Advanix™ Biliary M00534570Recalls & reported problems

Boston Scientific Corporation · Biliary Stent with NaviFlexTM RX Delivery System · Class 2 · Implant

All models of Advanix Biliary (59)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729787129), received between Mar 23, 2022 and Feb 26, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Difficult or Delayed Positioning1
  • Entrapment of Device1
  • Migration1
  • Retraction Problem1
  • Unsealed Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Pain1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 26, 2026Break; Difficult or Delayed Positioning; Entrapment of Device; Retraction Problem

    IT WAS REPORTED THAT AN ADVANIX STENT WITH NAVIFLEX DELIVERY SYSTEM WAS TO BE IMPLANTED IN THE DISTAL COMMON BILE DUCT TO TREAT CHOLANGITIS CAUSED BY A STRICTURE DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026 DURING ATTEMPTED DEPLOYMENT, THE HANDLE WAS UNLOCKED AND THE BLUE PULL WIRE WAS RETRACTED, ALONG WITH REMOVAL OF THE GUIDEWIRE. HOWEVER, THE STENT DID NOT DEPLOY AND REMAINED ATTACHED TO THE DELIVERY SYSTEM. THE SUTURE REMAINED INTACT, AND FURTHER RETRACTION OF THE BLUE PULL WIRE WAS NOT POSSIBLE. TO PREVENT POTENTIAL STENT MIGRATION FROM THE BILE DUCT, THE ENDOSCOPE ELEVATOR WAS USED TO STABILIZE THE STENT WHILE THE DELIVERY SYSTEM WAS WITHDRAWN. DURING WITHDRAWAL, THE BLUE PULL WIRE FRACTURED. UPON REMOVAL OF THE WHITE DELIVERY SYSTEM, IT WAS OBSERVED THAT THE BLACK INNER CATHETER REMAINED WITHIN THE LUMEN OF THE STENT, PREVENTING DISENGAGEMENT OF THE SUTURE AND SEPARATION OF THE STENT FROM THE DELIVERY SYSTEM. GRASPING FORCEPS WERE USED TO MANUALLY RETRIEVE THE BLACK INNER CATHETER WHILE MAINTAINING THE STENT POSITION WITHIN THE BILE DUCT. THE BLACK INNER CATHETER WAS SUCCESSFULLY REMOVED; HOWEVER, IT SUBSEQUENTLY BECAME L

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24457830 ↗
  • InjuryJan 29, 2026Migration

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX-NAVIFLEX RX BILIARY STENT WAS IMPLANTED IN THE LEFT INTRAHEPATIC DUCT (IHD) TO TREAT JAUNDICE DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. THERE WAS NO SERIOUS INJURY REPORTED AND THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE. SEVERAL DAYS AFTER THE INITIAL IMPLANTATION, THE PATIENT EXPERIENCED PAIN, AND THE PATIENT'S BILIRUBIN LEVELS DID NOT DECREASE AS EXPECTED. ON (B)(6) 2025, A SUBSEQUENT ERCP WAS PERFORMED, WHICH CONFIRMED DISTAL STENT MIGRATION. THE ORIGINAL STENT WAS REMOVED AND REPLACED WITH A NON-BOSTON SCIENTIFIC STENT. NO EXTENDED OBSERVATION PERIOD WAS REQUIRED, AND THE PATIENT'S CURRENT CONDITION WAS REPORTED TO BE FINE.

    Patient problems: Pain

    MAUDE report 24218181 ↗
  • MalfunctionSep 19, 2024Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY RX STENT WAS TO BE USED IN THE BILE DUCT TO TREAT STONES DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE (ERCP) PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, IT WAS NOTED THAT THERE WAS A TEAR IN THE INNER STERILE POUCH, AND THE STERILITY OF THE DEVICE WAS COMPROMISED. ANOTHER ADVANIX BILIARY STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20261740 ↗
  • MalfunctionMar 23, 2022Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING A PROCEDURE. THE EXACT EVENT DATE WAS NOT PROVIDED. IT WAS REPORTED THAT THE STENT BECAME BROKEN FROM THE TIP OF THE DUODENOSCOPE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13869264 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices