MUSTANG™ H74939171061510Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of MUSTANG (131)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Mustang balloon dilatation catheter is used to open narrowed peripheral arteries, such as in the leg, during angioplasty. The 31 reports mostly describe the balloon bursting or leaking during inflation, and in a few cases the balloon tip catching on a stent or coming off the wire inside the patient. One report involved surgery to remove balloon fragments; no deaths were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 08714729793458), received between Dec 3, 2018 and Nov 21, 2025 — classified as 1 injury, 30 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (24 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture24
- Detachment of Device or Device Component3
- Entrapment of Device2
- Break1
- Defective Device1
- Difficult to Remove1
- Failure to Advance1
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions17
- No Consequences Or Impact To Patient15
- Pain1
- Thrombosis/Thrombus1
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2025Fluid/Blood Leak; Inflation Problem; Fluid/Blood Leak; Inflation Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 06NOV2025. IT WAS REPORTED THAT BALLOON LEAK AND INFLATION FAILURE OCCURRED. THE 60 -70% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150,135CM MUSTANG BALLOON CATHETER WAS SELECTED FOR USE IN AN ANGIOPLASTY. DURING THE PROCEDURE, THE BALLOON WAS INFLATED AT 6 ATMOSPHERES, HOWEVER THE BALLOON LEAKED AND WAS UNABLE TO INFLATE. THE PROCEDURE WAS COMPLETED USING ORIGINAL METHOD OR DEVICE USED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED BALLOON TEAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23622292 ↗MalfunctionJun 30, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT THE BALLOON WAS BROKEN. A 6.0 X 150, 135CM, MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING INFLATION AT 14 ATMOSPHERES, THE BALLOON WAS BROKEN. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22360148 ↗MalfunctionFeb 4, 2025Difficult to Remove; Failure to Advance; Difficult to Remove; Failure to Advance
IT WAS REPORTED THAT DEVICE ENTRAPMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG WAS USED FOR DILATION. THE INNOVA SELF-EXPANDING STENT WAS PLACED IN THE ARTERY. WHEN THE BALLOON WAS ADVANCED INTO THE PLACED STENT, IT WAS NOTED THAT THE BALLOON TIP BECAME STUCK ON THE STENT STRUT AND THE BALLOON CATHETER COULD NOT PASS THROUGH. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21305824 ↗MalfunctionApr 30, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON THIRD INFLATION AT 24 ATMOSPHERES FOR 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19218101 ↗MalfunctionOct 4, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE SECOND INFLATION AT 20 ATMOSPHERES FOR 60 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS IN GOOD CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17867966 ↗InjurySep 8, 2023Material Rupture; Detachment of Device or Device Component; Material Rupture; Detachment of Device or Device Component
IT WAS REPORTED THAT THE BALLOON RUPTURED REQUIRING AN ARTERIOTOMY, TO REMOVE PORTIONS OF THE BALLOON. A 6.0 X 150, 135CM MUSTANG BALLOON WAS USED IN A BRACHIOCEPHALIC FISTULOPLASTY PROCEDURE FOR STENOSIS. THE 85% TO 90% STENOSED LESION WAS LOCATED IN THE BRACHIOCEPHALIC ARTERY, WITH MODERATE TORTUOSITY AND MILD CALCIFICATION. DURING THE PROCEDURE THE BALLOON WAS INFLATED TWO TIMES, FOR TWO MINUTES. DURING THE SECOND INFLATION, THE BALLOON RUPTURED AT 20 ATM. WHILE REMOVING THE BALLOON, A PORTION OF THE BALLOON REMAINED INSIDE THE BRACHIAL ARTERY. AT SOME POINT IN THE PROCEDURE THE PATIENT EXPERIENCED SEVERE PAIN AND A THROMBOSIS. LATER, THE PATIENT UNDERWENT AN ARTERIOTOMY TO REMOVE THE BALLOON MATERIAL. THE PATIENT HAS FULLY RECOVERED AND IS DOING BETTER. THE BALLOON WILL NOT BE RETURNED, AS IT WAS DISPOSED OF. IT WAS FURTHER REPORTED THAT THE MUSTANG BALLOON CATHETER DID NOT CONTRIBUTE TO THE PATIENT'S THROMBOSIS OR PAIN, AND THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; Pain; Thrombosis/Thrombus
MAUDE report 17712848 ↗MalfunctionApr 27, 2023Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. DURING UNPACKING OF A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER, IT WAS NOTICED THAT THE BALLOON AND THE SHAFT WAS COMPLETELY SEPARATED INTO TWO PIECES AT ABOUT 120CM FROM THE HUB. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16824892 ↗MalfunctionMar 8, 2023Defective Device; Detachment of Device or Device Component; Defective Device; Detachment of Device or Device Component
