Skip to content

MUSTANG™ H74939171120270Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of MUSTANG (131)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 08714729793922), received between Feb 6, 2019 and Aug 17, 2026 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture13
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient6

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 17, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50278474 ↗
  • MalfunctionJun 8, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED COMMON ILIAC ARTERY. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 10 ATMOSPHERES FOR 2 SECONDS, THE BALLOON RUPTURED FROM A PINHOLE AT NEAR THE MIDDLE. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25430720 ↗
  • MalfunctionOct 10, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED COMMON ILIAC ARTERY. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 8 ATMOSPHERES FOR 3 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20418732 ↗
  • MalfunctionSep 12, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED COMMON ILIAC ARTERY. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER DURING THE FIRST INFLATION AT 10 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17733629 ↗
  • MalfunctionNov 10, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS MODERATELY TORTUOUS AND SEVERELY CALCIFIED. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 14 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED AT THE CENTER PART OF THE BALLOON. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15773294 ↗
  • MalfunctionAug 22, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN A VESSEL. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING FIRST INFLATION AT NOMINAL PRESSURE, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT IS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15271456 ↗
  • MalfunctionMar 31, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 10 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED AND WAS VERTICALLY SEPARATED NEAR THE MIDDLE PART. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED AND THE PATIENT CONDITION WAS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13964164 ↗
  • MalfunctionJul 28, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS AN AREA OF IN-STENT RESTENOSIS OF A NON-BOSTON SCIENTIFIC STENT LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED ABDOMINAL AORTA. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 10 ATMOSPHERES FOR 1 SECOND, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND A PINHOLE WAS NOTED ON THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED AND THE PATIENT CONDITION WAS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12242291 ↗
  • MalfunctionAug 31, 2020Leak/Splash; Leak/Splash

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 10-AUG-2020. IT WAS REPORTED THAT HUB LEAK OCCURRED. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATAION. HOWEVER, PRESSURE COULD NOT BE APPLIED AND LEAKAGE WAS OBSERVED FROM THE HUB PART. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED BALLOON PINHOLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10473353 ↗
  • MalfunctionJul 25, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED ILIAC ARTERY. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, UPON FIRST INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS WAS STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8827908 ↗
  • MalfunctionJun 24, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED ILIAC ARTERY. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE SECOND INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8729820 ↗
  • MalfunctionMay 3, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN AN ILIAC ARTERY. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR PRE-DILATATION. HOWEVER, DURING THE FIRST INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8579581 ↗
  • MalfunctionMay 1, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE ILIAC VEIN. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR PRE-DILATATION. HOWEVER, DURING FIRST INFLATION THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT SERIOUS INJURY NOR COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8570281 ↗
  • MalfunctionFeb 6, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED ILIAC ARTERY. A 12.0 X 20, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR PRE-DILATATION. HOWEVER, UPON INFLATION AT 8 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A 12-40/5.8/75 MUSTANG BALLOON CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8312437 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page