Advanix™ Biliary M00534580Recalls & reported problems

Boston Scientific Corporation · Biliary Stent with NaviFlexTM RX Delivery System · Class 2 · Implant

All models of Advanix Biliary (59)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729787136), received between Jul 2, 2024 and Nov 25, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Material Deformation2
  • Activation, Positioning or Separation Problem1
  • Difficult to Remove1
  • Material Split, Cut or Torn1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 25, 2024Break; Difficult to Remove; Use of Device Problem; Material Deformation; Break; Difficult to Remove; Use of Device Problem; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY PRELOADED CENTER BEND STENT WAS IMPLANTED IN THE BILE DUCT TO TREAT A STRICTURE DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON AN UNKNOWN DATE. POST-STENT PLACEMENT, ON (B)(6) 2024, AN ATTEMPT WAS MADE TO REMOVE THE STENT USING A SNARE AND GRASPING FORCEPS. HOWEVER, THE END OF THE STENT BROKE INTO TWO PIECES AND THE STENT WAS UNABLE TO BE REMOVED. THE BROKEN PORTION BECAME STRETCHED AND STUCK IN THE DUCT. SUBSEQUENTLY, A SECOND STENT WAS PLACED, AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: NOTE: IT WAS REPORTED THAT THE GUIDEWIRE WAS INSIDE THE PATIENT DURING THE ATTEMPTED DEPLOYMENT. HOWEVER, THE ADVANIX BILIARY STENT WITH NAVIFLEX RX DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU) STATES, "FOR PRELOADED SYSTEM ONLY: IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE ENDOSCOPE, THE STENT CAN NOT BE FULLY DEPLOYED." THE PHYSICIAN DID NOT FOLLOW THE STEPS CITED IN THE IFU. A PHOTO OF THE DEVICE WAS PROVIDED, WHICH SHOWED THE STENT WAS BROKEN AND DAMAGED. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD-F

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20774077 ↗
  • InjuryJul 2, 2024Break; Activation, Positioning or Separation Problem; Material Deformation; Material Split, Cut or Torn; Break; Activation, Positioning or Separation Problem; Material Deformation; Material Split, Cut or Torn

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY PRELOADED CENTER BEND STENT WAS USED IN THE BILIARY DUCT TO TREAT STENOSIS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE STENT WAS UNABLE TO SEPARATE FROM THE INTRODUCER. SUBSEQUENTLY, THE INTRODUCER BROKE OFF WHEN IT WAS REMOVED AND WAS RETRIEVED. THE PROCEDURE WAS COMPLETED WITH ANOTHER ADVANIX BILIARY PRELOADED STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19656249 ↗