Advanix™ Biliary M00534220Recalls & reported problems
Boston Scientific Corporation · Biliary Stent with NaviflexTM RX Delivery System · Class 2 · Implant
All models of Advanix Biliary (59)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 08714729787310), received between May 29, 2018 and Jan 12, 2026 — classified as 5 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Positioning Failure4
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Difficult to Remove1
- Material Deformation1
- Material Split, Cut or Torn1
- Positioning Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryJan 12, 2026Break; Use of Device Problem; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED IN THE COMMON BILE DUCT FOR STENT PLACEMENT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE WIRE BROKE FROM THE DELIVERY SYSTEM. THE STENT WAS PLACED BUT HAD TO BE REPOSITIONED FROM WITH RAT TOOTH FORCEPS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT THE GUIDEWIRE WAS IN THE PATIENT DURING THE ATTEMPTED DEPLOYMENT. HOWEVER, THE ADVANIX BILIARY STENT WITH NAVIFLEX RX DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU) STATES, "FOR PRELOADED SYSTEM ONLY: IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE ENDOSCOPE, THE STENT CAN NOT BE FULLY DEPLOYED." THE PHYSICIAN DID NOT FOLLOW THE STEPS CITED IN THE IFU. THEREFORE, AR CODE (5191: 2457) WAS USED TO CAPTURE USER ERROR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24039855 ↗InjuryNov 26, 2024Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS TO BE IMPLANTED IN THE BILIARY DUCT TO TREAT A BILIARY STENOSIS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE (ERCP) PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE ADVANIX STENT WAS DEPLOYED. HOWEVER, WHEN THE GUIDEWIRE WAS REMOVED, THE GUIDE CATHETER WAS STUCK INSIDE THE DEPLOYED STENT AND WAS NOTED TO BE BROKEN. THE DETACHED GUIDE CATHETER WAS REMOVED USING FORCEPS. THE STENT REMAINS IMPLANTED, AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20786785 ↗MalfunctionMay 13, 2024Positioning Failure; Material Split, Cut or Torn; Positioning Failure; Material Split, Cut or Torn
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY AND NAVIFLEX RX DELIVERY SYSTEM WAS TO BE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: THIS EVENT HAS BEEN DEEMED AN MDR-REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF A STENT BARB TORN. PLEASE SEE BLOCK H10 FOR THE FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19306957 ↗MalfunctionSep 29, 2023Positioning Failure; Positioning Failure
IT WAS REPORTED TO BOSTON SCIENTIFIC (BSC) THAT AN ADVANIX BILIARY STENT WITH NAVIFLEX RX DELIVERY SYSTEM WAS SELECTED FOR USE IN THE PAPILLA/DUODENUM DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6), 2023. DURING THE PROCEDURE, ALTHOUGH THE GUIDEWIRE AND INNER CATHETER WERE RETRACTED AS INSTRUCTED PER THE DEVICE'S INSTRUCTIONS FOR USE, THE STENT DID NOT DEPLOY. THE ADVANIX BILIARY STENT WAS REMOVED FROM THE PATIENT, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER ADVANIX BILIARY STENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THAT THE GUIDE CATHETER WAS DETACHED/SEPARATED, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT. PLEASE SEE BLOCK H10 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17844612 ↗InjuryMar 8, 2023Break; Positioning Failure; Retraction Problem; Break; Positioning Failure; Retraction Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC THAT AN ADVANIX BILIARY STENT WAS ATTEMPTED TO BE PLACED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE ON (B)(6) 2023. DURING STENT PLACEMENT IN A VERY SICK PATIENT WITH A TIGHT STRICTURE, THE STENT WOULD NOT RELEASE FROM THE DELIVERY SYSTEM. ATTEMPTS WERE MADE TO PULL THE STENT AND DELIVERY SYSTEM BACK THROUGH THE SCOPE, BUT THE GUIDE CATHETER AND STENT GOT STUCK IN THE BILE DUCT. THE DELIVERY SYSTEM AND HALF OF THE GUIDE CATHETER WERE REMOVED THROUGH THE SCOPE. THE OTHER HALF OF THE GUIDE CATHETER AND THE STENT WERE SUCCESSFULLY REMOVED FROM THE BILE DUCT USING RAT TOOTH FORCEPS. THE PROCEDURE TO CREATE FLOW IN THE DUCT WAS COMPLETED USING A NEW ADVANIX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16509277 ↗MalfunctionJun 5, 2022Break; Positioning Failure; Material Deformation; Adverse Event Without Identified Device or Use Problem; Break; Positioning Failure; Material Deformation; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX RX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE DUODENAL PAPILLA, PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT, THE STENT FAILED TO DEPLOY AND IT WAS NOTICED THAT THE PUSH CATHETER AND GUIDE CATHETER BROKE. THE BROKEN GUIDE CATHETER AND PUSH CATHETER REMAINED IN ONE PIECE WITHIN THE DELIVERY SYSTEM ENABLING IT TO BE REMOVED ON ONE PIECE. