Coyote™ ES H74939134254010Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Coyote ES (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Coyote ES is a balloon catheter used to open narrowed arteries in the legs. There are 24 adverse event reports naming it, with 1 injury and no deaths. Almost all reports describe the balloon rupturing during inflation, usually in heavily calcified vessels, with procedures typically finished using a different device; one report mentions a fragment left in the patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 08714729767381), received between Aug 21, 2018 and Apr 20, 2026 — classified as 1 injury, 23 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (21 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture21
- Break1
- Detachment of Device or Device Component1
- Difficult to Remove1
- Entrapment of Device1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions17
- No Consequences Or Impact To Patient6
- Foreign Body In Patient1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 20, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 8 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM USING THE NORMAL METHOD, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24923370 ↗MalfunctionMar 30, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MILD TORTUOUS AND SEVERELY CALCIFIED VESSEL RIGHT LEG. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON SECOND INFLATION AT 10 ATMOSPHERES FOR 5 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY DEVICE FRAGMENTS LEFT, AND THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24719378 ↗MalfunctionFeb 20, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED BELOW THE KNEE VESSEL. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE FIRST INFLATION AT 9 ATMOSPHERES FOR 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24406203 ↗MalfunctionOct 21, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. A 2.5 MM X 40 MM X 145 CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED. NO ADDITIONAL INFORMATION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23346531 ↗MalfunctionJun 4, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON BURST EVEN WITHOUT REACHING THE BURST RATED PRESSURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22142136 ↗MalfunctionApr 2, 2025Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT THE BALLOON BECAME STUCK WITH THE WIRE. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED PERIPHERAL ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON BECAME STUCK WITH THE NON-BOSTON SCIENTIFIC WIRE BEFORE IT ENTERED THE PATIENT'S BODY. THE DEVICE WAS REMOVED TOGETHER WITH THE WIRE AND THE GUIDEWIRE WAS ABLE TO BE REMOVED FROM THE BALLOON ONCE OUTSIDE THE PATIENT, AND THE PROCEDURE WAS COMPLETED USING AN ALTERNATE DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21751930 ↗MalfunctionMar 11, 2025Leak/Splash; Leak/Splash
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 26FEB2025. IT WAS REPORTED THAT SHAFT LEAK OCCURRED. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING PREPARATION, THE FLUID WAS LEAKING THROUGH THE BALLOON PORT AND SHAFT LEAK WAS NOTED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING ORIGINAL METHOD AND/OR DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED A PINHOLE 6MM FROM THE TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21572357 ↗MalfunctionJan 29, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON INITIAL INFLATION AT 14 ATMOSPHERES FOR 5 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEMS, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21270187 ↗MalfunctionMar 24, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON BURST OCCURRED. THE PATIENT PRESENTED WITH A PERIPHERAL ARTERIAL DISEASE AND UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND CALCIFIED ANTERIOR TIBIAL ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, UPON THE FIRST INFLATION AT 6 ATMOSPHERES FOR 3 SECONDS, THE BALLOON BURST. THE DEVICE WAS REMOVED SUCCESSFULLY AND THERE WERE NO DEVICE FRAGMENTS LEFT INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18966458 ↗MalfunctionFeb 6, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. THE PATIENT PRESENTED WITH PERIPHERAL ARTERIAL DISEASE AND UNDERWENT ANGIOGRAM OF THE 90% STENOSED TARGET LESION. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS SELECTED FOR USE. HOWEVER, DURING PREPARATION, THE BALLOON CATHETER PULLED APART COMING OUT OF THE HOOP. THE DETACHED DISTAL TIP WAS INSIDE THE PROTECTIVE SHEATH AND WAS NOT INSERTED INTO THE PATIENT. THE DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16303484 ↗MalfunctionSep 5, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT NOMINAL PRESSURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15355853 ↗MalfunctionJun 3, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT PRESENTED WITH CHRONIC CRITICAL LIMB ISCHEMIA (CLI). THE 99% STENOSED, TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED TIBIALIS ANTERIOR ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 8 ATMOSPHERES FOR 15 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT WAS IN GOOD CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14588397 ↗MalfunctionMar 30, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 8 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED FROM THE PATIENT WITHOUT ANY PROBLEM AND A PINHOLE WAS NOTED. THE PROCEDURE WAS COMPLETED ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT STATUS WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13957366 ↗MalfunctionDec 2, 2021Break; Material Rupture; Break; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE AND FRACTURE OCCURRED. THE PATIENT PRESENTED WITH PERIPHERAL ARTERIAL DISEASE. