MUSTANG™ H74939171070670Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of MUSTANG (131)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Mustang balloon catheter is used to widen narrowed blood vessels. The 26 reports mostly describe the balloon tearing or bursting during inflation, with procedures finished using another device and no patient harm mentioned. There were no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 08714729793861), received between May 14, 2019 and Apr 25, 2026 — classified as 26 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (25 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture25
- Adverse Event Without Identified Device or Use Problem2
- Leak/Splash2
- Inflation Problem1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- No Consequences Or Impact To Patient5
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 25, 2026Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE TARGET LESION, EXHIBITING 80% STENOSIS, WAS LOCATED WITHIN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED ARTERIOVENOUS FISTULA (AVF). A 7.0 X 60 MM, 75 CM MUSTANG BALLOON CATHETER WAS ADVANCED TO THE LESION SITE FOR DILATION. DURING INFLATION AT 23 ATMOSPHERES, THE BALLOON FAILED TO EXPAND, AND A TEAR IN THE DEVICE WAS NOTED. THE PROCEDURE WAS COMPLETED USING A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE NOTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24991826 ↗MalfunctionApr 1, 2026Material Rupture
IT WAS REPORTED THAT BALLOON BREAK OCCURRED. THE STENOSED TARGET LESION WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON BREAK. THE PROCEDURE WAS COMPLETED USING WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24755628 ↗MalfunctionFeb 5, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS SELECTED FOR USE. DURING THE PROCEDURE, THE BALLOON RUPTURED. NO ADDITIONAL INFORMATION WAS PROVIDED. IT WAS FURTHER REPORTED THAT THE 100% STENOSED TARGET LESION WAS LOCATED WITHIN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED ARTERIOVENOUS FISTULA. DURING THE PROCEDURE, THE BALLOON WAS INFLATED TWICE. THE PROCEDURE WAS SUBSEQUENTLY COMPLETED USING A NON-BOSTON SCIENTIFIC DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24274636 ↗MalfunctionJan 15, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 7.0 X 60, 75CM MUSTANG WAS ADVANCE FOR DILATION. IT WAS NOTED THAT THE BALLOON WAS RUPTURED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED ARTERIOVENOUS FISTULA (AVF). THE BALLOON RUPTURED UPON SECOND INFLATION FOR 1 MINUTE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24079745 ↗MalfunctionDec 22, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND NON-CALCIFIED VESSEL. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 16 ATMOSPHERES FOR 1 MINUTE. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23872603 ↗MalfunctionSep 4, 2025Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT AND RIGHT ILIAC ARTERY. TWO 7.0 X 60, 75CM MUSTANG BALLOONS CATHETER WERE SELECTED AS FOR USE. THE MUSTANG BALLOONS WERE USED IN A KISSING BALLOON ANGIOPLASTY PROCEDURE SIMULTANEOUSLY ON BOTH THE LEFT AND RIGHT ILIAC PRIOR TO STENT IMPLANTATION. HOWEVER, ONE OF THE MUSTANG BALLOONS BURST BELOW THE RATED BURST PRESSURE DURING THE FIRST INFLATION AFTER 10 SECONDS. THIS MUSTANG BALLOON WAS REMOVED AND REPLACED WITH ANOTHER MUSTANG BALLOON. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22964045 ↗MalfunctionAug 13, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 85-90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED CEPHALIC VEIN. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING FIRST INFLATION AT 12 ATMOSPHERES FOR LESS THAN 1 MINUTE, THE BALLOON BURST. THE BALLOON WAS COMPLETELY REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22800490 ↗MalfunctionJun 12, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION IN A FISTULOPLASTY PROCEDURE. DURING THE PROCEDURE, THE BALLOON RUPTURED AT RATED PRESSURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE BALLOON RUPTURED UPON SECOND INFLATION AT 10 ATMOSPHERES FOR ONE MINUTE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22197379 ↗MalfunctionApr 28, 2025Inflation Problem; Material Rupture; Material Integrity Problem; Adverse Event Without Identified Device or Use Problem; Inflation Problem; Material Rupture; Material Integrity Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT BALLOON DAMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT UPPER LIMB ARTERIOVENOUS FISTULA. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING INFLATION, IT WAS NOTED THAT THE BALLOON COULD NOT BE INFLATED. WHEN WITHDRAWN AND INFLATED OUTSIDE THE BODY, IT WAS NOTED THAT THE BALLOON WAS DAMAGED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21916621 ↗MalfunctionNov 26, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) IN THE LEFT CEPHALIC VEIN. