Advanix™ Biliary M00534260Recalls & reported problems
Boston Scientific Corporation · Biliary Stent with NaviflexTM RX Delivery System · Class 2 · Implant
All models of Advanix Biliary (59)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729787358), received between Sep 1, 2021 and Dec 26, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Use of Device Problem3
- Material Deformation2
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Remove1
- Material Split, Cut or Torn1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Laceration(s)1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 26, 2025Break; Use of Device Problem; Break; Use of Device Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED IN THE BILE DUCT FOR A CHOLEDOCALITHIASIS DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE NURSE TRIED TO SHORTEN THE INTRODUCER BUT IT WAS NOT WORKING AND IT WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23899170 ↗MalfunctionFeb 16, 2024Break; Use of Device Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn; Break; Use of Device Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY WAS IMPLANTED IN THE BILE DUCT FOR BILE DRAINAGE DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, IT WAS NOTED THAT THERE WAS A THREAD HANGING ON THE STENT. THE ADVANIX BILIARY STENT REMAINS IMPLANTED, AND THE PROCEDURE WAS COMPLETED. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. NOTE: THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT BASED ON A MEDIA INSPECTION OF THE PHOTO PROVIDED BY THE CUSTOMER THAT DETERMINED THE STENT BARB WAS TORN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18727964 ↗InjuryDec 7, 2023Break; Use of Device Problem; Break; Use of Device Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY PRELOADED DOUBLE BEND STENT WAS IMPLANTED IN THE DUODENUM TO TREAT A BILE FLOW BLOCKAGE DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE TECHNICIAN PULLED THE STENT FORCEFULLY CAUSING THE BLACK INTRODUCER TO DETACH FROM THE DELIVERY SYSTEM INSIDE THE PATIENT. WHEN THE SCOPE WAS REMOVED, THE DETACHED BLACK INTRODUCER REMAINED IN THE PATIENT'S DUODENUM. THE TECHNICIAN ATTEMPTED TO RETRIEVE THE DETACHED INTRODUCER USING FORCEPS AND A SNARE BUT WAS UNSUCCESSFUL. AN UPPER SCOPE WAS THEN USED ALONG WITH A SNARE TO SUCCESSFULLY REMOVED THE DETACHED BLACK INTRODUCER. THE STENT REMAINED IMPLANTED, AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT THE GUIDEWIRE WAS INSIDE THE PATIENT DURING THE ATTEMPTED DEPLOYMENT. HOWEVER, THE ADVANIX BILIARY STENT WITH NAVIFLEX RX DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU) STATES, "FOR PRELOADED SYSTEM ONLY: IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE ENDOSCOPE, THE STENT CANNOT BE FULLY DEPLOYED." THE PHYSICIAN DID N
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18282443 ↗MalfunctionJan 12, 2022Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX RX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE BILE DUCT, PERFORMED ON AN (B)(6) 2021. POST PROCEDURE, ON (B)(6) 2021, WHEN THE PHYSICIAN ATTEMPTED TO REMOVE THE STENT WITH A SNARE, IT WAS DIFFICULT TO REMOVE. THE PROXIMAL FLANGE WAS PERPENDICULAR WHEN REMOVED AND CAUSED TRAUMA INSIDE THE PATIENTS' BILE DUCT AND HAD TO DRAIN THE FLUID. ANOTHER ADVANIX BILIARY STENT WAS IMPLANTED TO THE PATIENT AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WAS NO PATIENT COMPLICATION OF LACERATION REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO HAVE FULLY RECOVERED.
Patient problems: Laceration(s)
MAUDE report 13237892 ↗MalfunctionSep 1, 2021Break; Positioning Failure; Break; Positioning Failure
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX BILIARY STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE IN THE COMMON BILE DUCT PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE PHYSICIAN COULD NOT DEPLOY THE STENT AND IT WAS FOUND THAT THE SUTURE STRING BECAME BROKEN. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ADVANIX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THE STENT BROKEN, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12408577 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page