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED DISTAL POSTERIOR TIBIAL. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, ON THE FIRST INFLATION AT 6 ATMOSPHERES FOR FEW SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND REPLACED WITH A 2MM X 40MM X 144CM COYOTE ES BALLOON CATHETER, WHICH ALSO RUPTURED DURING INITIAL INFLATION AT 6 ATMOSPHERES FOR FEW SECONDS. IT WAS ALSO NOTICED THAT THE DEVICES WERE ALSO FRACTURED. BOTH DEVICES WERE REMOVED INTACT AND THE PROCEDURE WAS COMPLETED WITH A 2.0 X 150 COYOTE BALLOON CATHETER. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED. IT WAS FURTHER REPORTED THAT THE BALLOON WAS RUPTURED BUT THE CATHETERS WERE NOT NOTED TO BE FRACTURED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12920634 ↗InjuryMay 19, 2021Difficult to Remove; Material Rupture; Difficult to Remove; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED AND REMOVAL DIFFICULTIES WERE ENCOUNTERED. THE 90% STENOSED TAGET LESION WAS LOCATED IN THE NON-TORTUOUS AND SEVERELY CALCIFIED TIBIAL PERITONEAL TRUNK. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 3-4 ATMOSPHERES, THE BALLOON RUPTURED AND A PART OF BALLOON COULD NOT BE REMOVED. THE UNRETRIEVED PART WAS ATTEMPTED TO BE SNARED AND ANOTHER BALLOON WAS PLACED TO DRAG THE PIECE BUT FAILED TO BE REMOVED. A STENT WAS PLACED AND PINNED THE BALLOON PIECE AGAINST THE VESSEL WALL AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 11852704 ↗MalfunctionApr 19, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TOTALLY OCCLUDED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT POPLITEAL TIBIAL ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT 12 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER COYOTE ES BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT FULLY RECOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11689321 ↗MalfunctionFeb 23, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND HEAVILY CALCIFIED LEFT POSTERIOR TIBIAL ARTERY WITH NO SIGNIFICANT BEND. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER BALLOON AT THE SAME SIZE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED. IT WAS FURTHER REPORTED THAT PERCENTAGE OF STENOSIS WAS A 99% AND NOT 90% AS PREVIOUSLY REPORTED. THERE WERE NO SIGNIFICANT BENDS NOTED IN THE LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11369576 ↗MalfunctionFeb 10, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. VASCULAR ACCESS WAS OBTAINED UTILIZING AN IPSILATERAL ANTEGRADE APPROACH. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL BELOW THE KNEE. AFTER THE LESION WERE CROSSED WITH A NON-BSC GUIDEWIRE AND MICROCATHETER WITH THE SUPPORT OF A NON-BSC GUIDE CATHETER, A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 8 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A 3MM COYOTE ES BALLOON CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11306754 ↗MalfunctionDec 15, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED, TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED BELOW THE KNEE ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING POBA AT NOMINAL PRESSURE, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11008350 ↗MalfunctionJun 9, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10135869 ↗MalfunctionMay 3, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 60% STENOSED TARGET LESION WAS LOCATED IN THE PULMONARY ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, WHEN THE BALLOON WAS INFLATED AT 12 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8578201 ↗MalfunctionMar 18, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE CHRONIC TOTAL OCCLUSION (CTO) STENOSED TARGET LESION WAS LOCATED IN THE HEAVILY CALCIFIED NON TORTUOUS PERINEAL ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED OVER THE WIRE, IN WHOLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER WITH SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8428849 ↗MalfunctionAug 31, 2018Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED DISTAL ANTERIOR TIBIAL ARTERY. AFTER CROSSING A GUIDE WIRE, A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 8 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7840252 ↗MalfunctionAug 21, 2018Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 2.5MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION; HOWEVER, UPON INFLATION THE BALLOON RUPTURED DUE TO CALCIFICATION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7803436 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026