Advanix™ Biliary M00534640Recalls & reported problems

Boston Scientific Corporation · Biliary Stent with NaviFlexTM RX Delivery System · Class 2 · Implant

All models of Advanix Biliary (59)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729787198), received between Nov 20, 2018 and Jan 9, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Adverse Event Without Identified Device or Use Problem1
  • Difficult or Delayed Positioning1
  • Positioning Failure1
  • Unsealed Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 9, 2026Break; Difficult or Delayed Positioning

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX RX BILIARY PRELOADED STENT WAS IMPLANTED IN THE BILE DUCT TO TREAT STONES/STRICTURE DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE (ERCP) PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE INNER BLACK CATHETER COULD NOT BE REMOVED FULLY TO RELEASE THE STENT. THE CATHETER DETACHED FROM THE PULL WIRE. HOWEVER, THE PHYSICIAN WAS ABLE TO DEPLOY THE STENT BY REMOVING THE INNER CATHETER WITH RETRIEVAL FORCEPS. THE STENT REMAINS IMPLANTED, AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24020813 ↗
  • InjurySep 22, 2025Break; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX-NAVIFLEX BILIARY STENT WAS TO BE IMPLANTED IN THE DUODENUM, PROCEDURE PERFORMED ON (b)(6) 2025.DURING THE PROCEDURE, AFTER RELEASING THE STENT, THE DELIVERY DEVICE FAILED TO DETACH. PULLING ON IT CAUSED A BREAK IN THE WORKING CHANNEL. THE PROXIMAL PART WAS REMOVED, WHILE THE DISTAL PORTION REMAINED IN THE DUODENUM, STILL ATTACHED TO THE STENT. THE DETACHED CATHETER WAS REMOVED USING FORCEPS, AND THE PROCEDURE WAS COMPLETED.THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23118198 ↗
  • MalfunctionMar 9, 2023Break; Positioning Failure; Break; Positioning Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC THAT AN ADVANIX BILIARY STENT WAS ATTEMPTED TO BE PLACED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE ON (B)(6) 2023. THE REPORTED INTENDED USE WAS FOR CBD DRAINAGE AFTER STONE REMOVAL. AFTER THE STENT WAS DELIVERED INTO THE COMMON BILE DUCT, THE DELIVERY SYSTEM GUIDE CATHETER COULD NOT BE WITHDRAWN. IT WAS REPORTED THAT WHEN GUIDE CATHETER WAS PULLED OUTWARD, THE EXTERNAL TIP BECAME FRACTURED. THE GUIDE CATHETER WAS SUCCESSFULLY REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER ADVANIX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16512747 ↗
  • MalfunctionMay 1, 2019Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT UPON REVIEWING HOSPITAL INVENTORY, IT WAS NOTICED THAT THERE WAS A HOLE IN THE PACKAGING OF AN ADVANIX RX BILIARY STENT AND THE STERILITY OF THE DEVICE WAS COMPROMISED. THERE WAS NO ISSUE NOTED WITH THE OUTER FEDERAL EXPRESS BOX THAT THE DEVICE WAS SHIPPED IN. THERE WAS NO PATIENT OR PROCEDURE INVOLVED AT THE TIME THIS INCIDENT WAS NOTICED. **ADDITIONAL INFORMATION RECEIVED ON 01MAY2019** EVENT DATE WAS REPORTED TO BE (B)(6) 2019.

    Patient problems: No Patient Involvement

    MAUDE report 8571328 ↗
  • MalfunctionNov 20, 2018Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX RX BILIARY PRELOAD STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE, PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT, IT WAS NOTICED THAT THE GUIDE CATHETER BROKE. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8091074 ↗