM AND BILIARY SPHINCTEROTOMY WERE PERFORMED, AND THE STUDY STENT WAS IMPLANTED IN A SATISFACTORY MANNER. ON (B)(6) 2016, DUE TO DISEASES PROGRESSION, THE PATIENT WAS TRANSITIONED TO PALLIATIVE MANAGEMENT AND WILL NOT UNDERGO CURATIVE INTENT SURGERY. ON (B)(6) 2017, THE STUDY STENT WAS NOTED TO BE OCCLUDED DUE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6361386 ↗InjuryDec 28, 2016Partial Blockage; Partial Blockage
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON DECEMBER 06, 2016 THAT A WALLFLEX BILIARY RX UNCOVERED STENT WAS IMPLANTED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2015 AS PART OF THE (B)(6) CLINICAL TRIAL. THE PATIENT HAD A HISTORY OF PANCREATIC CANCER AND A PRIOR PLASTIC BILIARY STENT. THE PATIENT WAS ENROLLED IN THE STUDY FOR TREATMENT OF MALIGNANT NEOPLASMS WITH NO INTENDED SURGERY. ON (B)(6) 2015, AN ASSESSMENT OF BILIARY OBSTRUCTIVE SYMPTOMS (ABOS) WAS TAKEN AND REPORTED RIGHT UPPER QUADRANT PAIN. LIVER FUNCTION TESTS WERE ALSO PERFORMED ON (B)(6) 2015. THE PATIENT UNDERWENT AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), MAGNETIC RESONANCE CHOLANGIOPANCREATOGRAPHY (MRCP), CT SCAN AND ENDOSCOPIC ULTRASOUND (EUS) WITH FINE NEEDLE ASPIRATION (FNA) FOR IMAGING/TISSUE SAMPLING DURING THE STENT PLACEMENT PROCEDURE ON (B)(6) 2015. THE PROCEDURE WAS DONE IN AN INPATIENT SETTING AND A PANCREATOGRAM WAS PERFORMED DURING THE PROCEDURE. ADDITIONALLY, THE PLASTIC BILIARY STENT WAS REMOVED DURING THE PROCEDURE. THE STUDY STENT WAS IMPLANTED ON (B)(6) 2015 IN A SATISFACTORY MANNER DURING THE ERCP. ON (B
Patient problems: Jaundice; Nausea; Vomiting
MAUDE report 6210845 ↗InjuryNov 17, 2016Partial Blockage; Partial Blockage
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2016 THAT A WALLFLEX" BILIARY RX FULLY UNCOVERED STENT WAS IMPLANTED DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2016 AS PART OF (B)(6). A BASELINE ASSESSMENT OF BILIARY OBSTRUCTIVE SYMPTOMS WAS CONDUCTED AND THE FOLLOWING SYMPTOMS WERE REPORTED: DARK URINE, PALE STOOLS, AND JAUNDICE. THE PATIENT'S BASELINE KARNOFSKY SCORE WAS REPORTED AS 90, AND THE PATIENT'S WEIGHT WAS REPORTED TO BE (B)(6). BASELINE LIVER FUNCTION TESTS WERE TAKEN ON (B)(6) 2016. THE PATIENT UNDERWENT PANCREATIC PROTOCOL COMPUTED TOMOGRAPHY (CT), ENDOSCOPIC ULTRASOUND (EUS), CHEST X-RAY, AND EUS FINE NEEDLE ASPIRATION (FNA), AND A STAGE IIA T3 -N0 M0, A 4.3 CM ADENOCARCINOMA TUMOR WAS IDENTIFIED AT THE HEAD AND BODY OF THE PANCREAS. ON (B)(6) 2016, THE PATIENT UNDERWENT STUDY STENT PLACEMENT AS AN OUTPATIENT PROCEDURE. A SPHINCTEROTOMY WAS PERFORMED DURING THIS PROCEDURE, AND THE 10MM X 40MM WALLFLEX BILIARY UNCOVERED STENT WAS PLACED IN A SATISFACTORY MANNER. ON (B)(6) 2016, THE PATIENT EXPERIENCED A SERIOUS ADVERSE EVENT OF CHOLANGITIS. ACCORDING TO THE COMPLAINANT, THE EVENT IS RELATED TO THE STUDY STENT AND WAS NOTED TO BE SEVERE. ON (B)
Patient problems: Unspecified Infection
MAUDE report 6110595 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page