MUSTANG™ H74939171041010Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of MUSTANG (131)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08714729793717), received between Jul 11, 2019 and Jan 15, 2026 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture8
- Detachment of Device or Device Component1
- Material Integrity Problem1
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient4
- Foreign Body In Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryJan 15, 2026Nonstandard Device; Material Rupture; Detachment of Device or Device Component
IT WAS REPORTED THAT BALLOON DAMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE LOWER LIMB ARTERY. A 4.0 X 100, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED OVER THE GUIDEWIRE FOR DILATION. DURING INFLATION, THE BALLOON CRACKED. THE BALLOON WAS IMMEDIATELY WITHDRAWN AND A NEW ONCE WAS REPLACED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS 100% STENOSED, MILDLY TORTUOUS, AND SEVERELY CALCIFIED. ADDITIONALLY, THE DEVICE USED IN THIS PROCEDURE WAS A 3 MM X 150 MM STERLING BALLOON DILATION CATHETER. WHEN THE BALLOON REACHED THE RATED PRESSURE OF 6 ATM, A "POP" SOUND WAS HEARD. THE ANGIOGRAPHY REVEALED THAT THE BALLOON HAD RUPTURED INSIDE THE BODY. THEREFORE, THE DAMAGED BALLOON WAS SLOWLY REMOVED. AN EXTERNAL EXAMINATION OF THE "CORPSE" OF THE BALLOON SHOWED THAT IT WAS INCOMPLETE, AND SOME OF THE BALLOON'S OUTER SKIN HAD FALLEN INTO THE BLOOD VESSELS. ADDITIONALLY, IT WAS CONFIRMED THAT THE BALLOON DELIVERY CATHETER, TOGETHER WITH THE DAMAGED BALLOON, WAS REMOVED DIRECTLY ALONG THE GUIDEWIRE. THE BALLOON DEBRIS REMAINING IN THE BLOOD VESSELS WAS REMOVED USING A CAPTURE DEVICE. THE BALLOON WAS NOT INFLATED AND USED
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 24081665 ↗MalfunctionAug 7, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT TIP DAMAGE OCCURRED. A 4.0 X 100, 135 CM MUSTANG BALLOON CATHETER WAS SELECTED FOR USE. DURING INSERTION INTO THE GUIDE CATHETER, IT WAS NOTED THAT THE TIP HAS A BEVELED CUT. THE DEVICE WAS NOT USED IN THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22742350 ↗MalfunctionMay 15, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 4.0 X 100, 135CM MUSTANG BALLOON CATHETER WAS SELECTED FOR USE. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE SECOND INFLATION AT 18 ATMOSPHERES FOR 30SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22032382 ↗MalfunctionSep 29, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0 X 100, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 10 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15503230 ↗MalfunctionMay 17, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0 X 100, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. THE DEVICE WAS INFLATED TWICE AT 24 ATMOSHERES (ATM). HOWEVER,DURING THIRD INFLATION AT 24 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11832977 ↗MalfunctionDec 16, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA FEMORAL ARTERY UTILIZING A CONTRALATERAL APPROACH. THE STENOSED TARGET LESION WAS LOCATED IN THE RIGHT POPLITEAL ARTERY. AFTER A GUIDEWIRE CROSSED THE LESION, A 4.0 X 100, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING FIRST INFLATION AT 15 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS 80% STENOSED, MODERATELY TORTUOUS, AND MODERATELY CALCIFIED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11023972 ↗MalfunctionMay 5, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 4.0 X 100, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THIRD INFLATION AT 18 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10026521 ↗MalfunctionNov 7, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. DURING A PROCEDURE INVOLVING A 4.0 X 100, 135CM MUSTANG BALLOON CATHETER, THE BALLOON RUPTURED ON THE SECOND INFLATION AT 18 ATMOSPHERES FOR 60 SECONDS. THE DEVICE WAS REMOVED AND THE PROCEDURE COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9293203 ↗MalfunctionJul 11, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A 4.0 X 100, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8783135 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page