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20789819 ↗MalfunctionNov 10, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN AN INTERNAL FISTULA IN THE RIGHT FOREARM RADIAL ARTERY TO CEPHALIC VEIN. AFTER OPENING THE CEPHALIC VEIN STENOSIS OF THE RIGHT UPPER ARM WITH A 4F SINGLE CURVED CATHETER AND 0.035 GUIDEWIRE, THE GUIDEWIRE ENTERED THE RIGHT SUBCLAVIAN VEIN. THE CATHETER WAS WITHDRAWN AND A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED ALONG THE GUIDEWIRE FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18109898 ↗MalfunctionNov 3, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE THIRD INFLATION AT 24-26 ATMOSPHERES FOR 30-60 SECONDS, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18064443 ↗MalfunctionJul 6, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE ILIAC ARTERY. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INITIAL INFLATION, THE BALLOON DEFLATED BEFORE IT REACHED TO ITS RATED BURST PRESSURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICES. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT STATUS WAS STABLE. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS 60% STENOSED, NON-TORTUOUS AND NON-CALCIFIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17264267 ↗MalfunctionMay 30, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND MODERATELY CALCIFIED SUBCLAVIAN ARTERY. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT 4 BAR AND BEFORE REACHING THE NOMINAL PRESSURE, THE BALLOON BURST. THE DEVICE WAS RETRIEVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17021084 ↗MalfunctionMay 24, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN SUPERFICIAL FEMORAL ARTERY. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE PROCEDURE, THE BALLOON BURST WHILE BLOWING INSIDE THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16996604 ↗MalfunctionMar 21, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION THE BALLOON BURST. NO PATIENT COMPLICATIONS REPORTED. IT WAS FURTHER REPORTED THAT THE 70% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED ARTERIOVENOUS FISTULA. THE BALLOON RUPTURED WHEN IT WAS INFLATED AT 14 ATMOSPHERES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16587167 ↗MalfunctionDec 19, 2022Material Rupture; Material Rupture
IT WAS REPORTED BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE CALCIFIED AND NON-TORTUOUS ARTERIOVENOUS FISTULA INNOMINATE VEIN. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING FIRST INFLATION AT BELOW 20 ATMOSPHERES FOR 5-10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16008255 ↗MalfunctionAug 19, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUBCLAVIAN ARTERY. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT NOMINAL BAR, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS 80% STENOSED, MODERATELY CALCIFIED AND NON-TORTUOUS. IN ADDITION, THE BALLOON RUPTURED ON THE FIRST INFLATION AT 3 ATMOSPHERES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15262587 ↗MalfunctionDec 7, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT PRESENTED FOR TREATMENT OF A LEFT ARTERIOVENOUS FISTULA. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, UPON INITIAL INFLATION, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12957659 ↗MalfunctionOct 5, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE AORTA ABDOMINALIS. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT LESS THAN RATED BURST PRESSURE FOR FEW SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12576681 ↗MalfunctionJun 9, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 60% STENOSED TARGET LESION WAS LOCATED IN A NON-TORTUOUS AND NON-CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS WAS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11970069 ↗MalfunctionNov 2, 2020Leak/Splash; Material Rupture; Leak/Splash; Material Rupture
IT WAS REPORTED THAT BALLOON PINHOLE OCCURRED. THE TARGET LESION WAS LOCATED IN A RENAL VESSEL. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, THE BALLOON COULD NOT BE INFLATED AND LEAKING OF CONTRAST WAS NOTICED DUE TO A PINHOLE ON THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10771577 ↗MalfunctionSep 24, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE ARTERIOVENOUS FISTULA. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT'S STATUS WAS GOOD. IT WAS FURTHER REPORTED THAT THE BALLOON RUPTURED DURING THE INITIAL INFLATION AT BELOW RATED BURST PRESSURE AND THE DEVICE WAS COMPLETELY REMOVED FROM THE PATIENT'S BODY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10579859 ↗MalfunctionMar 11, 2020Leak/Splash; Leak/Splash
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 19FEB2020. IT WAS REPORTED THAT BALLOON LEAK OCCURRED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. AFTER AN ELUVIA DRUG-ELUTING STENT WAS DEPLOYED, A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR POST-DILATION. HOWEVER, AFTER BALLOONING, CONTRAST AGENT APPEARED IN ANGIOGRAPHY. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS GOOD. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED BALLOON PINHOLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9819255 ↗MalfunctionFeb 5, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN A VESSEL IN THE ARM. A 7.0 X 60, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE SECOND INFLATION BELOW THE RATED BURST PRESSURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9672086 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page