Skip to content

MUSTANG™ H74939171080610Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of MUSTANG (131)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729794714), received between Sep 16, 2022 and Jul 25, 2024 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture2
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Failure to Deflate1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1
  • Hemorrhage/Blood Loss/Bleeding1
  • Perforation1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 25, 2024Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 86% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED ABDOMINAL AORTA. A 8.0 X 60, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER NOT EXCEEDING BURST PRESSURE IN THE FIRST INFLATION. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19833655 ↗
  • MalfunctionJun 21, 2024Leak/Splash; Leak/Splash

    REPORTABLE BASED ON THE DEVICE ANALYSIS COMPLETED ON 30MAY2024. IT WAS REPORTED THAT BALLOON LEAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE CORONARY VEIN, AND THE PATIENT UNDERWENT A TRANSJUGULAR PORTAL SHUNT. AN 8.0 X 60, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED TO DILATE A NON-BOSTON SCIENTIFIC STENT. DURING THE PROCEDURE, INDIRECT PORTAL VEIN ANGIOGRAPHY WAS CONDUCTED TO VISUALIZE THE RELATIONSHIP BETWEEN THE PORTAL VEIN AND THE LIVER VEIN. DIRECT PORTAL VEIN ANGIOGRAPHY REVEALED THAT THE MAIN PORTAL VEIN WAS UNOBSTRUCTED, THOUGH THE GASTRIC CORONARY VEIN APPEARED THICKENED AND SLUGGISH. PORTAL PRESSURE WAS RECORDED AT 37 MMHG. HOWEVER, UPON INFLATION, THE BALLOON LEAKED, AND PRESSURE COULD NOT BE MAINTAINED EVEN AFTER THE BALLOON WAS REMOVED FROM THE PATIENT. THE DEVICE WAS REPLACED FOR ANOTHER OF THE SAME MODEL TO COMPLETE THE PROCEDURE. NO COMPLICATIONS REPORTED, AND PATIENT STATUS WAS STABLE. HOWEVER, DEVICE RETURNED ANALYSIS REVEALED A PINHOLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19587909 ↗
  • MalfunctionAug 1, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA). THE TARGET LESION WAS LOCATED IN THE COMMON ILIAC ARTERY. AFTER A .035 UNSPECIFIED GUIDEWIRE WAS INSERTED, AN 8.0 X 60, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION; HOWEVER, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED THAT THE STENOSED TARGET LESION WAS TOTALLY OCCLUDED LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED COMMON ILIAC ARTERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17435857 ↗
  • InjuryNov 8, 2022Failure to Deflate; Failure to Deflate

    IT WAS REPORTED THAT BALLOON DEFLATION OCCURRED RESULTING IN ADDITIONAL INTERVENTION. THE TARGET LESION WAS LOCATED IN CEPHALIC VEIN. AN 8.0 X 60, 135CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION DURING THE TREATMENT OF AN ARTERIOVENOUS FISTULA THROMBOSIS FOR HEMODIALYSIS. HOWEVER, DURING PROCEDURE THE BALLOON FAILED TO DEFLATE. THE MAXIMUM PRESSURE HAS NOT BEEN EXCEEDED. THE PHYSICIAN TRIED TO USE INSUFFLATOR IN CONVENTIONAL WAY TO DEFLATE THE DEVICE AND SYRINGE THEREAFTER BUT FAILED. THE INSUFFLATOR USED WAS DISCONNECTED BUT THE BALLOON REMAINED FULL AND WITH SIGNIFICANT PRESSURE. THE BALLOON WAS PUNCTURED PERCUTANEOUSLY AND REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THE PATIENT WAS STABLE AND THERE WERE NO PATIENT COMPLICATIONS AND INJURIES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15756393 ↗
  • InjurySep 16, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT BALLOON DETACHMENT AND UN-RETRIEVED DEVICE FRAGMENT OCCURRED; THE PATIENT EXPERIENCED PERFORATION, BLOOD LOSS, PAIN, AND VOMITING. BOTH GROINS WERE ACCESSED AND 5FR SHEATHS WERE INSERTED. THE CLOSE TO 100% STENOSED, CHRONIC TOTALLY OCCLUDED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND SEVERELY CALCIFIED RIGHT ILIAC ARTERY IN THE COMMON-EXTERNAL ILIAC JUNCTION. A 6X60X135 MUSTANG BALLOON CATHETER WAS FIRST USED TO PREDILATE THE AREA AND THE BALLOON WAS REMOVED WITHOUT ISSUE. A 8.0 X 60, 135CM MUSTANG BALLOON CATHETER WAS THEN ADVANCED FOR FURTHER DILATATION AND WAS HELD DILATED FOR 30 SECONDS BEFORE IT WAS DEFLATED. THE BALLOON WAS PULLED OUT WITH INTENSE CARE AS TO NOT OBSTRUCT THE VIEW OF THE AREA JUST TREATED AND THEN TOOK A SUBTRACTED RUN. A PERFORATION THEN WAS DETECTED. THE PHYSICIAN RE-ADVANCED THE SAME BALLOON BACK INTO POSITION AND RE-INFLATED UNTIL THE STENT COULD BE PREPPED FOR DEPLOYMENT. WHEN THE STENT WAS PREPPED, THE BALLOON WAS THEN DEFLATED. HOWEVER, DURING REMOVAL THE DEVICE, THE WHOLE BALLOON CAME OFF THE SHAFT. THE WIRE ALONG WITH THE SHAFT CAME COMPLETELY OUT OF THE BODY. THE PHYSICIAN RE-INSERTED THE GUIDEWIRE QUICKLY AND DEPLOYED TWO

    Patient problems: Foreign Body In Patient; Hemorrhage/Blood Loss/Bleeding; Perforation

    MAUDE report 15437981 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page