Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729793410), received between Jul 29, 2020 and Aug 23, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Difficult to Remove1
- Leak/Splash1
- Material Rupture1
Patient problems
- No Consequences Or Impact To Patient2
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 23, 2024Break; Difficult to Remove; Break; Difficult to Remove
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BLOOD LOSS OR HEMORRHAGE. A 5.0 X 120, 135CM MUSTANG WAS SELECTED FOR VASCULAR CASE IN THE RIGHT BRACHIAL ARTERY. DURING THE PROCEDURE, THE BALLOON CATHETER WAS INTRODUCED THROUGH A NON-BSC INTRODUCER SHEATH. WHILE INSERTION WAS SUCCESSFUL, REMOVAL OF THE BALLOON AFTER IT WAS USED WAS DIFFICULT. IT REQUIRED SIGNIFICANT FORCE, AND WHICH RESULTED IN STRETCHING OF THE SHEATH AND THEN THE BALLOON SNAPPED INSIDE THE SHEATHS HUB. THE SHEATH HAD TO BE REMOVED, AND MANUAL PRESSURE WAS APPLIED TO THE BRACHIAL ARTERY FOR REMAINDER OF CASE TO COMPLETE THE PROCEDURE FEMORALLY. CONSEQUENTLY, THE PATIENT EXPERIENCED BLOOD LOSS OR HEMORRHAGE DUE TO THIS EVENT. THE PROCEDURE WAS COMPLETED USING AN ALTERNATE METHOD. NO FURTHER COMPLICATIONS WERE REPORTED, AND THE PATIENT IS EXPECTED TO FULLY RECOVER.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20045018 ↗MalfunctionAug 26, 2020Leak/Splash; Leak/Splash
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 05AUG2020. IT WAS REPORTED THAT BALLOON LEAK OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 5.0 X 120, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR POST-DILATATION. HOWEVER, AFTER DILATATION, ANGIOGRAPHY IMAGE SHOWED LEAKING OF CONTRAST AGENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT CONDITION WAS FINE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED BALLOON PINHOLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10457045 ↗MalfunctionJul 29, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN AN UNOBSTRUCTED AND NON-TORTUOUS LEFT COMMON ILIAC ARTERY. A 5.0 X 120, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, ON INITIAL INFLATION, THERE WAS LEAKAGE OF DYE NOTED FROM THE BALLOON AND THE BALLOON COULD NOT INCREASE PRESSURE. A VERY SMALL HOLE WAS NOTED AT THE EDGE OF THE BALLOON. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10341083 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page