Skip to content

MUSTANG™ H74939171081210Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of MUSTANG (131)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729794448), received between Apr 28, 2022 and Apr 5, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Leak/Splash1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 5, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BALLOON LEAK WAS ENCOUNTERED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE SUBCLAVIAN VEIN. AFTER A 6F SHEATH AND A 5F GUIDE CATHETER WERE CROSSED THE LESION, A 8.0 X 120, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, WHEN THE BALLOON WAS FULLY INFLATED AT 14 ATMOSPHERES, THE PRESSURE WAS FOUND TO BE INCREASED AND THE INFLATION WAS ABNORMAL. IT WAS NOTED THAT THE CONTRAST AGENT WAS OVERFLOWING, AND THE BALLOON WAS LEAKING GAS UPON CHECKING. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT CONDITION WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16683778 ↗
  • MalfunctionApr 28, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON PINHOLE OCCURRED. THE STENOSED TARGET LESION WAS LOCATED IN THE CENTRAL VEIN. A 0.035 GUIDE WIRE WAS USED TO DELIVER A 8.0 X 120, 135CM MUSTANG BALLOON CATHETER TO THE LESION AND THE PUMP WAS USED TO SLOWLY PRESSURIZE THE BALLOON TO A WORKING PRESSURE OF 10KPA FOR FEW SECONDS. HOWEVER, DURING FIRST INFLATION, THE BALLOON FAILED TO INFLATE, AND A CONTRAST LEAKAGE WAS NOTED. THE PUMP INDEX DROPPED RAPIDLY, AND IT WAS SLOWLY WITHDRAWN FROM THE BODY AND A BALLOON PINHOLE WAS NOTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14242048 ↗