IT WAS REPORTED THAT BALLOON DETACHMENT OCCURRED. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING PROCEDURE, THE BALLOON FAILED AND CAME OFF THE WIRE IN THE PATIENT. NO PATIENT HARM RESULTED IN RELATION TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16504219 ↗MalfunctionFeb 27, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT BALLOON DETACHMENT OCCURRED. THE 89% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED RIGHT SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON WAS ADVANCED OVER AN 0.035 WIRE AND THE DEVICE WAS INFLATED AT 10 ATMOSPHERES FOR A MINUTE. WHILE REMOVING THE DEVICE, THE BALLOON RIPPED APART. THE PHYSICIAN WAS ABLE TO RETRIEVE THE BALLOON WITH NO COMPLICATION AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16443762 ↗MalfunctionJun 15, 2022Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT ENTRAPMENT OF THE DEVICE OCCURRED. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE PROCEDURE, IT WAS NOTED THAT THE DEVICE MATERIAL FAILED WITH THE GUIDEWIRE LOCKING IN ITS LUMEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14697876 ↗MalfunctionDec 4, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 80-90% STENOSED TARGET LESION WAS LOCATED IN A FISTULA IN THE MILDLY TORTUOUS AND MODERATE TO SEVERELY FIBROTIC BRACHIAL VESSEL. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, ON INITIAL INFLATION AT 20 ATMOSPHERES FOR 25 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12930693 ↗MalfunctionOct 31, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 20 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT WAS IN GOOD CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12725547 ↗MalfunctionOct 21, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT 18 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT STATUS WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12676947 ↗MalfunctionJun 30, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150MM, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 22 ATMOSPHERES, THE BALLOON RUPTURED AND A PINHOLE WAS NOTED AFTER THE BALLOON WAS WITHDRAWN. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE AND NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12096727 ↗MalfunctionMay 4, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED RIGHT SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11773290 ↗MalfunctionFeb 22, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, ON INITIAL INFLATION AT 20 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11358665 ↗MalfunctionDec 30, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN SEVERELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 20 ATMOSPHERES IN 10 SECONDS, THE BALLOON RUPTURED AND PINHOLE WAS NOTED ON THE BALLOON MATERIAL. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT STATUS WAS GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11094402 ↗MalfunctionDec 1, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED AND INITIALLY INFLATED AT 8 ATMOSPHERES (ATM) FOR 30 SECONDS. HOWEVER, DURING THE SECOND INFLATION AT 14 ATM FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10928572 ↗MalfunctionNov 5, 2020Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT CATHETER ENTRAPMENT OCCURRED. THE 50-60% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, AFTER DEFLATION, THE BALLOON REFUSED TO EXIT AND BECAME STUCK OVER THE WIRE. THE DEVICE WAS REMOVED WITH THE WHOLE SYSTEM AND COMPLETED THE PROCEDURE WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10793205 ↗MalfunctionJul 17, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 14 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10287650 ↗MalfunctionJul 15, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED, TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE THIRD INFLATION AT 20 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED WITHOUT ANY ISSURE NOTED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVCE. NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10280428 ↗MalfunctionJun 15, 2020Material Rupture; Material Rupture
IT WAS REPORTED BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION; HOWEVER, DURING INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10155718 ↗MalfunctionJun 3, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 15 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10113827 ↗MalfunctionJun 1, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT SUPERFICIAL FEMORAL ARTERY. A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10104020 ↗MalfunctionMar 20, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. AFTER CROSSING THE LESION WITH A GUIDEWIRE, A 5.00MM / 2.0CM / 135CM PERIPHERAL CUTTING BALLOON WAS ADVANCED FOR PRE-DILATATION. HOWEVER, DURING INITIAL INFLATION, THE BALLOON RUPTURED DUE TO SEVERELY CALCIFIED AREA PROTRUDING INSIDE THE BLOOD VESSEL. SUBSEQUENTLY, A 6.0 X 150, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED BUT THE BALLOON ALSO RUPTURED DURING THE THIRD INFLATION DUE TO CALCIFICATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9862194 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page