Skip to content

iLab Ultrasound Imaging System H749ILAB240CART0Recalls & reported problems

Boston Scientific Corporation · Cart System · Class 2

All models of iLab Ultrasound Imaging System (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729777694), received between Jan 31, 2024 and Mar 18, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Application Program Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Problem1
  • Computer Operating System Problem1
  • Computer Software Problem1
  • Failure to Power Up1
  • Poor Quality Image1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2025Computer Operating System Problem; Computer Operating System Problem

    IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. A CAPITAL EQUIPMENT ILAB ULTRASOUND IMAGING SYSTEM WAS SELECTED FOR USE. DURING PROCEDURE, IMAGING FAILED TO START ERROR OCCURRED. THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21630910 ↗
  • MalfunctionJun 25, 2024Computer Software Problem; Failure to Power Up; Computer Software Problem; Failure to Power Up

    IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. A CAPITAL EQUIPMENT ILAB ULTRASOUND IMAGING SYSTEM WAS SELECTED FOR USE. DURING PROCEDURE, ERROR 1002 OCCURRED AND ACQUISITION PC DID NOT RESTART AFTER RESET. THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19608606 ↗
  • MalfunctionJan 31, 2024Poor Quality Image; Application Program Problem; Poor Quality Image; Application Program Problem

    IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. A CAPITAL EQUIPMENT ILAB ULTRASOUND IMAGING SYSTEM WAS SELECTED FOR USE. DURING PROCEDURE, A FIRST RUN WAS SUCCESSFULLY PERFORMED. WHEN A SECOND WAS ATTEMPTED, NO IMAGE WAS DISPLAYED. THE MDU5+ COMPONENT SEEMED TO WORK BUT THE IMAGE WAS BLACK. THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18614851 ↗

Other COMPUTER, DIAGNOSTIC, PROGRAMMABLE devices

All COMPUTER, DIAGNOSTIC, PROGRAMMABLE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page