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE METHOD. A PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED AND SHOWED THAT THE PULLWIRE WAS KINKED AND THE PUSH CATHETER WAS BROKEN. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14599618 ↗MalfunctionApr 4, 2022Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE IN THE BILE DUCT PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, WHEN ADVANCING THE STENT INTO THE BILE DUCT, IT DEPLOYED PREMATURELY EVEN THOUGH THE HANDLE WAS IN THE LOCKED POSITION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ADVANIX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THAT THE STENT DAMAGED, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13995619 ↗InjuryFeb 23, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE COMMON BILE DUCT, PERFORMED ON (B)(6) 2021. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE AND INSIDE THE PATIENT, WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT, IT WAS NOTICED THAT THE GUIDE CATHETER BROKE. THE BROKEN PIECE OF GUIDE CATHETER WAS RETRIEVED WITH A USE OF A PAIR OF GRASPING FOCEPS. ANOTHER ADVANIX RX BILIARY STENT WAS OPENED AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE OKAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11366849 ↗MalfunctionFeb 8, 2021Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED IN THE BILIARY. THE EXACT DATE OF THE PROCEDURE WAS NOT REPORTED. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT, THE GUIDEWIRE WAS STUCK AND THE STENT FAILED TO DEPLOY. THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER ADVANIX BILIARY STENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF GUIDE CATHETER DETACHED/SEPARATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11298592 ↗InjuryMay 6, 2019Break; Premature Activation; Difficult to Remove; Break; Premature Activation; Difficult to Remove
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED IN THE BILE DUCT OF A PATIENT ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT THROUGH THE STRICTURE IN THE BILE DUCT, THE SUTURE BROKE AND THE STENT PREMATURELY DEPLOYED AND BECAME STUCK IN THE PATIENT'S DUCT. THE PHYSICIAN ATTEMPTED TO REMOVE THE STENT USING SNARES AND FORCEPS, HOWEVER, UPON ATTEMPTED REMOVAL, THE STENT TORE MULTIPLE TIMES. THE PROCEDURE WAS COMPLETED BY LEAVING THE REMAINING STENT IN PLACE AND THE BROKEN STENT PIECES WERE LEFT TO PASS OUT THE PATIENT NATURALLY. THE PHYSICIAN NOTED THAT THE TREATMENT LOCATION WAS A SUTURED BILE DUCT (SURGICAL COMPLICATION OF LAPAROSCOPIC CHOLECYSTECTOMY). THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8582224 ↗MalfunctionMay 29, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX RX BILIARY STENT WAS IMPLANTED IN A PATIENT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE IN THE COMMON BILE DUCT, PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING A PTC (PERCUTANEOUS TRANSHEPATIC CHOLANGIOGRAPHY) RENDEVOUS CASE WHERE THE GUIDEWIRE WAS PLACED BY INTERVENTIONAL RADIOLOGY, THE ADVANAIX STENT WAS PLACED USING THE GUIDEWIRE. WHEN THE PHYSICIAN WAS GOING TO DEPLOY THE STENT, HE REQUESTED THE GUIDEWIRE BE CUT BY THE BIOPSY CHANNEL OF THE DUODENOCOPE. THE PHYSICIAN REQUESTED THE GUIDEWIRE TO BE CUT TO PREVENT CONTAMINATION, AS THE PHYSICIAN WAS EXPECTING THIS PORTION OF THE GUIDEWIRE TO BE PULLED BACK THROUGH THE PATIENT'S LIVER. THE CUT PORTION OF THE GUIDEWIRE WAS ATTEMPTED TO BE REMOVED FROM THE PATIENT VIA THE PTC SITE BUT BECAME STUCK WHEN REMOVING FROM THE PATIENT. THE PHYSICIAN ASKED FOR THE STENT TO BE DEPLOYED WITH THE GUIDEWIRE STILL IN PLACE. AFTER DEPLOYMENT, THE GUIDEWIRE WAS STILL UNABLE TO BE REMOVED FROM THE PATIENT VIA THE PTC SITE, SO THE PHYSICIAN REMOVED THE DELIVERY CATHETER AND GUIDEWIRE THROUGH THE ERCP SCOPE. ALL DEVICES WERE REMOVED FROM THE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7549748 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page