Skip to content

Advanix™ Biliary M00532260Recalls & reported problems

Boston Scientific Corporation · Biliary Stent · Class 2 · Implant

All models of Advanix Biliary (59)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729787037), received between May 9, 2019 and Feb 24, 2026 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Migration” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration3
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Material Deformation1

Patient problems

  • No Consequences Or Impact To Patient2
  • Perforation2
  • Hemorrhage/Blood Loss/Bleeding1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 24, 2026Migration

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX STENT WAS IMPLANTED IN THE COMMON BILE DUCT TO TREAT CHOLEDOCHOLITHIASIS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (b)(6) 2025.ON (b)(6) 2026; DURING THE SCHEDULED EXPLANT PROCEDURE, IT WAS NOTED THAT THE ADVANIX STENT HAD MIGRATED BACK INTO THE STOMACH, AND IT WAS EXTRACTED USING AN STANDARD GASTROSCOPE AND AN ENDOSCOPIC SNARE. THIS WAS PERFORMED WITHOUT ANY DIFFICULTY.THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24436786 ↗
  • InjuryOct 13, 2023Migration; Migration

    IT WAS REPORTED TO BOSTON SCIENTIFIC THAT AN ADVANIX BILIARY STENT WAS IMPLANTED DURING A PROCEDURE IN THE DUODENUM. THE PROCEDURE DATE IS UNKNOWN. DURING AN EMERGENT FOLLOW-UP ON (B)(6) 2023, COMPUTED TOMOGRAPHY (CT) SCAN WAS PERFORMED AND IT WAS FOUND THAT THE STENT MIGRATED, AND ONE SIDE OF THE PIGTAIL PERFORATED THROUGH THE DUODENUM INTO THE PERITONEUM. THE PERFORATED DUODENUM WAS CLOSED USING A BOSTON SCIENTIFIC MANTIS CLIP DEVICE AND THE ADVANIX BILIARY STENT WAS RETRIEVED USING RAT TOOTH FORCEPS WHICH SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WAS NO NEW STENT IMPLANTED TO REPLACE THE EXPLANTED ADVANIX BILIARY STENT. IT WAS REPORTED THAT THE PATIENT HAS EXPERIENCED SYMPTOMS CAUSING THE EMERGENT FOLLOW UP. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: Perforation

    MAUDE report 17932332 ↗
  • InjuryOct 16, 2019Material Deformation; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PREVIOUSLY IMPLANTED ADVANIX BILIARY DOUBLE PIGTAIL STENT WAS REMOVED FROM A PATIENT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE COMMON BILE DUCT (CBD) PERFORMED ON (B)(6) 2019. THE EXACT IMPLANT DATE WAS NOT REPORTED. ON (B)(6) 2019, THE REPEAT ERCP PROCEDURE WAS PERFORMED AS PLANNED TO REMOVE THE STENT THAT WAS IMPLANTED IN THE COMMON BILE DUCT. PRIOR TO REMOVAL OF THE STENT, AN X-RAY WAS PERFORMED AND CONFIRMED THAT THE ADVANIX BILIARY DOUBLE PIGTAIL STENT WAS IN ITS EXPECTED PIGTAIL SHAPE IN THE CBD. ACCORDING TO THE COMPLAINANT, DURING THE PLANNED STENT REMOVAL PROCEDURE, THE ADVANIX BILIARY STENT WAS REMOVED FROM THE PATIENT'S DUCT USING RAT TOOTH FORCEPS WITHOUT ANY ISSUES. AFTER REMOVING THE STENT FROM THE DUCT, THE STENT COULD NOT BE RETRACTED THROUGH THE BIOPSY CHANNEL OF THE ENDOSCOPE. IT WAS NOTED BY THE PHYSICIAN THAT THE STENT APPEARED TO HAVE BECOME KINKED IN THREE PLACES ON THE BODY OF THE STENT DURING ATTEMPTED REMOVAL THROUGH THE SCOPE, MAKING THE REMOVAL THROUGH THE SCOPE VERY DIFFICULT. THE PHYSICIAN DECIDED TO REMOVE THE SCOPE ENTIRELY FROM THE PATIENT IN ORDER TO REMOVE THE S

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 9197785 ↗
  • InjuryOct 8, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PREVIOUSLY IMPLANTED ADVANIX BILIARY STENT WAS REMOVED FROM A PATIENT DURING A PLANNED STENT REMOVAL PROCEDURE IN THE COMMON BILE DUCT, PERFORMED ON APPROXIMATELY TWO TO THREE WEEKS FOLLOWING THE STENT PLACEMENT PROCEDURE. THE STENT WAS IMPLANTED ON (B)(6) 2019 WITHOUT ANY ISSUES REPORTED. ACCORDING TO THE COMPLAINANT, APPROXIMATELY TWO TO THREE WEEKS FOLLOWING THE STENT PLACEMENT PROCEDURE, A PLANNED STENT REMOVAL WAS PERFORMED. THE PHYSICIAN NOTICED THAT THE STENT PERFORATED THE COMMON BILE DUCT OF THE PATIENT. THE STENT WAS RETRIEVED USING FORCEPS AND THE PROCEDURE WAS COMPLETED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. **CORRECTED INFORMATION RECEIVED ON 01OCT2019** THE STENT WAS IMPLANTED ON (B)(6) 2019 WITHOUT ANY ISSUES REPORTED. THE STENT WAS RETRIEVED USING A SNARE AND THE PROCEDURE WAS COMPLETED.

    Patient problems: Perforation

    MAUDE report 9167216 ↗
  • MalfunctionJul 25, 2019Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY IN THE MIDDLE PART OF THE COMMON BILE DUCT PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE MIDDLE PART OF THE STENT WAS FRACTURED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DIFFERENT MODEL OF THE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8826216 ↗
  • MalfunctionMay 9, 2019Migration; Migration

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE COMMON BILE DUCT, PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, POST PROCEDURE, THE PATIENT WAS SEEN BY THE PHYSICIAN FOR A PLANNED STENT REMOVAL ON (B)(6) 2019. IT WAS NOTICED UNDER A FLOUROSCOPY THAT THE STENT MIGRATED DOWN TO THE GASTROINTESTINAL TRACT. THERE WAS NO SERIOUS INJURY, NOR WERE THERE ANY ADVERSE PATIENT EFFECTS REPORTED AS A RESULT OF THIS EVENT. THE CONDITION OF THE PATIENT AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8595776 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page