Coyote™ ES H74939134302010Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Coyote ES (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 08714729767398), received between Nov 1, 2018 and Dec 8, 2025 — classified as 1 injury, 15 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture15
- Deflation Problem1
- Difficult to Advance1
- Difficult to Remove1
- Material Integrity Problem1
Patient problems
- No Consequences Or Impact To Patient9
- No Clinical Signs, Symptoms or Conditions6
- No Patient Involvement1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 8, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 3 SECONDS, THE BALLOON RUPTURED IN A PINHOLE FORM. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23739417 ↗MalfunctionApr 17, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT ANTERIOR TIBIAL ARTERY. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 10 ATMOSPHERES FOR 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19125128 ↗MalfunctionSep 19, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING FIRST INFLATION AT 10 ATMOSPHERES FOR FEW SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17778087 ↗MalfunctionMay 26, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL BELOW THE KNEE. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE SECOND INFLATION AT 6 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14518118 ↗MalfunctionDec 29, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT THE BALLOON RUPTURED. A 3MM X 20MM X 143CM COYOTE BALLOON CATHETER WAS SELECTED FOR USE IN A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY (SFA). THE TARGET LESION WAS 100% OCCLUDED, MODERATELY TORTUOUS AND SEVERELY CALCIFIED. THE FIRST INFLATION WAS 10 SECONDS IN DURATION AT 8ATM. HOWEVER, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13109111 ↗InjuryApr 8, 2021Deflation Problem; Difficult to Remove; Difficult to Advance; Material Integrity Problem; Deflation Problem; Difficult to Remove; Difficult to Advance; Material Integrity Problem
IT WAS REPORTED THAT DEFLATION FAILURE OCCURRED. THE 60-70% STENOSED TARGET LESION WAS LOCATED IN THE NON-CALCIFIED VESSEL OF THE LOWER EXTREMITY. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. THE DEVICE HAD DIFFICULTY ADVANCING OVER THE WIRE UPON INTRODUCTION INTO THE VESSEL. THE DEVICE WAS REMOVED, AND THE WIRE PORT WAS FLUSHED. THE BALLOON WAS RE-INTRODUCED AND WAS ABLE TO CROSS THE LESION. THE BALLOON WAS INFLATED AT 14 ATMOSPHERES FOR 2 MINUTES. THE DEVICE WAS THEN DEFLATED, HOWEVER, IT FAILED TO DEFLATE DESPITE MULTIPLE ATTEMPTS TO APPLY NEGATIVE PRESSURE WITH BOTH A BALLOON INFLATOR DEVICE AND A LARGE SYRINGE. THE PHYSICIAN RETRACTED THE BALLOON THROUGH THE VESSELS AND THE SHEATH WHILE IT WAS STILL INFLATED. THE BALLOON ENCOUNTERED SIGNIFICANT RESISTANCE AT THE SHEATH BUT WAS ABLE TO BE REMOVED. UPON REMOVAL, THE BALLOON SEEMED TO BE INVERTED ON ITSELF, MOST LIKELY DUE TO THE FORCE OF PULLING IT THROUGH THE SHEATH WHILE STILL INFLATED. THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS, AND THERE WAS NO COMPLICATIONS OR FURTHER INTERVENTION REQUIRED. THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11634479 ↗MalfunctionDec 31, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED POPLITEAL ARTERY. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE SECOND INFLATION AT 8 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS IN GOOD CONDITION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11100052 ↗MalfunctionNov 30, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, ON INITIAL INFLATION AT 8 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10920831 ↗MalfunctionNov 18, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, ON INITIAL INFLATION AT 14 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT PROBLEM AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10862160 ↗MalfunctionAug 10, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL BELOW THE KNEE. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 14 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10388689 ↗MalfunctionApr 8, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 3MM X 20MM X 143 CM COYOTE ES BALLOON CATHETER WAS SELECTED FOR USE FOR A PERIPHERAL LOWER EXTREMITY STENT PROCEDURE IN A MODERATELY CALCIFIED, MODERATELY TORTUOUS AND 90% STENOSED LESION IN THE LEFT LEG. DURING THE PROCEDURE, THE BALLOON WAS INFLATED AND RUPTURED AT 1 ATMOSPHERE. THE DEVICE WAS REMOVED COMPLETELY, LEAVING NO REMAINING PARTS WITHIN THE PATIENT, AND ANOTHER OF THE SAME BALLOON WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9934432 ↗MalfunctionFeb 1, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING SECOND INFLATION AT 6 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9656515 ↗MalfunctionOct 30, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON TEAR WAS ENCOUNTERED. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS SELECTED FOR USE. HOWEVER, UPON OPENING THE PACKAGE, IT WAS NOTICED THAT THE BALLOON HAD A TEAR. THE PROCEDURE WAS COMPLETED WITH A 3X30 COYOTE BALLOON CATHETER. NO FURTHER COMPLICATIONS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 9256789 ↗MalfunctionAug 13, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, UPON INITIAL INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8891130 ↗MalfunctionJan 16, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERLY CALCIFIED RIGHT ANTERIOR TIBIAL ARTERY. A 3MM X 20MM X 143CM CAYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 4 ATMOSPHERE FOR 6 SECONDS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8253074 ↗MalfunctionNov 1, 2018Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND NON CALCIFIED VESSEL. AFTER A GUIDE WIRE CROSSED THE LESION, A 3MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR PRE-DILATION. HOWEVER, DURING THE FIRST INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER COYOTES ES BALLOON CATHETER. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8028699 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page