iLab Ultrasound Imaging System H749ILAB220CART0Recalls & reported problems
Boston Scientific Corporation · Cart System · Class 2
All models of iLab Ultrasound Imaging System (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 08714729780205), received between Jul 15, 2020 and Dec 12, 2025 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Application Program Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Application Program Problem2
- Device Displays Incorrect Message2
- No Display/Image2
- Application Network Problem1
- Complete Loss of Power1
- Computer Operating System Problem1
- Computer Software Problem1
- Image Display Error/Artifact1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2025Application Program Problem; Application Program Problem
IT WAS REPORTED THAT A DEVICE ISSUE OCCURRED RESULTING IN AN ABORTED PROCEDURE. A CAPITAL EQUIPMENT ILAB ULTRASOUND IMAGING SYSTEM WAS SELECTED FOR USE. DURING THE PROCEDURE, THE ACQUISITION PROCESSOR (ACQ PC) FAILED TO START. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23790970 ↗MalfunctionDec 1, 2025Application Program Problem; Application Program Problem
IT WAS REPORTED THAT A DEVICE ISSUE OCCURRED, RESULTING IN AN ABORTED PROCEDURE. AN ILAB ULTRASOUND IMAGING SYSTEM WAS USED FOR INTRAVASCULAR ULTRASOUND (IVUS) EXAMINATION OF THE TARGET LESION. DURING PREPARATION PRIOR TO THE IVUS PROCEDURE, IT WAS FOUND THAT THE ACQUISITION PROCESSOR (ACQ PC) COULD NOT START UP. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23684977 ↗MalfunctionOct 22, 2025Computer Operating System Problem; Computer Operating System Problem
IT WAS REPORTED THAT A DEVICE ISSUE OCCURRED, RESULTING IN AN ABORTED PROCEDURE. AN ILAB ULTRASOUND IMAGING SYSTEM WAS USED FOR INTRAVASCULAR ULTRASOUND (IVUS) EXAMINATION OF THE TARGET LESION. DURING THE IVUS PROCEDURE, THE DEVICE COULD NOT BE STARTED. THE PROCEDURE WAS NOT COMPLETED AND WAS RESCHEDULED. NO PATIENT COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED THAT ONLY THE IMAGING PROCEDURE WAS CANCELLED DUE TO THE ILAB ISSUE. THE PATIENT DID NOT RECEIVE THE PLANNED TREATMENT, AND THE PROCEDURE WAS CANCELLED BEFORE THE PATIENT WAS SEDATED. LOCAL ANESTHESIA HAD BEEN PLANNED, BUT WAS NOT ADMINISTERED SINCE THE PROCEDURE WAS CANCELLED BEFOREHAND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23358616 ↗MalfunctionOct 21, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. A CAPITAL EQUIPMENT ILAB ULTRASOUND IMAGING SYSTEM WAS SELECTED FOR USE. DURING THE PROCEDURE, THE SCREEN WENT BLACK. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23345521 ↗MalfunctionOct 15, 2025Operating System Becomes Nonfunctional; Operating System Becomes Nonfunctional
IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. A CAPITAL EQUIPMENT ILAB ULTRASOUND IMAGING SYSTEM WAS SELECTED FOR USE. DURING PREPARATION, THE SCREEN TURNED BLUE WHEN THE DEVICE WAS POWERED ON. THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23279279 ↗MalfunctionOct 15, 2025Image Display Error/Artifact; Device Displays Incorrect Message; Complete Loss of Power; Image Display Error/Artifact; Device Displays Incorrect Message; Complete Loss of Power
IT WAS REPORTED THAT A DEVICE ISSUE OCCURRED RESULTING IN AN ABORTED PROCEDURE. A CE ILAB WHICH IS USED FOR ULTRASOUND EXAMINATION OF A TARGET LESION. DURING THE PROCEDURE, THE CATHETER DISPLAYED A BLACK SCREEN AFTER INSERTION. UPON REPLACEMENT WITH ANOTHER CATHETER, AN ERROR MESSAGE APPEARED. THE PROCEDURE COULD NOT BE COMPLETED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23279242 ↗MalfunctionJul 31, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. THE CE ILAB WAS SELECTED FOR EXAMINATION OF THE TARGET LESION. DURING THE PROCEDURE, IT WAS FOUND THAT THE SCREEN HAD NO DISPLAY. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED, AND THE PATIENT STATUS WAS STABLE. IT WAS FURTHER REPORTED THAT ONLY THE IMAGING PROCEDURE WAS CANCELLED BUT THE TREATMENT PROCEDURE WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22674116 ↗MalfunctionApr 25, 2025Computer Software Problem; Computer Software Problem
IT WAS REPORTED THAT A DEVICE ISSUE OCCURRED RESULTING IN AN ABORTED PROCEDURE. A CE ILAB WAS SELECTED FOR THE ULTRASOUND EXAMINATION OF THE TARGET LESION. DURING THE IVUS PROCEDURE, THE DEVICE AUTOMATICALLY RESTARTED. THE PROCEDURE WAS NOT COMPLETED. THERE WERE NO PATIENT COMPLICATIONS, AND THE PATIENT WAS STABLE. IT WAS FURTHER REPORTED THAT ONLY THE IMAGING PROCEDURE WAS CANCELLED, AND THAT THE PATIENT HAD A STENT IMPLANTED AS PLANNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21907999 ↗MalfunctionJan 8, 2025Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT A DEVICE ISSUE OCCURRED, RESULTING IN AN ABORTED PROCEDURE. AN ILAB ULTRASOUND IMAGING SYSTEM WAS USED FOR THE INTRAVASCULAR ULTRASOUND (IVUS) EXAMINATION OF THE TARGET LESION. DURING PREPARATION, THE PHYSICIAN FOUND THAT THE IVUS CATHETER COULD NOT BE SEARCHED. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS ISSUE AND WAS RESCHEDULED. THE PATIENT WAS STABLE, AND NO COMPLICATIONS OR ADDITIONAL INTERVENTIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21099073 ↗MalfunctionMar 26, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT A CAPITAL EQUIPMENT ISSUE OCCURRED. A CE ILAB, WHICH IS USED FOR ULTRASOUND EXAMINATION OF A TARGET LESION, WAS UNDERGOING ROUTINE INSPECTION. DURING THIS ACTIVITY, IT WAS NOTED THAT A WHEEL HAD FALLEN OFF. THERE WAS NO PROCEDURE OR PATIENT INVOLVED. THE ISSUE WAS RESOLVED ON SITE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18980833 ↗MalfunctionJul 15, 2020Application Network Problem; Application Network Problem
IT WAS REPORTED THAT NETWORK FAILURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE CORONARY ARTERY. A 220V CE ILAB ULTRASOUND IMAGING SYSTEM WAS USED. PRIOR THE PROCEDURE, THE PHYSICIAN FOUND A NETWORK FAILURE IN THE EQUIPMENT SYSTEM. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THE PATIENT IS STABLE, NO PATIENT COMPLICATION OR OTHER ADDITIONAL INTERVENTION REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10277156 ↗
Other COMPUTER, DIAGNOSTIC, PROGRAMMABLE devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026