Advanix™ Biliary M00534680Recalls & reported problems
Boston Scientific Corporation · Biliary Stent with NaviFlexTM RX Delivery System · Class 2 · Implant
All models of Advanix Biliary (59)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08714729787235), received between Oct 28, 2018 and Aug 23, 2024 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Activation, Positioning or Separation Problem1
- Contamination1
- Difficult or Delayed Positioning1
- Material Deformation1
- Migration1
- Positioning Failure1
- Unsealed Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
- No Patient Involvement1
- Perforation1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryAug 23, 2024Break; Difficult or Delayed Positioning; Activation, Positioning or Separation Problem; Break; Difficult or Delayed Positioning; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WITH NAVIFLEX RX DELIVERY SYSTEM WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE (ERCP) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, IT WAS IMPOSSIBLE TO PULL THE SUTURE WHEN THE STENT WAS ATTEMPTED TO BE RELEASED. THE STENT WAS EVENTUALLY DEPLOYED AFTER SEVERAL ATTEMPTS. HOWEVER, THE GUIDE CATHETER GOT STUCK INSIDE THE STENT. THE GUIDE CATHETER WAS THEN REMOVED USING RAPTOR FORCEPS, AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20052630 ↗MalfunctionJul 24, 2022Contamination; Contamination
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS OPENED TO BE USED FOR AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO BE PERFORMED ON (B)(6) 2022. DURING THE PREPARATION, IT WAS NOTED THAT THE PACKAGE WAS DAMAGED AND THE SEAL WAS DESTROYED. THE STENT IS NO LONGER STERILE AND COULD NOT BE USED ANYMORE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ADVANIX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15090661 ↗MalfunctionJun 13, 2022Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE. THE PROCEDURE DATE WAS NOT REPORTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THAT THE GUIDE CATHETER WAS DETACHED, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT. SEE INVESTIGATION SUMMARY FOR MORE DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14674069 ↗InjuryJan 6, 2022Migration; Migration
IT WAS REPORTED THAT AN ADVANIX BILIARY STENT WAS IMPLANTED IN THE PATIENT ON AN UNKNOWN DATE. DURING TWO ROUTINE FOLLOW UPS IN THE CLINIC AFTER PLASTIC STENT PLACEMENT, IT WAS REPORTED THAT THE PATIENT CAME IN WITH PAIN IN THE MIDDLE OF THE NIGHT AND SHOWED THAT THE STENT HAD MIGRATED INTO THE SIGMOID COLON AND HAD A PERFORATION. THE STENT WAS REMOVED VIA SURGERY. NO NEW STENT WAS IMPLANTED IN PLACE OF THE MIGRATED STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE "EXPECTED TO FULLY RECOVER".
Patient problems: Perforation
MAUDE report 13181649 ↗MalfunctionNov 1, 2021Positioning Failure; Material Deformation; Positioning Failure; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING AN HEMOSTASIS PROCEDURE IN THE COMMON BILE DUCT ON (B)(6) 2021. DURING THE PROCEDURE, IT WAS NOTED THAT STENT WOULD NOT DISENGAGE FROM THE PUSHER. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ADVANIX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THE STENT WAS DAMAGED, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12727395 ↗MalfunctionMay 26, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE IN THE COMMON BILE DUCT PERFORMED ON (B)(6) 2021. DURING PREPARATION, WHEN THEY UNPACKED THE STENT, THE STENT WAS FOUND BROKEN. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ADVANIX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11883099 ↗MalfunctionFeb 13, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY IN THE BILE DUCT, PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHILE ATTEMPTING TO DEPLOY THE STENT, THE GUIDE CATHETER WAS COMPLETELY DETACHED FROM THE DELIVERY SYSTEM. REPORTEDLY, THE BROKEN INNER GUIDE CATHETER REMAINED WITHIN THE PUSH CATHETER, ENABLING THE DELIVERY SYSTEM TO BE REMOVED IN ONE PIECE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE ORIGINAL ADVANIX BILIARY STENT . THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9706639 ↗MalfunctionJul 31, 2019Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS IMPLANTED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, AFTER THE PULL WIRE WAS RETRACTED, THE PHYSICIAN HAD DIFFICULTY REMOVING THE GUIDEWIRE FROM THE PATIENT, WHEN ATTEMPTED TO PULL THE GUIDEWIRE WITH FORCE, THE GUIDEWIRE WAS MANAGED TO BE REMOVED FROM THE STENT. HOWEVER, DUE TO THE MANEUVER, THE GUIDEWIRE LOST ITS PTFE COATING IN PROXIMITY OF THE EXIT PORT AND WAS BLOCKED INSIDE THE GUIDE CATHETER. REPORTEDLY, THE SUTURE WAS STUCK AND THE STENT HAD A BREAK IN THE VICINITY OF THE FIN DUE TO THE MANEUVER YET THE ADVANIX BILIARY STENT WAS IMPLANTED SUCCESSFULLY. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE CONDITION OF THE PATIENT AT THE CONCLUSION OF THE PROCEDURE WAS NOTED TO BE GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8846402 ↗MalfunctionOct 28, 2018Unsealed Device Packaging; Unsealed Device Packaging
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT WHEN THE DEVICE WAS RECEIVED, IT WAS NOTICED THAT THE PACKAGING OF AN ADVANIX BILIARY STENT WAS BROKEN AND THE STERILITY OF THE DEVICE WAS COMPROMISED. THE EXACT DATE OF THIS EVENT WAS NOT REPORTED. THERE WAS NO PATIENT OR PROCEDURE INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 8011169 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page