MUSTANG™ H74939171040210Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of MUSTANG (131)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729793724), received between Sep 29, 2020 and Jun 17, 2022 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Rupture3
- Deflation Problem1
- Detachment of Device or Device Component1
- Difficult to Remove1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJun 17, 2022Deflation Problem; Difficult to Remove; Detachment of Device or Device Component; Deflation Problem; Difficult to Remove; Detachment of Device or Device Component
IT WAS REPORTED THAT DEFLATION FAILURE, REMOVAL DIFFICULTY AND BALLOON DETACHMENT OCCURRED. THE TARGET LESION WAS LOCATED IN SUPERFICIAL FEMORAL ARTERY. A 4.0 X 20, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. THE BALLOON WAS INFLATED AT 10 ATMOSPHERES FOR 2 MINUTES. HOWEVER, DURING DEFLATION, THE BALLOON FAILED TO DEFLATE. TO REMOVE THE BALLOON, THEY FIRST PULLED NEGATIVE WITH THE STOPCOCK ON, THEN OFF, BUT IT STILL DID NOT COME DOWN. THE SHEATH WAS THEN USED TO SWALLOW THE BALLOON, BUT IT WAS DIFFICULT TO DO SO. THE SHAFT TO THE BALLOON BROKE AS IT WAS PULLED INTO THE SHEATH. THE SURGEON HAD TO POP THE BALLOON TO REMOVE IT. IT ENDED COMING OUT IN TWO PIECES AND DETACHED FROM THE SHAFT. THE PROCEDURE COULD NOT BE COMPLETED AS HE LOST ACCESS DUE TO THE REMOVAL OF THE BALLOON. ADDITIONAL PROCEDURE WAS TO BE RE-SCHEDULED, AND THE PATIENT WAS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14733984 ↗MalfunctionJan 26, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED ILIAC ARTERY. A 4.0 X 20, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, ON INITIAL INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11228527 ↗MalfunctionJan 19, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0 X 20, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 20 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED FROM THE PATIENT'S BODY WITHOUT PROBLEM AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11194484 ↗MalfunctionSep 29, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE CHRONIC TOTALLY OCCLUDED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED POPLITEAL ARTERY. A 4.0 X 20, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATIONS. HOWEVER, DURING INITIAL INFLATION BEFORE RATED BURST PRESSURE, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT'S STATUS WAS GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10603892 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page